EDX Series Conditioner Cards

  • CE
  • TEDS

Conditioner Cards

Summary

· Conditioner cards for EDX-100A, EDX-200A, and EDX-5000A.
EDX-5000A does not support CAN card CAN-41A.

Features

The Cards are for measuring strain, voltage, acceleration, and temperature. CAN data is supported. Isolated amplifier cards and carrier-wave amplifier cards also available for countermeasure against noise.

Products table

All display

CDV-40B
Strain/Voltage Measurement Card

CDV-40B-F
  • CE
  • TEDS

Specifications

Input Channels 8 (Centralized connector)
Input Types Strain Measurement: Balanced differential.
Voltage Measurement: Unbalanced
Input Resistance Strain Measurement: Approx. 10 + 10 MΩ.
Voltage Measurement: Approx. 1 MΩ
Coupling DC/AC (DC cut)
Gage Factor 2
Applicable Bridge Resistance 120 Ω to 1 kΩ
Bridge Power Supply 2 VDC ±2%
Balance Adjustment Resistance ±2.4% (±12000 μm/m)
Measurement Range Strain Measurement: 500, 1 k, 2 k, 5 k, 10 k, 20 k, 50 k μm/m, OFF
Voltage Measurement: 0.1, 0.2, 0.5, 1, 2, 5, 10 V, OFF
Accuracy: Each range ±0.2%FS of each range
Calibration Value ±100%, ±50% of each range
Nonlinearity ±0.1%FS
Frequency Response DC coupling: DC to 50 kHz, dev. + 1 dB, -3 dB
AC coupling (DC cut): 0.2, 1 Hz to 50 kHz (Refer to high-pass flter. )
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 k, 3 k, 10 kHz and FLAT (8 steps)
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
HPF Cutoff Frequency : 0.2, 1 Hz
Attenuation : -6 dB/oct.
A/D Resolution 16 bits
Added Functions Reading of information using TEDS built-in sensor
Applicable Standard
EMC Directive EN61326-1 (Class A)
Optional Accessories Conversion adapter FV-1A
8-channel input cable U-38 (1.5 m), U-39 (1 m), U-40 to 48 (50 cm)

CDV-40B-F
Strain/Voltage Measurement Card, with Antialiasing LPF

CDV-40B-F
  • CE
  • TEDS

Specifications

Input Channels 8 (Centralized connector)
Input Types Strain Measurement: Balanced differential.
Voltage Measurement: Unbalanced
Input Resistance Strain Measurement: Approx. 10 + 10 MΩ.
Voltage Measurement: Approx. 1 MΩ
Coupling DC/AC (DC cut)
Gage Factor 2
Applicable Bridge Resistance 120 Ω to 1 kΩ
Bridge Power Supply 2 VDC ±2%
Balance Adjustment Resistance ±2.4% (±12000 μm/m)
Measurement Range Strain Measurement: 500, 1 k, 2 k, 5 k, 10 k, 20 k, 50 k μm/m, OFF
Voltage Measurement: 0.1, 0.2, 0.5, 1, 2, 5, 10 V, OFF
Accuracy: Each range ±0.2%FS of each range
Calibration Value ±100%, ±50% of each range
Nonlinearity ±0.1%FS
Frequency Response DC coupling: DC to 50 kHz, dev. + 1 dB, -3 dB
AC coupling (DC cut): 0.2, 1 Hz to 50 kHz (Refer to high-pass flter. )
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 k, 3 k, 10 k Hz and FLAT (8 steps)
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
Antialiasing Filter Transfer Characteristic : 8th order Butterworth
Notes: Provided that LPF is set to AUTO on DCS-100A.
Cutoff Frequency : Automatic setting at x0.25 sampling frequency
Attenuation : -48 dB±5 dB (At sampling frequency x 0.5)
HPF Cutoff Frequency : 0.2 Hz, 1 Hz
Attenuation : -6 dB/oct.
A/D Resolution 16 bits
Applicable Standard
EMC Directive EN61326-1 (Class A)
Added Functions Reading of information using TEDS built-in sensor
Optional Accessories Conversion adapter FV-1A
8-channel input cable U-38 (1.5 m), U-39 (1 m), U-40 to 48 (50 cm)

CVM-41A
Strain/Voltage/Acceleration Measurement Card

CVM-41A
  • CE
  • TEDS

Strain Measurement Mode

Applicable Sensors Strain measurement: Strain gages, Strain-gage transducers
*When measuring with the strain gage, using the bridge box.

