CAB-E
Strain Generator

Summary

· Generated strain level is determined by setting of 2 dials.
· Remote sensing possible
· No power supply is required

Features

The CAB-E is a compact and lightweight device, which generates equivalent strains to check strain measuring instruments.

Products table

All display

CAB-120E
120 Ω

Specifications

Equivalent Strain RANGE dials: 4 steps of ×-500, ×-100, ×100 and ×500
STRAIN dials: 11 steps of 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, and 10 ×10-6 strain

Generated strain level is determined by setting of both dials.
Accuracy Within ±(1.5% of setting +5 ×10-6 strain)
Gage Factor 2.0 fixed
I/O Resistance & Accuracy 120 Ω, -10% to 1%
Excitation Voltage 4 VDC or less
Operating Temperature 0 to 45°C
Operating Humidity 20 to 80% (Non-condensing)
Output Connector NDIS4102 (7 pins) connector
Dimensions 122 W × 70 H ×52 D mm
Weight Approx. 350 g
Standard Accessories Connection cable
 (With NDIS4102 (7 pins) Connector at both ends, 1 m)

Dimensions

CAB-350E
350 Ω

CAB-E

Specifications

Equivalent Strain RANGE dials: 4 steps of ×-500, ×-100, ×100 and ×500
STRAIN dials: 11 steps of 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, and 10 ×10-6 strain

Generated strain level is determined by setting of both dials.
Accuracy Within ±(1.5% of setting +5 ×10-6 strain)
Gage Factor 2.0 fixed
I/O Resistance & Accuracy 350 Ω, -10% to 1%
Excitation Voltage 12 VDC or less
Operating Temperature 0 to 45ºC
Operating Humidity 20 to 80% (Non-condensing)
Output Connector NDIS4102 (7 pins) connector
Dimensions 122 W × 70 H ×52 D mm
Weight Approx. 350 g
Standard Accessories Connection cable
 (With NDIS4102 (7 pins) Connector at both ends, 1 m)

Dimensions

Page Top

Remarks

1. Since the CAB-E is designed to be compatible with remote sensing, it mustn't be used for systems such as MCF, CDV cards, DIS, etc. with which F and G terminals of input NDIS4102 (7 pins) connector are used for other purposes.
2. It is not recommended to use for carrier-type strain amplifiers such as DPM series.
3. Since the CAB-E has a special circuit structure, the stated accuracy may not be satisfied depending on measuring instruments under test.
4. The CAB-E is designed for checking and is not for calibration.