EDX-5000A
Memory Recorder/Analyzer

  • CE
  • TEDS

EDX-5000A

Summary

· High-speed sampling at 200 kHz/32 channels
· Up to 80 input channels
· Simultaneous recording of measuring data and video
· Conditioner cards selectable for specific applications
· Effective real-time processing capability

Features

EDX-5000A is an advanced all-in-one measuring instrument having sophisticated features and high-speed processing capabilities.
It is the high-end model of EDX Series.

Products table

All display

EDX-5000A-64
8 slots, 64 channels

EDX-5000A
  • TEDS

Specifications

Model EDX-5000A-64-HE: Hard disk drive (HDD) , English
EDX-5000A-64-SE: Solid state drive (SSD) , English
Compatible Conditioner Cards CDV-40B (-F)
CVM-41A
DPM-42B (-F)
DPM-42B-I (-F)
CTA-40A
CFV-40A
CCA-40A (-F)
CDV-44AS
CDA-44AS
CDA-45AS
AD-40AS (-F)

*The number of CFV-40A that able to be mounted is up to 2 units.
Channels Max. 64 (CDV-40B × 8)
Analog Input See spec of respective conditioner cards.
Digital I/O
I/O Points Max. 8
I/O Settings Sets every bit to digital input, digital output, or remote-control input with the same common.

*Remote-control input:
 Start and stop measuring
 Execute BAL, etc.
Input Modes Isolated
TTL level input
Input Voltage Max. 5 VDC
Isolation Modes Digital isolation
Output Modes Isolated open-collector output
(With 10 kΩ built-in pull-up resistors)
Output Voltage 5 VDC
Output Current Max. 25 mA (per point)
Voice Input 1 channel (To be recorded together with measuring data.)
An optional remote control unit RCU-42 is required.
Sampling Frequencies
Sampling System Simultaneous sampling of all channels
Sampling Frequencies 1-2-5 series
1 Hz to 200 kHz: When recording data up to 32 channels (16 channels)
1 Hz to 100 kHz: When recording data up to 64 channels (32 channels)
1 Hz to 50 kHz: When recording data up to 80 channels (64 channels)
1 Hz to 20 kHz: When recording data up to 80 channels (80 channels)
1 Hz to 10 kHz: When real-time simultaneous process is enabled
2n series
2 Hz to 131.072 kHz: When recording data up to 32 channels (16 channels)
2 Hz to 65.536 kHz: When recording data up to 64 channels (32 channels)
2 Hz to 32.768 kHz: When recording data up to 80 channels (64 channels)
2 Hz to 16.384 kHz: When recording data up to 80 channels (80 channels)
2 Hz to 8.192 kHz: When real-time simultaneous process is enabled
[Note] AD data format: In 16-bit mode
[Note] In the 24-bit mode, the number of channels mentioned in parentheses.
Data Storage Device Hard Disk Drive (HDD) Capacity: 300 GB or more
Solid State Drive (SSD) Capacity: 30 GB or more
Display Channel status LED (OVER value of every channel is settable)
REC, BUSY, BATTERY, POWER LED
With 12.1-inch wide touch panel mounted display
Operation Keys Touch panel and front mounted keys(REC, STOP, BAL., OPT).
Operation with external keyboard and mouse is also possible.
External Control Connectors CONT IN, CONT OUT (remote control,synchronous operation)
External I/O Connectors External triggers: TRG IN, TRG OUT
External clock: CLK IN, CLK OUT
(Output at any frequency division ratio)
External Interfaces USB type keyboard
USB type mouse
VGA connector for monitor

USB I/F: 3 USB 2.0 in front, 2 USB 3.0 in rear
*The USB 2.0 port in the front is limited to about 100 mA that can be supplied.