Voltage Measurement: Voltage, Voltage output type sensors

Acceleration measurement (Piezoelectric): Piezoelectric accelerometer (Built-in amplifier)

Specifications

Input Channels 8
Input Types Strain measurement: Balanced differential input

Voltage Measurement: Balanced differential input
*When using the voltage conversation adapter FV-1A and the voltage input box VI-8A, this becomes unbalanced input.
*Common mode input voltage range ±20 VDC, absolute maximum input voltage range ±50 VDC.

Acceleration measurement (Piezoelectric): Unbalanced input
*The voltage conversation adapter FV-1A usage possible.
Input Impedance Voltage Measurement: (1 MΩ + 1 MΩ) Within ±10%
*When using the voltage conversation adapter FV-1A (At unbalanced input), within 1 MΩ±10%.
Bridge Power Supply Strain measurement: Constant voltage output (BV 2V: 2 VDC, BV 5V: 5 VDC, BV 10V: 10 VDC)

Voltage Measurement: Constant voltage output (2 VDC (±1 V), 5 VDC (±2.5 V), 10 VDC (±5 V) or OFF, 20 mA/CH or less)

Acceleration measurement (Piezoelectric): Constant current output (Approx. 4 mA, Excitation voltage: Approx. 23 VDC, Load: 1 kΩ or less)

*When installed in EDX-100A, number of channels that can be set to bridge excitation BV10 V or sensor power supply 10 VDC (±5 V) mode, will be installed number of CVM-41Ax3 channels or less.
Gage Factor Strain measurement: 2.00 fixed
Applicable Bridge Resistance Strain measurement:
BV 2V: 120 to 1000 Ω, BV 5V: 350 to 1000 Ω, BV 10V: 500 to 1000 Ω
Balance (Zero suppression) Strain measurement:
[Auto-balance enabled]
Cancel the unbalanced bridge portion in the analog circuit, and zero the measurement value.
[Auto-balanced disabled]
Do not cancel the unbalanced bridge portion.
(The initial unbalanced value in the bridge circuit can be confirmed. )

Voltage Measurement:
[Zero suppression enabled]
Cancel the input voltage in the analog circuit, and zero the measurement value.
[Zero suppression disabled]
Do not cancel the input voltage in the analog circuit.
(Display the input voltage as is. )
Balance Adjustment Strain measurement:
BV 2V: Resistance ±10% (±50000 μm/m)
BV 5V: Resistance ±4% (±20000 μm/m)
BV 10V: Resistance ±2% (±10000 μm/m)

Voltage Measurement: ±5 V
Measurement Range Strain measurement:
BV2V: 2 k, 5 k, 10 k, 20 k, 50 k, 100 k, 200 k, 500 k μm/m
BV5V: 2 k, 5 k, 10 k, 20 k, 50 k, 100 k, 200 k μm/m
BV10V: 2 k, 5 k, 10 k, 20 k, 50 k, 100 k μm/m

Voltage Measurement:
1, 2, 5, 10, 20, 50 V

Acceleration measurement (Piezoelectric):
100, 200, 500, 1000, 2000, 5000 mV
Range Accuracy Strain measurement:
Within ±0.2%FS

Voltage Measurement:
Within ±0.2%FS

Acceleration measurement (Piezoelectric):
Within ±1.0%FS
Nonlinearity Strain measurement:
Within ±0.1%FS

Voltage Measurement:
Within ±0.1%FS

Acceleration measurement (Piezoelectric):
Within ±0.2%FS
Calibration Value Strain measurement: ±100% and ±50% of each range and SHUNT
*When SHUNT CAL has 350 Ω load connected, Approx. 257 μm/m output.

Voltage Measurement: ±100% and ±50% of each range

Acceleration measurement (Piezoelectric): ±100% and ±50% of each range
Frequency Response Strain measurement:
DC coupling: DC to 5 kHz (Deviation + 1 dB, -3 dB)
coupling: 0.2, 1 Hz to 5 kHz (Deviation + 1 dB, -3 dB)
(Refer to HPF)

VAColtage measurement:
DC coupling: DC to 5 kHz (Deviation + 1 dB, -3 dB)
coupling: 0.2, 1 Hz to 5 kHz (Deviation + 1 dB, -3 dB)
(Refer to HPF)

