UCAM-550A Series
Fast Data Logger

  • TEDS

Summary

· All channels simultaneous sampling
· Synchronously sample at 50 Hz for 1000 channels
· 5 kinds of measuring units available

Features

The UCAM-550A Series is a Fast Data Logger, which can repeatedly measure a maximum of 1000 channels at an interval of 0.02 s.
Being capable of high-speed synchronous measurement, this unit can measure a wide range of phenomena, from static to dynamic phenomena.

Block diagrams

UCAM-550A

Products table

All display

UCAM-550A
Fast Data Logger

Specifications

Model UCAM-550A: With DCS-100A
UCAM-550A-0: Without DCS-100A

DCS-100A UCAM-550A Control Specifications

Interfaces LAN

Specifications

Channels Maximum of 50 channels/unit (Mixed palletizing is possible up to 5
units of the measuring unit)
(Each measuring unit can measure 10 channels per one unit.)
Measurement is possible up to 1000 channels at maximum using the
multi-unit synchronous operation.
*The public command corresponds up to 20 units (Maximum 1000
channels).
*DCS-100A corresponds up to 6 units (Maximum 300 channels).
Measuring Range
Measuring Units Note : *Indication of channel number is different such as;
・51A series; Start from "0"
・51B series; Start from "1"
Sampling Frequencies 1, 2, 10, 20, 50 Hz
*Response frequencies depends on the measuring units.
USS-51A/B, USV-51A/B, UST-51A/B: DC to 7.8 Hz
(Deviation: 0.5 to −3.5 dB)
Measuring Functions Original measurement
Major measurement
Interfaces 10 BASE-T, 100BASE-TX
Between PC-UCAM
LAN cable (Straight) Max. 30 m
Between UCAM-UCAM
STP straight cable (See notes) Max. 100 m
Note: "STP" in STP cable is short for Shield Twisted Pair, and an STP cable is a shielded LAN cable
Display LCD (20 digits x 2 lines)
Status display LED: POWER (When power ON, lit green)
MASTER (When master, lit green, when sleep, not lit)
TRANSFER (When communications, flashing green)
Sampling Formats Simultaneous sampling of all channels)
Operation Keys UP, down, left, right key
Storage Measurement data is saved in a PC (No internal storage)
Operating Temperature Operating Temperature: 0 to 40ºC
Operating Humidity: 20 to 85%RH (Non-condensing)
Power Supply 100 to 240 VAC
Approx. 50 VA (With 5 USS-51B strain units installed,
and 120 Ω load on all channels connected)
Dimensions 426 W x 132.5 H x 305 D mm
(Excluding protrusions)
Weight Approx. 7 kg (With 5 USS-51B strain unit installed)
Standard Accessories DVD (DCS-100A, instruction manual)
AC power cable P-18 (With 2-pin conversion plug CM-39)
Ground wire
Measuring Range
USS-51B
Strain Gages and Strain-gage Transducers Quarter bridge system: 120 Ω, 350 Ω
Half bridge system 120 to 1000 Ω: Active-dummy System, Active-active System
Full bridge system 120 to 1000 Ω: Opposite side active System, Bridge System
Potentiometer Sensors -50% to 50%
USV-51B
Voltage ±20.0 V
UST-51B
Temperature Thermocouple K, T, E, J, and R

DCS-100A UCAM-550A Control Specifications

Connection Units Max. 6 (Max. 300 channels)
Max. 20 (Max.1000 channels), optional software
DCS-106 is required.
Recorded The measured data is stored to the PC hard disk as a KS2 file.
Sampling Frequencies 1, 2, 10, 20, and 50 Hz
Measuring Modes Manual, manual (Preset data points)
Interval, and analog trigger
Measuring Functions Major, original
Major: Measured values = Sensor output values - Initial values
Original: Measured valuse = Sensor output values
Calibration Factor Computation ON/OFF setting in all channels of one batch
 Calibration factor compensation: Measured value X Calibration factor + Offset
Channel Conditions Measurement, mode, range,
calibration factor, offset, unit, initial value, CH name,
measuring range, decimal digits, upper limit, lower limit, rated capacity, rated output
(Selection of any item for display is possible)
Initial Value Measurement Measures the initial value of each sensor.
Manual Measurement Measurement is made from a press of the REC
button to a press of the STOP button or to
completion of recording to the preset data points.
Interval Measurement Measurement is made automatically at preset
intervals from the preset starting time.
Analog Trigger Measurement Start and/or stop recording based on specified trigger conditions
(Trigger level value is fixed)
End Trigger : Settable
Delay : Both start and end max. 3000 points/channel
Trigger Channels : Any channel
Trigger Level : Sets in physical quantity
Trigger Slope : Up, down
Changing Stroke Changes the data, before the stroke and after the
stroke, when using a displacement transducer.