LAN I/F: 100/1000BASE-T
Power Supply 100 to 240 VAC and 10 to 30 VDC, dual supply
Built-in batteries against momentary power failure
Current Consumption Approx. 1.8 A
(For 100 VAC, CDV-40B x 8)
Operating Temperature 0 to 40 ºC
Operating Humidity 20 to 80% (Non-condensing)
Storage Temperature -20 to 60 ºC
Vibration Resistant [EDX-5000A-64-HE, EDX-5000A-64-SE]
49.0 m/s2 (5 G), 5 to 55 Hz (When not operating)
29.4 m/s2 (3 G), 5 to 55 Hz (When operating)

[EDX-5000A-64-HE]
9.8 m/s2 (1 G), 10 to 200 Hz (When operating)

[EDX-5000A-64-SE]
19.6 m/s2 (2 G), 10 to 200 Hz (When operating)
Impact Resistant 196.1 m/s2 (20 G), 11 ms
Dimensions 365 W × 159 H × 300 D mm (Excluding protrusions)
Weight Approx. 11.5 kg (Mainframe only)
Synchronous Operation Records data synchronously by connecting multiple EDX-5000A units.
Able to connect EDX-5000A and EDX-3000B, EDX-5000A and EDX-200A.

Connectable units: 10
(When connecting EDX-5000A-80 units only: Max. 800 channels)

When connecting the EDX-200A, the connectable units is max. 8.
Optional Functions CAN data acquisition (Input channels: Max. 512)
GPS data acquisition (Tracing data acquisition, time synchronous function)
Online Control Controlled by DCS-100A
Compliance Compliance Directive 2014/30/EU (EMC)
Directive 2014/35/EU (LVD) (AC model only)
Directive 2011/65/EU, (EU)2015/863
(10 restricted substances) (RoHS)
Standard Accessories AC power cable P-18 (With 2-pin conversion plug CM-52)
DC power cable P-70
Ground wire P-72
Simplified instruction manual
CD-R (Instruction manual)
EDX accessory bag
AC fuse
DC fuse
Soldering plug for DIO
Shell case for DIO plug
Optional Accessories Synchronous cable N-128
Remote control unit RCU-42A
Belt hook
Dummy panel for EDX EDX1P-DUMMY
EDX2000-DUMMY

Software ETA-55A Specifications

Measuring Condition Settings
Measuring Channel Conditions Measurement ON/OFF
Measuring modes
Range
HPF
LPF
Digital HPF
Digital LPF
Balance ON/OFF
CAL
Calibration coefficient
Offset
Offset zero ON/OFF
Unit
Channel name
Measuring range
Rated capacity
Rated output
Upper-limit check value
Lower-limit check value
Numeric display digits (Selects any items to display)
Inner sensitivity register ON/OFF
Input cable
Cable compensation value
Recordable Data Up to remaining disk space of built-in memory
Manual Measurement Recording from REC to STOP, or preset data starts from REC
Interval Measurement Automatic recording according to designated starting time and recording interval
Trigger Measurement Recording and stopping according to preset trigger conditions

• Common trigger conditions
End trigger: Settable
Pre-trigger: Max. 2097152 points per channel
Post-trigger: Max. 2097152 points per channel
 (When AD is set to 24 bits)
Trigger data correlates with the number of channels and the AD data.

• Analog trigger conditions
Trigger channels: Any 1 analog channel
Trigger level: Sets in physical quantity.
Trigger slope: Up, down

• Digital trigger conditions
Trigger bit: Any 1 bit
Trigger level: 0, 1

• External trigger conditions
Trigger slope: Up, down

• Complex trigger conditions
Trigger source:
Selection of any 4 analog/digital channels, an external trigger channel, or a manual trigger channel
AND/OR:
Be used for analog trigger, digital trigger and external trigger.
Trigger level:
An physical quantity is set for the analog channel, and 0 or 1 for the digital channel.
Trigger slope: Up, down
TEDS Reads sensor's information and sets to channel condition automatically.
Load/Save Measurement Conditions File Load ME5 file and configure settings according to loaded measurement conditions.
Save the measurement conditions in ME5 file.