Acceleration measurement (Piezoelectric):
0.5 Hz to 5 kHz (Deviation + 1 dB, -3 dB)
LPF Transfer Characteristics : 5th order Butterworth
Cutoff Frequency : 30, 100, 300, 1 k, 3 k Hz, FLAT and AUTO
*With AUTO settings, the cutoff frequency is set to 1/4 of the sampling frequency.
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -30±3 dB/oct.
HPF Cutoff Frequency : 0.2, 1 Hz
Attenuation : 6 dB/oct.
A/D Resolution 24 bits
*When installed in EDX-100A, will operate in 16-bit resolution.
Distortion Factor Acceleration measurement (Piezoelectric): 1% or less
Monitor Output Accuracy: ±5 V±0.5% (With ±FS)
Non linearity: Within ±0.5%FS
Dimensions 22 W x 119 H x 213 D mm (Excluding protrusions)
Weight Approx. 400 g
Added Functions Reading of information using TEDS built-in sensor
Remark EDX-10A and EDX-2000A/B series are not supported.
Applicable Standard
EMC Directive EN61326-1 (Class A)
Optional Accessories CVM input cable U-121 to 123
Input conversion adapter FV-1A
Centralized output cable U-62
CCA input cable U-111
CVM concentrated input cable N-121
Voltage input box VI series VI-8A (-T)
Small-sized bridge box for 1-gage system DBS-120A-8 (T), DBS-350A-8 (T)
One-touch lock type bridge box DB-120V-8, DB-350V-8

DPM-42B
Dynamic Strain Measurement Card

  • CE
  • TEDS

Specifications

Applicable Sensors Strain gages, Strain-gage transducers
Input Channels 4
Frequency Response DC to 5 kHz (Deviation: ±10%)
Carrier Wave Frequencies 12 kHz
Applicable Bridge Resistance 120 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Power Supply 2 Vrms, 0.5 Vrms switching, 12 kHz sine wave
Balance Adjustment Resistance: ±2.4% (±12000 μm/m)
Capacity: 2000 pF
Balance Adjustment Methods Resistance: pure electronic auto balance (Saved in nonvolatile memory).
Capacity: CST method (Automatic tracking)
Measurement Range With bridge excitation 2 Vrms: 200, 500, 1000, 2000, 5000, 10000, 20000 μm/m and OFF, 8 steps.
With bridge excitation 0.5 Vrms: 1000, 2000, 5000, 10000, 20000, 50000 μm/m and OFF, 7 steps.
Calibration Value Output at ±100% and ±50% of each range
Nonlinearity Within ±0.2% FS
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 kHz and FLAT, 6 steps
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
A/D Resolution 16 bits
Added Functions Input check functions: Input resistance into one side of the bridge, and check input
Reading of information using TEDS built-in sensor
Monitor Output Accuracy: Within ±5 V±0.5% (With ±FS)
Nonlinearity: Within ±0.5% FS
Applicable Standard
EMC Directive EN61326-1 (Class A)
Optional Accessories Monitor output cable U-64 (2 m)

DPM-42B-F
Dynamic Strain Measurement Card, with Antialiasing LPF

DPM-42B-F
  • CE
  • TEDS

Specifications

Applicable Sensors Strain gages, Strain-gage transducers
Input Channels 4
Frequency Response DC to 5 kHz (Deviation: ±10%)
Carrier Wave Frequencies 12 kHz
Applicable Bridge Resistance 120 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Power Supply 2 Vrms, 0.5 Vrms switching, 12 kHz sine wave
Balance Adjustment Resistance: ±2.4% (±12000 μm/m)
Capacity: 2000 pF
Balance Adjustment Methods Resistance: pure electronic auto balance (Saved in nonvolatile memory).
Capacity: CST method (Automatic tracking)
Measurement Range With bridge excitation 2 Vrms: 200, 500, 1000, 2000, 5000, 10000, 20000 μm/m and OFF, 8 steps.
With bridge excitation 0.5 Vrms: 1000, 2000, 5000, 10000, 20000, 50000 μm/m and OFF, 7 steps.
Calibration Value Output at ±100% and ±50% of each range
Nonlinearity Within ±0.2% FS
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 kHz and FLAT, 6 steps
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
Antialiasing Filter Transfer Characteristic : 8th order Butterworth, Notes:
Provided that LPF is set to AUTO on DCS-100A.
Cutoff Frequency : Automatic setting at x0.25 sampling frequency
Attenuation : -48 dB±5 dB (At sampling frequency x 0.5)
A/D Resolution 16 bits
Added Functions Input check functions: Input resistance into one side of the bridge, and check input
Reading of information using TEDS built-in sensor
Monitor Output Accuracy: Within ±5 V±0.5% (With ±FS)
Nonlinearity: Within ±0.5% FS
Applicable Standard
EMC Directive EN61326-1 (Class A)
Optional Accessories Monitor output cable U-64 (2 m)