DCS-100A NTB-500A Control Specification

Static Measurement Every time the DCS-100A starts recording data, the DCS-100A additionally saves the moving averaged
measured data in a single CSV format file in manual and interval modes.

DCS-100A UCAM-550A Control Specifications

Burnout Check For USM-51B/52B only
TEDS Reads sensor's information and sets to channel condition automatically. (USM-51B/52B only)

Dimensions

UCAM-550A

USS-51B
Strain Unit

USS-51B

Specifications

Channels 10
Gage Factor 2.00 fixed
(Correction is possible at 2.00/Ks with the engineering value conversion function)
Bridge Excitation 2 VDC constant voltage
(Applied constuntly)
Power Supply to Potentiometers 2 VDC constant voltage
Measuring Targetss Strain gages and Strain gage transducers
Potentiometer sensors
Measuring Range, Resolution, Accuracy (DC Input) (Strain Range L)
Measuring range: 0 to ±19000 μm/m, Resolution: 1 μm/m, Accuracy: ±0.05% FS
(Strain Range H)
Measuring range: 0 to ±200000 μm/m, Resolution: 10 μm/m, Accuracy: ±0.05% FS
(Potentiometer Sensor)
Measuring Range: 0 to ±50%, Resolution: 0.01%, Accuracy: ±0.1% FS
Note: Measuring range is indicated when the initial measurement and the original measurement are performed. In the case of a major measurement, the value is of the initial measurement value deducted in advance from the original measurement value.
Optional Input section protection/terminal cover for protection from dust UT-50A

UST-51B
Thermocouple Unit

UST-51B

Specifications

Channels 10
Measuring Targetss Temperature (Thermocouples)
Measuring Range, Resolution, Accuracy (DC Input) (K Range L)
Measuring range: -200.0 to 437.0ºC, Resolution: 0.1ºC, Accuracy: ±0.8ºC
(K Range H)
Measuring range: -200.0 to 1200.0ºC, Resolution: 0.1ºC, Accuracy: ±2.8ºC
(T)
Measuring range: -200.0 to 350.0ºC, Resolution: 0.1ºC, Accuracy: ±0.7ºC
(E Range L)
Measuring range: -200.0 to 260.0ºC, Resolution: 0.1ºC, Accuracy: ±0.5ºC
(E Range H)
Measuring range: -200.0 to 800.0ºC, Resolution: 0.1ºC, Accuracy: ±1.7ºC
(J Range L)
Measuring range: 0 to 330.0ºC, Resolution: 0.1ºC, Accuracy: ±0.6ºC
(J Range H)
Measuring range: 0 to 750.0ºC, Resolution: 0.1ºC, Accuracy: ±2.0ºC
(R)
Measuring range: 0 to 1600.0ºC, Resolution: 0.2ºC, Accuracy: ±2.2ºC
* Accuracy of the Internal Reference-junction Compensator, when temperature balanced at input terminals, and the ambient temperature is 25 ±10 ºC.
Type K, T, E, J, and R: Within ±0.5 ºC
Type N: Within ±1.0 ºC
Note: Accuracy does not include internal standard connection accuracy. Switching between internal and external standard connect compensators is possible. Thermocouple resistance 300 Ω or less (K type).
Standard Accessories Input section protection/terminal cover for protection from dust UT-50A

USV-51B
Voltage Unit

USV-51B

Specifications

Channels 10
Measuring Targetss DC voltage, voltage output type sensor
Measuring Range, Resolution, Accuracy (DC Input) Measuring range: 0 to ±20.000 V, Resolution: 1 mV, Accuracy: ±0.05% FS, Signal Source Resistance: 50 Ω or less
Optional Input section protection/terminal cover for protection from dust UT-50A