*You can also load/save ME5 file using DCS-100A and EDX-3000B COMPATIBLE SOFTWARE.
Measuring Operations
  Monitor measurement, recording start, pause, stop, balancing, CAL output, etc.
Real-time Processing Simultaneous monitoring and recording of data

The sampling frequencies up to 10 kHz are available.
Video Data Acquisition with a Web Camera
Camera DirectX compatible Web camera
(Recognized by the OS as an imaging device)
Number of Cameras 1
Resolution Max. 640 x 480
Frame Rate Max. 30 fps
Saving File Format AVI format

*Resolution and frame rate depend on the camera.
*The Web camera is optional.
Measuring Conditions During Recording Video Manual mode
Manual mode (Data points preset)
Arithmetic Operation
Digital Filters Butterworth filters (IIR)
Type of filters: LPF, HPF
Order of filter: 1 to 8
Amplitude ratio at cut off point: -3 dB
Attenuation: -6 x N dB/oct. (N is the filtering orders)

*Simultaneous use with built-in LPF of conditioner is possible.
Applicable Operators and Constants Calculation channels: Max. 64

Calculation conditions:
 ON/OFF
 Operators (Max. 200 characters)
 Unit
 Numeric display digits
 Channel name (Max. 40 characters)
 Calculation zero
 Calculation zero value

Applicable operators and constants:
+, -, *, /, ^ [power], PI [π], ()

Functions:
 SQR Square root
 ABS Absolute value
 SIN Sine
 COS Cosine
 TAN Tangent
 ASIN Arc sine (Return value: Radian)
 ACOS Arc cosine (Return value: Radian)
 ATAN Arc tangent (Return value: Radian)
 DSIN (Return value: Angle)
 DCOS (Return value: Angle)
 DTAN (Return value: Angle)
 LOG Common logarithm
 LN Natural logarithm
 EXP Exponent
 HMX Max. principal strain
 HMN Min. principal strain
 HSM Max. shearing strain
 SMX Max. principal stress
 SMN Min. principal stress
 SSM Max. shearing stress
 DEG Principal strain direction
FFT Analysis
Analysis Types Linear spectrum
Power spectrum
Cross spectrum
Auto-correlation
Cross-correlation
Window Functions OFF
Hamming
Hanning
Fejer
Blackman
Gaussian
Analyzed Data Points 256, 512, 1024, 2048, 4096, and 8192
Analysis Channels 4 channels/window
Saving File Format Kyowa standard file format (KS2)
KS2 file version: 01.06
Monitor
Y-time Graph X axis indicates the time, and Y axis the physical values of measurement for a maximum of 8 channels.
X-Y Graph X and Y axes indicate the measured data, max. 4 channels
Bar Graph Up to 80 channels are contained in a graph.
Displays max. and min. values are possible.
Circle Meter Any one channel is displayed in a circle meter.
Numeric Display Up to 80 channels are listed.
Displays max. and min. values of each channel are possible.
FFT Graph Up to 4 channels of analyzed results are displayed.
Web Camera Displays captured images.
Over Input Indication Displays the over-input values, background, etc. in red.
Graph Scale Capable of displaying auto-scale value on the Y-time graph (Y axis), X-Y graph (X, Y axes) and bar graph (Y axis).
Cursors Available on Y-time graph, X-Y graph, physical values appear at the cursor positions
Windows Up to 12 different graph windows are available at the same time.
Saving Images Saves each graph as an image file.
Environmental Settings
Synchronous Operation Standalone, sync host, or sync guest is selected
Others
  Inner, external oscillators switching, operating beep, balance reference value, AD bits switching, and optional switches selection
GPS/Multi-channel CAN Optional Functions
CAN Data Recording Up to 512 channels of CAN data can be recorded for each unit.
(CAN data is saved in E4A file format.)