DPM-42B-I
Dynamic Strain Measurement Card, Low Inverter Noise Type

  • CE
  • TEDS

Specifications

Applicable Sensors Strain gages, Strain-gage transducers
Input Channels 4
Frequency Response DC to 5 kHz (Deviation: ±10%)
Carrier Wave Frequencies 12 kHz
Applicable Bridge Resistance 120 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Power Supply 2 Vrms, 0.5 Vrms switching, 12 kHz sine wave
Balance Adjustment Resistance: ±2.4% (±12000 μm/m)
Capacity: 2000 pF
Balance Adjustment Methods Resistance: pure electronic auto balance (Saved in nonvolatile memory).
Capacity: CST method (Automatic tracking)
Measurement Range With bridge excitation 2 Vrms: 200, 500, 1000, 2000, 5000, 10000, 20000 μm/m and OFF, 8 steps.
With bridge excitation 0.5 Vrms: 1000, 2000, 5000, 10000, 20000, 50000 μm/m and OFF, 7 steps.
Calibration Value Output at ±100% and ±50% of each range
Nonlinearity Within ±0.2% FS
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 kHz and FLAT, 6 steps
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
A/D Resolution 16 bits
Added Functions Input check functions: Input resistance into one side of the bridge, and check input
Reading of information using TEDS built-in sensor
Monitor Output Accuracy: Within ±5 V±0.5% (With ±FS)
Nonlinearity: Within ±0.5% FS
Withstand Voltage
Applicable Standard
EMC Directive EN61326-1 (Class A)
Optional Accessories Monitor output cable U-64 (2 m)

DPM-42B-I-F
Dynamic Strain Measurement Card, with Antialiasing LPF, Low Inverter Noise Type

DPM-42B-I-F
  • CE
  • TEDS

Specifications

Applicable Sensors Strain gages, Strain-gage transducers
Input Channels 4
Frequency Response DC to 5 kHz (Deviation: ±10%)
Carrier Wave Frequencies 12 kHz
Applicable Bridge Resistance 120 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Power Supply 2 Vrms, 0.5 Vrms switching, 12 kHz sine wave
Balance Adjustment Resistance: ±2.4% (±12000 μm/m)
Capacity: 2000 pF
Balance Adjustment Methods Resistance: pure electronic auto balance (Saved in nonvolatile memory).
Capacity: CST method (Automatic tracking)
Measurement Range With bridge excitation 2 Vrms: 200, 500, 1000, 2000, 5000, 10000, 20000 μm/m and OFF, 8 steps.
With bridge excitation 0.5 Vrms: 1000, 2000, 5000, 10000, 20000, 50000 μm/m and OFF, 7 steps.
Calibration Value Output at ±100% and ±50% of each range
Nonlinearity Within ±0.2% FS
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 k Hz and FLAT, 6 steps
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
Antialiasing Filter Transfer Characteristic : 8th order Butterworth
Notes: Provided that LPF is set to AUTO on DCS-100A.
Cutoff Frequency : Automatic setting at x0.25 sampling frequency
Attenuation : -48 dB±5 dB (At sampling frequency x 0.5)
A/D Resolution 16 bits
Added Functions Input check functions: Input resistance into one side of the bridge, and check input
Reading of information using TEDS built-in sensor
Monitor Output Accuracy: Within ±5 V±0.5% (With ±FS)
Nonlinearity: Within ±0.5% FS
Applicable Standard
EMC Directive EN61326-1 (Class A)
Optional Accessories Monitor output cable U-64 (2 m)

CTA-40A
Thermocouple Card

  • CE

Specifications

Applicable Sensors Thermocouples
Input Channels 8
Applicable Thermocouples K, T
Thermocouple Resistance 200 Ω or less (With burnout ON)
1000 Ω or less (With burnout OFF)
Measurement Range K1230, K480, K240, T400, T210 and OFF, 6 steps
K1230 (-200 to 1230ºC)
K480 (-200 to 480ºC)
K240 (-200 to 240ºC)
T400 (-200 to 400ºC)
T210 (-200 to 210ºC)
General Accuracy ambient temperature 20±3ºC, within ± (0.5%rdg+1) ºC
At ambient temperature 0 to 40ºC, within ± (0.5%rdg+2) ºC
Calibration Value Output at 100% and 50% and 0ºC as absolute value of each range
Frequency Response DC to 10 Hz
A/D Resolution 16 bits
Burnout Built-in: At burnout [Burnout display], with ON/OFF
(Notes: If high thermocouple resistance, turn the burnout function OFF to improve accuracy
Monitor Output Accuracy: Within 5 V±0.5% (At + FS)
Nonlinearity: Within ±0.5%FS
Withstand Voltage
Applicable Standard
EMC Directive EN61326-1 (Class A)
Standard Accessories 8 channel input cable U-104 x1, temperature measuring adapter CT-2A x8
Optional Accessories Integrated output cable U-62 (1.1 m)