USM-51B
Strain/Voltage/Thermocouple Unit

USM-51B
  • TEDS

Specifications

Input NDIS connector+Screw soldering terminal block
Channels 10
Connection Targets Quarter bridge: 120 /350 Ω (2-wire/3-wire changed using internal circuit)
Half bridge/Full bridge: 120 to 1000 Ω
Strain-gage transducer: 120 to 1000 Ω
Potentiometer Sensor: 1 k to 10 k Ω
Voltage: ±20 V
Temperature (Thermocouple): K, T, E, J, R, N (Resistance of thermocouple: 1 kΩ or less)
Sampling Formats Simultaneous sampling of all channels (Temperature measurement: scanning, simultaneously update it approx. 1 s)
Bridge Excitation 2 VDC constant voltage
Power Supply to Potentiometers 2 VDC constant voltage
Gage Factor 2.00 fixed (Correction is possible at 2.00/ Ks with the engineering value conversion function)
Frequency Response DC to 7.8 Hz (Deviation: 0.5 dB to -3.5 dB, other than temperature measurement)
Check Functions When you select, burnout check runs in this moment.
TEDS Read from TEDS sensor
Dimensions 60 W x 120 H x 290 D mm (Excluding protrusions)
Weight Approx. 850 g
Measuring Range, Resolution, Accuracy (DC Input) (Strain range L)
Measuring range: 0 to ±19000 μm/ m, Resolution: 1 μm/m, Accuracy: Within ±0.08%FS
(Strain range H)
Measuring range: 0 to ±300000 μm/ m, Resolution: 10 μm/m, Accuracy: Within ±0.08%FS
(Potentiometer sensor)
Measuring range: 0 to ±50.00%, Resolution: 0 to ±50.00%, Accuracy: Within ±0.1%FS
(Voltage)
Measuring range: 0 to ±20.000 V, Resolution: 1 mV, Accuracy: Within ±0.08%FS
(Thermocouple K)
Measuring range: -200.0 to +1200.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +1200.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple T)
Measuring range: -200.0 to +350.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +350.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple E)
Measuring range: -200.0 to +800.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +800.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple J)
Measuring range: -200.0 to +750.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +750.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple R)
Measuring range: 0.0 to +1600.0ºC,
Resolution: 0.1ºC,
Accuracy: 0.0 to +100.0ºC or less (±(0.6% of reading+1.2ºC)), +100.0 to +1600.0ºC (±(0.5% of reading+1.0ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple N)
Measuring range: -200.0 to +1250.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +1250.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)

USM-52B
Strain/Voltage/Thermocouple Unit

USM-52B
  • TEDS

Specifications

Input NDIS connector+One-touch terminal block
Channels 10
Connection Targets Quarter bridge: 120 /350 Ω (2-wire/3-wire changed using internal circuit)
Half bridge system/Full bridge: 120 to 1000 Ω
Strain-gage transducer: 120 to 1000 Ω
Potentiometer Sensor: 1 k to 10 k Ω
Voltage: ±20 V
Temperature (Thermocouple): K, T, E, J, R, N (Resistance of thermocouple: 1 kΩ or less)
Sampling Formats Simultaneous sampling of all channels (Temperature measurement: scanning, simultaneously update it approx. 1 s)
Bridge Excitation 2 VDC constant voltage
Power Supply to Potentiometers 2 VDC constant voltage
Gage Factor 2.00 fixed (Correction is possible at 2.00/ Ks with the engineering value conversion function)
Frequency Response DC to 7.8 Hz (Deviation: 0.5 dB to -3.5 dB, other than temperature measurement)
Check Functions When you select, burnout check runs in this moment.
TEDS Read from TEDS sensor
Dimensions 60 W x 120 H x 290 D mm (Excluding protrusions)
Weight Approx. 750 g
Measuring Range, Resolution, Accuracy (DC Input) (Strain range L)
Measuring range: 0 to ±19000 μm/ m, Resolution: 1 μm/m, Accuracy: Within ±0.08%FS
(Strain range H)
Measuring range: 0 to ±300000 μm/ m, Resolution: 10 μm/m, Accuracy: Within ±0.08%FS
(Potentiometer sensor)
Measuring range: 0 to ±50.00%, Resolution: 0 to ±50.00%, Accuracy: Within ±0.1%FS
(Voltage)
Measuring range: 0 to ±20.000 V, Resolution: 1 mV, Accuracy: Within ±0.08%FS
(Thermocouple K)
Measuring range: -200.0 to +1200.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +1200.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple T)
Measuring range: -200.0 to +350.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +350.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple E)
Measuring range: -200.0 to +800.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +800.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple J)
Measuring range: -200.0 to +750.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +750.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple R)
Measuring range: 0.0 to +1600.0ºC,
Resolution: 0.1ºC,
Accuracy: 0.0 to +100.0ºC or less (±(0.6% of reading+1.2ºC)), +100.0 to +1600.0ºC (±(0.5% of reading+1.0ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)
(Thermocouple N)
Measuring range: -200.0 to +1250.0ºC,
Resolution: 0.1ºC,
Accuracy: -200.0 to -100.0ºC or less (±(0.3% of reading+0.8ºC)), -100.0 to +1250.0ºC (±(0.2% of reading+0.6ºC)),
Accuracy of internal reference contact compensator (At input terminal temperature balance): Within ±1.0ºC (Ambient temperature range: 25±10ºC), Within ±2.0ºC (Ambient temperature range: other than temperature ranges 25±10ºC)

Page Top

Notes

DCS-100A of a standard accessory can be measured up to 300 channels. Measurement up to 1000 channels requires DCS-106A of an option.