*Use the ELA-55A for CAN condition settings.
CAN Trigger Start/stop recording data by using analog triggers, digital triggers, and complex triggers.

*Use the ELA-55A for CAN condition settings.
Graph Display Display numerical and analog data in graphs simultaneously.
Point ZERO (Manual) Measurement Able to start manual measurement at zero second (at the moment when ms is zero) using the time data received from GPS satellites.
Interval (GPS Synchronization) Measurement Able to start interval measurement at zero second (at the moment when ms is zero) using the time data received from GPS satellites.
GPS Data Recording Able to monitor and record GPS data (latitude, longitude, speed, bearing, etc.).
GPS data is saved in NMEA file format.

Software ELA-55A Specifications

Measuring Condition Settings
Measuring Channel Conditions Measurement ON/OFF
Measuring modes
Range
HPF
LPF
Digital HPF
Digital LPF
Balance ON/OFF
CAL
Calibration coefficient
Offset
Offset zero ON/OFF
Unit
Channel name
Measuring range
Rated capacity
Rated output
Upper-limit check value
Lower-limit check value
Numeric display digits (Selects any items to display)
Inner sensitivity register ON/OFF
Input cable
Cable compensation value
Recordable Data Up to remaining disk space of built-in memory
Manual Measurement Recording from REC to STOP, or preset data starts from REC
Interval Measurement Automatic recording according to designated starting time and recording interval
Trigger Measurement • Common trigger conditions
End trigger: Settable
Pre-trigger or post-trigger:
 Max. 2097152 points per channel when AD is set to 24 bits.
Trigger data correlates with the number of channels and the AD data.

• Analog trigger conditions
Trigger channels: Any 1 analog channel
Trigger level: Sets in physical quantity.
Trigger slope: Up, down

• Digital trigger conditions
Trigger bit: Any 1 bit
Trigger level: 0, 1

• External trigger conditions
Trigger slope: Up, down

• Complex trigger conditions
Trigger source:
Selection of any 4 analog/digital channels, an external trigger channel, or a manual trigger channel
AND/OR:
Be used for analog trigger, digital trigger and external trigger.
Trigger level:
An physical quantity is set for the analog channel, and 0 or 1 for the digital channel.
Trigger slope: Up, down
Reading/Saving Measuring Condition Files Capable of reading/saving measuring condition files
Reading/Saving Sensors Files Capable of reading/saving sensor's CSV files by channel condition
Digital I/O Settings I/O points: Max. 8

I/O settings:
 Sets every bit to digital input, digital output, or remote-control
TEDS Reads sensor's information and sets to channel condition automatically.
Load/Save ME5 File Load ME5 file and configure settings according to loaded measurement conditions.
Save the measurement conditions in ME5 file.

*You can also load/save ME5 file using DCS-100A and MAIN SOFTWARE.
Measuring Operations
  Monitor measurement, recording start, pause, stop, balancing, CAL output, etc.
Real-time Processing Simultaneous monitoring and recording of data

The sampling frequencies up to 10 kHz are available.
Video Data Acquisition with a Web Camera
Camera DirectX compatible Web camera
(Recognized by the OS as an imaging device)
Number of Cameras 1
Resolution Max. 640 x 480
Frame Rate Max. 30 fps
Saving File Format AVI format

*Resolution and frame rate depend on the camera.
*The Web camera is optional.
Measuring Conditions During Recording Video Manual mode
Manual mode (Data points preset)
Arithmetic Operation
Digital Filters Butterworth filters (IIR)
Type of filters: LPF, HPF
Order of filter: 1 to 8
Amplitude ratio at cut off point: -3 dB
Attenuation: -6 x N dB/oct. (N is the filtering orders)

*Simultaneous use with built-in LPF of conditioner is possible.
Applicable Operators and Constants Calculation channels: Max. 64

