EDX-10 Series
Compact Recording System

  • CE

EDX-10series

Summary

· Compact & lightweight
· With stacked connection, no synchronization cable is required, therefore wiring-saving.
· Synchronous sampling up to 20 kHz for 4 channels
· Sensors can be easily connected with input cables or input adapters.

Features

The EDX-10 Series is measuring instrument, which can measure strain, voltage, and temperature (Thermocouples) simply by connecting a PC via USB interface.
Up to 4 measuring units for 16 channels can be powered by USB interface, no separated power supply is required.

Products table

All display

EDX-10B
Control Unit

EDX-10B
  • CE

Specifications

Interfaces USB2.0 compliant
Connector configuration: USB standard B receptacle
Installed Measuring Units Max. 4 (16 channels)
Sampling Frequencies 1 Hz to 20 k Hz (1 to 4 channels)
1 Hz to 10 k Hz (1 to 8 channels)
1 Hz to 5 k Hz (1 to 16 channels)
Operating Temperature 0 to 40ºC
Power Supply 5 VDC by USB bus power or a AC adapter
Notes:
1. When power supply from a USB port, please connect the EDX-10B to the PC directly. No USB hub is required.
2. The AC adapter operates any 4 measuring units.
3. The combination of measuring units for power supply by USB port are as follows.
Current Consumption 140 mA or less (5 VDC)
Weight Approx. 170 g
Dimensions 84.0 W x 26.6 H x 84.0 D mm (Excluding protrusions)
Control Software DCS-100A
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Standard Accessories USB cable N-38 (1 m)
Ground wire P-72 (5 m)
DVD (Dynamic data acquisition software DCS-100A)
Optional Accessories AC adapter UN310-0515 (For U.S.A.: UN312-0520)

Dimensions

EDX-10B

EDX-14A
Strain Measuring Unit

EDX-14A
  • CE

Specifications

Measuring Targets Strain-gage transducers, strain gages
*Bridge boxes or input adapters are required for strain measurement.
Channels 4
Measuring Range 10 k, 50 k ×10-6 strain (2 steps)
Applicable Bridge Resistance 120 Ω to 1 kΩ
Bridge Excitation 1 VDC
Gage Factor 2.00 fixed
Range Accuracy Each range within ±0.3%FS
Nonlinearity Within ±0.1%FS
A/D Converter 24 bits
Frequency Response DC to 2 kHz
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 100 Hz, 2 k Hz
Operating Temperature 0 to 40ºC
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 140 mA or less (120 Ω load with all channels connected, at power supply 5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Standard Accessories Strain input cable U-124 (30 cm)
Optional Accessories Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A
Bridge adapter for quarter bridge system UI-53A-120/350
Bridge adapter for quarter bridge system UI-54A-120/350
One-touch type input adapter UI-55A

Dimensions

EDX-14A

EDX-15A
Low-power Strain/Voltage Measuring Unit

EDX-15A
  • CE

Specifications

Measuring Targets Strain-gage transducers, strain gages
*Bridge boxes or input adapters are required for strain measurement.
Voltage
Channels 4
Measuring Range 10 k, 50 k ×10-6 strain (2 steps) (Strain-gage transducers, strain gages)
10, 50 V(Voltage)
Applicable Bridge Resistance 120 Ω to 1 kΩ
Bridge Excitation 1 VDC
Gage Factor 2.00 fixed
Range Accuracy Each range within ±0.3%FS
Nonlinearity Within ±0.1%FS
A/D Converter 24 bits
Frequency Response DC to 2 kHz
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 100 Hz, 2 k Hz
Operating Temperature 0 to 40ºC
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 150 mA or less (120 Ω load with all channels connected, at power supply 5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Standard Accessories Strain input cable U-124 (30 cm)
Conversion adapter FV-1A x4
Optional Accessories Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A
Bridge adapter for quarter bridge system UI-53A-120/350
Bridge adapter for quarter bridge system UI-54A-120/350
One-touch type input adapter UI-55A

Dimensions

EDX-15A

EDX-12A
Voltage Measuring Unit

EDX-12A
  • CE

Specifications

Measuring Targets Voltage
Channels 4 (Single end)
Measuring Range 10 V, 50 V (2 steps)
Range Accuracy Each range within ±0.3%FS
Nonlinearity Within ±0.1% FS
A/D Converter 24 bits
Frequency Response DC to 2 kHz
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 100 Hz, 2 k Hz
Operating Temperature 0 to 40ºC
Input Connectors D-sub 37-pin connector
Power Supply 5 V DC supplied by control unit
Current Consumption 110 mA or less (5 V DC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Standard Accessories Input adapter UI-51A
Optional Accessories BNC input cable U-125 (30 cm)
Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
One-touch type input adapter UI-52A

Dimensions

EDX-12A

EDX-13A
Thermocouple Measuring Unit

EDX-13A
  • CE

Specifications

Measuring Targets Thermocouples
Channels 4
Applicable Thermocouples K, T, J, N (Resistance of thermocouple: 1 kΩ or less)
(See the table below for details about the temperature measuring range, etc.)
Check Functions Burnout check
A/D Converter 24 bits
Sampling System Scanning
Inside Samping Frequencies Approx. 0.5 Hz, approx. 2.0 Hz
Input Connectors Screw type terminal box
Power Supply 5 V DC supplied by control unit
Current Consumption 120 mA or less (5 V DC)
Weight Approx. 130 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Range K: -200.0 to 1370.0ºC
T: -200.0 to 400.0ºC
J: -200.0 to 1200.0ºC
N: -200.0 to 1300.0ºC
Accuracy (Resolution: 0.1 °C )
K
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1370.0°C: ± (0.1% of reading + 0.4ºC)
T
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 400.0°C: ± (0.1% of reading + 0.4ºC)
J
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1200.0°C: ± (0.1% of reading+0.4ºC)
N
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1300.0°C: ± (0.1% of reading+0.4ºC)
Accuracy of internal reference contact compensator ±1.0°C 〔Input terminal temperature at equilibrium〕
〔Ambient temperature: 25±10°C〕
Mount the EDX-13A on the bottom when using it with measuring units other than the EDX-13A.
±2.0°C 〔Input terminal temperature in equilibrium〕
For temperatures other than those in the ambient temperature and operating temperature described above
Standard Accessories Terminal box 1piece, screwdriver 1piece

Dimensions

EDX-13A

EDX-11A
Strain Measuring Unit

EDX-11A
  • CE

Specifications

Measuring Targets Strain-gage transducers, strain gage
*Bridge boxes or input adapters are required for strain measurement.
Channels 4
Measurement Range 10k, 50k ×10-6 strain (2 steps)
Applicable Bridge Resistance 120 Ω to 1 kΩ
Bridge Excitation 2 VDC
Gage Factor 2.00 fixed
Range Accuracy Each range within ±0.3%FS
Nonlinearity Within ±0.1% FS
A/D Converter 24 bits
Frequency Response DC to 2 kHz
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 100 Hz, 2 k Hz
Operating Temperature 0 to 40ºC
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 180 mA or less(120 Ω load with all channels connected, at power supply 5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Standard Accessories Strain input cable U-124 (30 cm)
Optional Accessories Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A, UI-55A
Bridge adapter for quarter bridge system UI-53A-120/350
Bridge adapter for quarter bridge system UI-54A-120/350

Dimensions

EDX-11A

Page Top

Download