CAB-E
Strain Generator

Summary
· Generated strain level is determined by setting of 2 dials.
· Remote sensing possible
· No power supply is required
Features
The CAB-E is a compact and lightweight device, which generates equivalent strains to check strain measuring instruments.
Products table
All display
CAB-120E
120 Ω

Specifications
Equivalent Strain | RANGE dials: 4 steps of ×-500, ×-100, ×100 and ×500 STRAIN dials: 11 steps of 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, and 10 ×10-6 strain Generated strain level is determined by setting of both dials. |
---|---|
Accuracy | Within ±(1.5% of setting +5 ×10-6 strain) |
Gage Factor | 2.0 fixed |
I/O Resistance & Accuracy | 120 Ω, -10% to 1% |
Excitation Voltage | 4 VDC or less |
Operating Temperature | 0 to 45°C |
Operating Humidity | 20 to 80% (Non-condensing) |
Output Connector | NDIS4102 (7 pins) connector |
Dimensions | 122 W × 70 H ×52 D mm |
Weight | Approx. 350 g |
Standard Accessories | Connection cable (With NDIS4102 (7 pins) Connector at both ends, 1 m) |
---|
Dimensions

CAB-350E
350 Ω

Specifications
Equivalent Strain | RANGE dials: 4 steps of ×-500, ×-100, ×100 and ×500 STRAIN dials: 11 steps of 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, and 10 ×10-6 strain Generated strain level is determined by setting of both dials. |
---|---|
Accuracy | Within ±(1.5% of setting +5 ×10-6 strain) |
Gage Factor | 2.0 fixed |
I/O Resistance & Accuracy | 350 Ω, -10% to 1% |
Excitation Voltage | 12 VDC or less |
Operating Temperature | 0 to 45ºC |
Operating Humidity | 20 to 80% (Non-condensing) |
Output Connector | NDIS4102 (7 pins) connector |
Dimensions | 122 W × 70 H ×52 D mm |
Weight | Approx. 350 g |
Standard Accessories | Connection cable (With NDIS4102 (7 pins) Connector at both ends, 1 m) |
---|
Dimensions

Remarks
1. Since the CAB-E is designed to be compatible with remote sensing, it mustn't be used for systems such as MCF, CDV cards, DIS, etc. with which F and G terminals of input NDIS4102 (7 pins) connector are used for other purposes.
2. It is not recommended to use for carrier-type strain amplifiers such as DPM series.
3. Since the CAB-E has a special circuit structure, the stated accuracy may not be satisfied depending on measuring instruments under test.
4. The CAB-E is designed for checking and is not for calibration.