Calculation conditions:
 ON/OFF
 Operators (Max. 200 characters)
 Unit
 Numeric display digits
 Channel name (Max. 40 characters)
 Calculation zero
 Calculation zero value

Applicable operators and constants:
+, -, *, /, ^ [power], PI [π], ()

Functions:
 SQR Square root
 ABS Absolute value
 SIN Sine
 COS Cosine
 TAN Tangent
 ASIN Arc sine (Return value: Radian)
 ACOS Arc cosine (Return value: Radian)
 ATAN Arc tangent (Return value: Radian)
 DSIN (Return value: Angle)
 DCOS (Return value: Angle)
 DTAN (Return value: Angle)
 LOG Common logarithm
 LN Natural logarithm
 EXP Exponent
 HMX Max. principal strain
 HMN Min. principal strain
 HSM Max. shearing strain
 SMX Max. principal stress
 SMN Min. principal stress
 SSM Max. shearing stress
 DEG Principal strain direction
FFT Analysis
Analysis Types Linear spectrum
Power spectrum
Cross spectrum
Auto-correlation
Cross-correlation
Window Functions OFF
Hamming
Hanning
Fejer
Blackman
Gaussian
Analyzed Data Points 256, 512, 1024, 2048, 4096, and 8192
Analysis Channels 4 channels/window
Max. 8 channels
Saving Analyzed results are saved in the FFT analyzed result file format (CSV).
Saving File Format Kyowa standard file format (KS2)
KS2 file version: 01.06
Monitor
Y-time Graph X axis indicates the time, and Y axis the physical amount of measurement for a maximum of 16 channels.
1 to 10 graphs per window is possible.
Y-time (All channel) Graph X axis indicates the time, and Y axis the physical amount of measurement.
The line color is the same for all channels.
Y-time (DIV) Graph X axis indicates the time, and Y axis the physical amount of measurement for a maximum of 16 channels possible.
Allows channel's zero position to be set on the Y axis.
X-Y Graph Any combination of 8 channels is plotted on X and Y axis
Bar Graph Up to 32 channels are contained in a graph.
1 to 4 graphs are indicated on a window.
Peak hold ON/OFF
Digital Graph X axis indicates the time, and Y axis bit data of a digital channel (Up to 16 bits).
1 to 4 graphs are displayed on a window.
Circle Meter Any one channel is displayed in a circle meter.
Bar Meter Any one channel is displayed in a horizontal or vertical bar meter.
Numeric Display Any one channel, 16 channels or all channels are listed.

The max. and min. values of each channel are possible.
FFT Graph Up to 4 channels of analyzed results are displayed
Web Camera Displays captured images
Over Input Indication Capable of displaying the excessive channel values in red, except FFT graphs.
Graph Scale Capable of displaying auto-scale and full scale values on the Y-time graph (Y axis), X-Y graph (X, Y axes) and bar graph (Y axis).
Y axis of Y-time graph are switchable among 1-axis, 2-axis, and channel-by-channel
Display Color Any color is selectable per window.
Title, Labels Any title or X and Y axis labels are specified.
Windows 8 numeric windows
8 graph windows
(Including an FFT window)
Auxiliary Lines Capable of displaying the desired auxiliary lines on the Y-time Graphs (X axis and Y axis), X-Y Graphs (X axis and Y axis), and Bar Graphs (X axis and Y axis).
(Up to 4 auxiliary lines each for both X axis and Y axis.)
Comparative Data Displays the comparative data (existing KS2 format data) on the Y-time graphs, neither Y-time (All channels) graphs nor Y-time (DIV) graphs, and X-Y graphs for comparing the monitor data.