CFV-40A
F/V Converter Card

  • CE

Specifications

Applicable Sensors Alternating signal output sensors
Input Channels 4
Input Signals Alternating (Zero cross), TTL level (Including open collector signals)
Input Voltage ± (0.5 V to 50 V): High hysteresis
± (0.1 V to 50 V): Low hysteresis
Measurement Range 0, 100, 500, 1 k, 2 k, 5 k, 10 k, 20 kHz and OFF, 9 steps
Accuracy: Within ±0.1%FS
Calibration Value Output at 100%, 50% (Added), and 0% (Absolute value) of each range
Response Time Below 10 μs (When the input pulse is repeated) two cycles of input frequency + below 50 μs (When the input pulse is broken)
A/D Resolution 16 bits
Power Supply to Sensors Within 12 VDC (Each channel 50 mA or less)
Monitor Output Accuracy: Within 5 V±0.5% (At + FS)
Nonlinearity: Within ±0.1%FS
Withstand Voltage
Remark This card can be mounted up to two cards on the EDX-2000A/B-32 mainframe. When two cards are mounted on the EDX-2000A/B-64, up to four other cards can be mounted, and when one card is mounted, up to six other cards can be mounted
Applicable Standard
EMC Directive EN61326-1 (Class A)
Standard Accessories Conversion adapter FV-1A x4
Optional Accessories Input cable U-12 (1.5 m), U-13 (1.5 m)
Monitor output cable U-64 (2 m)

CCA-40A
Charge Amplifier Card

  • CE
  • TEDS

Specifications

Applicable Sensors piezoelectric accelerometer
Input Channels 8
Power Supply to Sensors rated current power (Rated current: 4 mA, excitation voltage: approx. 24 VDC, load 1 kΩor less)
Frequency Response 1 to 20 kHz (Deviation: + 1 dB, -3 dB)
Measurement Range 20, 50, 100, 200, 500, 1000, 2000, 5000 mV and OFF, 9 steps.
Accuracy: Within ±1%FS
Calibration Value DC CAL: ±100% and ±50% of each range, Accuracy: Within ±0.2%FS
AC CAL: 100% and 50% of each range, Accuracy: Within ±1%FS Frequency accuracy: Within 100 Hz±5%
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 300, 1 k, 3 k, 10 k, and FLAT (5 steps)
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
Distortion Factor 1% or less
A/D Resolution 16 bits
Monitor Output Accuracy: Within 5 V±1% (At +FS)
Added Functions Reading of information using TEDS built-in sensor
Applicable Standard
EMC Directive EN61326-1 (Class A)
Standard Accessories Input cable U-111 (1 m)
Optional Accessories ntegrated output cable U-62 (1.1 m),
Input cable U-109 (1 m)
Conversion adapter (BNC-miniature) BNCP-C25J-A
Conversion adapter (Miniature-Tajimi) CCA-1B
Conversion adapter (BNC-miniature) CCA-2B

