CAB-E
Strain Generator

Summary

· Generated strain level is determined by setting of 2 dials.
· Remote sensing possible
· No power supply is required

Features

The CAB-E is a compact and lightweight device, which generates equivalent strains to check strain measuring instruments.

Products table

All display

CAB-120E
120 Ω

Specifications

Equivalent Strain RANGE dial: 4 steps of x-500, x-100, x100 and x500
STRAIN dial: 11 steps of 0, 1, 2, 3, 4, 5, 6, 7,8, 9, and 10 ×10-6 strain
Generated strain level is determined by setting of both dials.
Accuracy Within (±1.5% of setting + 5 ×10-6 strain)
Gage Factor 2.0 fixed
I/O Resistance & Accuracy 120 Ω, + 1, -10%
Excitation Voltage 4 VDC or less
Operating Temperature 0 to 45°C
Operating Humidity 20 to 80% (Non-condensing)
Output Connectors NDIS connector
Dimensions 122 W × 70 H ×52 D mm
Weight Approx. 350 g
Standard Accessories Connection cable with NDIS connectors at both ends (1 m).

Dimensions

CAB-350E
350 Ω

CAB-E

Specifications

Equivalent Strain RANGE dial: 4 steps of x-500, x-100, x100 and x500
STRAIN dial: 11 steps of 0, 1, 2, 3, 4, 5, 6, 7,8, 9, and 10 ×10-6 strain
Generated strain level is determined by setting of both dials.
Accuracy Within (±1.5% of setting + 5 ×10-6 strain)
Gage Factor 2.0 fixed
I/O Resistance & Accuracy 350 Ω, + 1, -10%
Excitation Voltage 12 VDC or less
Operating Temperature 0 to 45ºC
Operating Humidity 20 to 80% (Non-condensing)
Output Connectors NDIS connector
Dimensions 122 W × 70 H ×52 D mm
Weight Approx. 350 g
Standard Accessories Connection cable with NDIS connectors at both ends (1 m)

Dimensions

Page Top

To Ensure Safe Usage

1. Since the CAB-E is designed to be compatible with remote sensing, it cannot be used for the systems such as CDV cards, DIS, MCF etc. with which F and G terminals of input NDIS connector are used for other purposes.
2. It is not recommended to use for carrier-type strain amplifiers such as DPM series.
3. Since the CAB-E has a special circuit structure, the stated accuracy may not be satisfied depending on measuring instruments under test.
4. The CAB-E is designed for checking and cannot be used for calibration.

Download

Catalogs