CCA-40A-F
Charge Amplifier Card, with Antialiasing LPF

CCA-40A-F
  • CE
  • TEDS

Specifications

Applicable Sensors piezoelectric accelerometers
Input Channels 8
Power Supply to Sensors rated current power (Rated current: 4 mA, excitation voltage: approx. 24 VDC, load 1 kΩor less)
Frequency Response 1 to 20 kHz (Deviation: + 1 dB, -3 dB)
Measurement Range 20, 50, 100, 200, 500, 1000, 2000, 5000 mV and OFF, 9 steps
Accuracy: Within ±1%FS
Calibration Value DC CAL: ±100% and ±50% of each range, Accuracy: Within ±0.2%FS
AC CAL: 100% and 50% of each range, Accuracy: Within ±1%FS Frequency accuracy: Within 100 Hz±5%
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 300, 1 k, 3 k, 10 k Hz, and FLAT (5 steps)
Amplitude Ratio at Cutoff Point : -3±1 dB
Attenuation : -12±1 dB/oct.
Antialiasing Filter Transfer Characteristic : 8th order Butterworth type
Notes: Enabled when "AUTO" set in EDX-2000A/B LPF settings
Cutoff Frequency : Automatic setting at x0.25 sampling frequency
Attenuation : -48 dB±5 dB (At sampling frequency x 0.5)
Distortion Factor 1% or less
A/D Resolution 16 bits
Monitor Output Accuracy: Within 5 V±1% (At +FS)
Added Functions Reading of information using TEDS built-in sensor
Applicable Standard
EMC Directive EN61326-1 (Class A)
Standard Accessories Input cable U-111 (1 m)
Optional Accessories Integrated output cable U-62 (1.1 m)
Input cable U-109 (1 m)
Conversion adapter (BNC-miniature) BNCP-C25J-A
Conversion adapter (Miniature-Tajimi) CCA-1B
Conversion adapter (BNC-miniature) CCA-2B

CAN-41A
CAN Card, Measurement ID Numbers Max. 32ID

CAN-41A
  • CE

Specifications

Number of CAN ports 2 (2 nodes)
Connector Configuration Dsub 9-pin (Male)
Supported CAN Version CAN 2.0B active support (ISO-11898 specifications-compliant)
High-speed CAN/low-speed CAN switching
Measurement ID Numbers Max. 32 ID
CAN Controller Operation Clock 40 MHz, 32 MHz
Baud rates With high-speed CAN: 1000, 800, 500, 250, 125, 100, 83.3, 62.5, 50, 33.3, 25, 20, and and 10 kbps.
With low-speed CAN: 125, 100, 83.3, 62.5, 50, 33.3, 25, 20, and and 10 kbps
Communication Conditions Sample points, sample rotations, resynchronization jump width selection
Graph Display Simultaneous graph display of numerical value display, frame display, and analog data
Others Only one can be mounted in the unit, in the last slot When measuring CAN data, the sampling frequency is restricted to EDX-3000B: 10 kHz, EDX-200A: 2048 Hz, EDX-100A: 1 kHz
Channel Conditions Start bit, bit length, data type, correction coefficient
(Conditions for converting CAN data to clipping physical quantity)
Applicable Standard
EMC Directive EN61326-1 (Class A)

CDV-44AS
Strain/Voltage Measurement Insulation Card

  • TEDS

Specifications

Applicable Sensors Strain gage (Full bridge), Strain-gage transducers, voltage
Input Channels 4
Input Resistance Strain Measurement: Approx. 10 + 10 MΩ.
Voltage Measurement: Approx. 1 MΩ
Input Types Strain Measurement: Balanced differential.
Voltage Measurement: Unbalanced
IMRR 120 dB (When 500 μm/m range)
Gage Factor 2.00 fixed (Strain mode)
Applicable Bridge Resistance 120 Ω to 1 kΩ (Strain mode)
Frequency Response With DC coupling: DC to 5 kHz, deviation within + 1 dB, -3 dB.
with AC coupling (DC cut): 0.2 Hz (Refer to HPF)
Bridge Power Supply Within 2 VDC±2%
Range Accuracy Within ±0.3% FS
Balance Adjustment Within ±2.4% (±12000 μm/m) (At strain measurement).
Within ±5 V (At voltage measurement)
Measurement Range Strain Measurement: 500, 1 k, 2 k, 5 k, 10 k, 20 k μm/m, OFF
Voltage Measurement: 1, 2, 5, 10, 20, 50 V, OFF
Accuracy: Each range ±0.3%FS
ZERO Accuracy Within ±0.3% FS (Voltage OFF mode)
Nonlinearity Within ±0.1% FS
Calibration Value Output at ±100% and ±50%of each range.
Accuracy: Within ±0.3%FS
Monitor Output Accuracy: Within ±5 V±0.5% (±5 V of full scale of each range)
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 k Hz, FLAT
Amplitude Ratio at Cutoff Point : Within -3±1 dB
Attenuation : whitin -12±1 dB/oct.
HPF Cutoff Frequency : 0.2 Hz
Attenuation : whthin -6±1 dB/oct.
A/D Resolution 16 bits
Added Functions Reading of information using TEDS built-in sensor
Withstand Voltage
Standard Accessories Insulation amplifier voltage conversion adapter (FV-2A x4)
Optional Accessories Monitor output cable U-64 (2 m)

CDA-44AS
Direct Rated Current Expansion Card, Applicable Bridge Resistance 120 Ω

  • TEDS

Specifications

Applicable Sensors Strain gages (Full bridge), Strain-gage transducers, voltage
Input Channels 4
Input Resistance Approx. 10 MΩ + 10 MΩ (Strain mode).
Approx. 1 MΩ (Voltage mode)
Input Types Balanced differential input (Strain mode)
Unbalanced input (Voltage mode)
IMRR 120 dB (When 500 μm/m range)
Frequency Response With DC coupling: DC to 200 Hz, deviation within + 1 dB, -3 dB
with AC coupling (DC cut ): 0.2 Hz (Refer to HPF)
Gage Factor 2.00 fixed (Strain mode)
Bridge Power Supply Approx. DC16.7 mA (Rated current)
*If sensitivity or temperature resistance is in the transducer bridge exitation line, then sensitivity and temperature characteristics are not corrected
Cable Length Applicable Bridge Resistance Cable length within 500 m (When cross-section 0.5 mm 2)
Applicable Bridge Resistance 120 Ω (Strain mode)
Range Accuracy Within ±0.3% FS
Measurement Range 500, 1 k, 2 k, 5 k, 10 k, 20 k μm/m, OFF (Strain mode)
1, 2, 5, 10, 20, 50 V, OFF (Voltage mode)
Balance Adjustment Within ±2.4% (±12000 μm/m) (At strain measurement).
Within ±5 V (At voltage measurement)
ZERO Accuracy Within ±0.3% FS (Voltage OFF mode)
Nonlinearity Within ±0.1% FS
Calibration Value Output at ±100% and ±50% of each range.
Accuracy: Within ±0.3%FS
Monitor Output Accuracy: Within ±5 V±0.5%
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 1, 3, 10, 30, 100 Hz and FLAT, 6 steps
Amplitude Ratio at Cutoff Point : Within -3±1 dB
Attenuation : whitin -12±1 dB/oct.
HPF Cutoff Frequency : 0.2 Hz
Attenuation : Within -6±1 dB/oct.
A/D Resolution 16 bits
Added Functions Reading of information using TEDS built-in sensor
Withstand Voltage
Standard Accessories Insulation amplifier voltage conversation adapter (FV-2A x4)

CDA-45AS
Direct Rated Current Expansion Card, Applicable Bridge Resistance 350 Ω

CDA-45AS
  • TEDS

Specifications

Applicable Sensors Strain gage (Full bridge), Strain-gage transducers, voltage
Input Channels 4
Input Resistance Approx. 10 + 10 MΩ (Strain mode).
Approx. 1 MΩ (Voltage mode)
Input Types Balanced differential input (Strain mode)
Unbalanced input (Voltage mode)
IMRR 120 dB (When 500 μm/m range)
Frequency Response With DC coupling: DC to 200 Hz, deviation within + 1 dB, -3 dB
with AC coupling (DC cut ): 0.2 Hz (Refer to HPF)
Gage Factor 2.00 fixed (Strain mode)
Bridge Power Supply Approx. DC5.7 mA (Rated current)
*If sensitivity or temperature resistance is in the transducer BV line, then sensitivity and temperature characteristics are not corrected
Cable Length Applicable Bridge Resistance Cable length within 1000 m (When cross-section 0.5 mm 2)
Applicable Bridge Resistance 120 Ω (Strain mode)
Range Accuracy Within ±0.3% FS
Measurement Range 500, 1 k, 2 k, 5 k, 10 k, 20 k μm/m, OFF (Strain mode)
1, 2, 5, 10, 20, 50 V, OFF (Voltage mode)
Balance Adjustment Within ±2.4% (±12000 μm/m) (At strain measurement).
Within ±5 V (At voltage measurement)
ZERO Accuracy Within ±0.3% FS (Voltage OFF mode)
Nonlinearity Within ±0.1% FS
Calibration Value Output at ±100% and ±50% of each range.
Accuracy: Within ±0.3%FS
Monitor Output Accuracy: Within ±5 V±0.5%
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 1, 3, 10, 30, 100 and FLAT, 6 steps
Amplitude Ratio at Cutoff Point : Within -3±1 dB
Attenuation : whitin -12±1 dB/oct.
HPF Cutoff Frequency : 0.2 Hz
Attenuation : Within -6±1 dB/oct.
A/D Resolution 16 bits
Added Functions Reading of information using TEDS built-in sensor
Withstand Voltage
Standard Accessories Insulation amplifier voltage conversation adapter (FV-2A x4)

AD-40AS
A/D Converter

Specifications

Input Channels 8
Measurement Range ±5, ±10 V, OFF
Input Types Unbalanced (Not balanced differential)
Input Resistance Approx. 1 MΩ
A/D Resolution 16 bits (± 32000 counts/FS)
Method: Successive approximation
Conversion Accuracy Within ±0.2% FS
Nonlinearity Within ±0.1% FS
Input Frequencies DC to 50 kHz Deviation: 1 to -3 dB
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 k, 3 k, 10 kHz and FLAT (8 steps)
Amplitude Ratio at Cutoff Point : Within -3±1 dB
Attenuation : -12±1 dB/oct.
Power Supply to Sensors ±2.5 V each channel
Added Functions TEDS compatible (Load TEDS information)
Optional Accessories 8-channel input cable: U-127 (1.5 m)
Voltage input box: VI-8A with a cable N-121 (1.5 m)

AD-40AS-F
A/D Converter, with Antialiasing LPF

Specifications

Input Channels 8
Measurement Range ±5 V, ±10 V, OFF
Input Types Unbalanced (Not balanced differential)
Input Resistance Approx. 1 MΩ
A/D Resolution 16 bits (± 32000 counts/FS)
Method: Successive approximation
Conversion Accuracy Within ±0.2% FS
Nonlinearity Within ±0.1% FS
Input Frequencies DC to 50 kHz Deviation: 1 to -3 dB
LPF Transfer Characteristics : 5th order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 k, 3 k, 10 k Hz and FLAT (8 steps)
Amplitude Ratio at Cutoff Point : Within -3±1 dB
Attenuation : -12±1 dB/oct.
Antialiasing Filter Transfer Characteristic : 8th order Butterworth
Cutoff Frequency : Automatic setting at x0.25 sampling frequency
Attenuation : Within (-48±5 dB) /oct. * (Set LPF to [AUTO])
Power Supply to Sensors ±2.5 V each channel
Added Functions TEDS compatible (Load TEDS information)
Optional Accessories 8-channel input cable: U-127 (1.5 m)
Voltage input box: VI-8A with a cable N-121 (1.5 m)

CDV-46AS
Strain/Voltage Measurement Isolation Card

CDV-46AS
  • TEDS

Specifications

Applicable Sensors Strain gages (Full-bridge system)
Strain-gage transducers, voltage
Input Channels 4
Input Types Balanced differential input (Strain mode)
Unbalanced input (Voltage mode)
Input Resistance Approx. 10 + 10 MΩ (Strain mode)
Approx. 1 MΩ (Voltage mode)
IMRR 120 dB (When 2k ×10-6 strain range)
Frequency Response With DC coupling DC to 5 kHz,
deviation within +1 dB, -3 dB
DC cut (With AC coupling) 0.2 Hz
(See the HPF)
Gage Factor 2.00 fixed (Strain mode)
Bridge Power Supply Within 2 VDC±2% (Strain mode)
Applicable Bridge Resistance 120 to 1000 Ω (Strain mode)
Measurement Range 2k, 5k, 10k, 20k, 50k, 100k ×10-6 strain, OFF
(At strain measurement)
1, 2, 5, 10, 20, 50, OFF (At voltage measurement)
Range Accuracy Within ±0.3%FS
Balance Adjustment Within ±2.4% (±12000 ×10-6 strain)
(At strain measurement)
Within ±5 V (At voltage measurement)
ZERO Accuracy Within ±0.3% FS (Voltage OFF mode)
Nonlinearity Within ±0.1%FS
Calibration Value Output at ±100% and ±50% of each range
Accuracy: Within ±0.3%FS
LPF Transfer Characteristics : 2nd order Butterworth
Cutoff Frequency : 10, 30, 100, 300, 1 kHz and FLAT (6 steps)
Amplitude Ratio at Cutoff Point : -3 ±1 dB
Attenuation : (-12±1) dB/oct.
HPF Cutoff Frequency : 0.2 Hz
Attenuation : Within (-6±1) dB/oct.
A/D Resolution 24 bits
Monitor Output Accuracy: Within ±5 V±0.5% (±5 V of full scale of each range)
TEDS Reads information from TEDS-installed sensors.
Withstand Voltage Between input and case (output):
Withstand voltage 500 VDC, 1 min.
Between channels: Withstand voltage 500 VDC, 1 min.
Standard Accessories Conversion adapter FV-2A × 4

Page Top