The size of the data file is maximum 10 MB.
If the file size exceeds 10 MB, displays the 10 MB data from its head.
Configuration
Operation Modes Standalone, sync host, sync guest setting
Control Modes Off-line, online switching
Automatic File Conversion Automatic file conversion upon measurement
(CSV, XLS, XLSX, or RPCIII format)
Others
  Inner, external oscillators switching, operating beep, balance reference value, AD bits switching, and optional switches selection
GPS/Multi-channel CAN Optional Functions
Recording CAN Data Capable of recording up to 512-channel CAN data per one unit.
(In the EDX-5000A data drive, as the E4A file).
CAN Trigger • Analog trigger conditions
Trigger channel: Any CAN channel
Trigger level: Set in CAN data
Trigger slope: Up, down

• Digital trigger conditions
Trigger channel: Any 1 bit of CAN channel
Trigger level: 0, 1

• Complex trigger conditions
Trigger source: Any 4 channels
AND/OR
 Analog trigger: Logic is possible.
 Digital trigger: Logic is possible.
Trigger level
 Analog channel: CAN data
 Digital channel: 0, 1
 Trigger slope: Up, down
Communications Conditions Communication standards
Communication speed
Terminal resistance

*Sample points, sampling frequency, resynchronization jump width are set automatically.
Measuring ID Conditions Port No
Format
Frame ID
Data length
Measuring Channel Conditions Start bit
Bit length
Data type
Endian
Calibration coefficient
Offset
(Conditions for conversion of extracting CAN data to physical quantity)
Graph Display Simultaneous display of graph
Numerical value
Analog data
Read CANdb Files File: DBC format
The number of “Network Node”: Maximum 1024
The number of “Message”: Maximum 2048
The number of “Signal”: Maximum 4096
Point ZERO (Manual Measurement) Starts the manual measurement at 0 ms by using the time data, received from the GPS satellite.
Interval (GPS Synchronization Measurement) Starts the interval measurement at 0 ms by using the time data, received from the GPS satellite.
Recording GPS Data Monitors/records the GPS data (latitude, longitude, speed, azimuth, etc.).
The ELA-55A saves the GPS data in the EDX-5000A data drive (in the NMEA format).

Data Reproduction, Data Analysis Specifications

Data Reproduction
Graph Display 4 patterns of display condition are set for a graph.
Allows 1, 2, or 4 graphs to be displayed
Y-time Graph Up to 16 channels per graph, and allows Y axis to be scaled automatically.
X-Y Graph 1 graph fixed, allows any 4 channels to be plotted on X- and Y-axis.
Both axes to be scaled automatically.
All Data Display Displays all data by every 4 channels.
Numeric Data Display Lists any 16 channels data
Cursor Numeric display of the engineering value of cursor position.
Zoom in data between 2 cursors.
Display max. and min. data between 2 cursors.
Scrolling Scrolls X axis on Y-time graphs.
Data File Editing Extracting of an arbitrary range or arbitrary channel from collected data file and conversion to a CSV format file are possible.
Data file titles, comments, channel conditions display and editing are possible.
Max. and Min. Display Showing max. and min. data of each channel (Max. 5-data) , 400-data around the max. or min. are shown in graphs.
KS2 File Block number support (1 block display, all block display).
Displays and plays back the audio data.
Video Playback Playback
Backward
Repeat
Frame-by-frame forward
Frame-by-frame backward
Jump to the beginning
Jump to the end
Set the beginning position
Set the end position

Playback speed: ×0.1 to ×20

Plays back the measured data, video and GPS data at the same time.
AVI Files Playback frame rate
Start frame No. (time)
Static Measuring Files Reproduces the static measuring files (CSV format files).
Data Analysis
Static Processing A list of maxima, minima, averages and standard deviations in a desired section of the data file.
The results are saved in CSV files.
Arithmetic Operations Inter-channel operations for up to 2 files.
The result is saved in a new file (Up to 320 expressions is possible).
Expression: Up to 200 characters

Applicable operators and constants:
+, -, ×, /, ^ [power], PI [π], ()

Functions:
 SQR Square root
 ABS Absolute value
 SIN Sine
 COS Cosine
 TAN Tangent
 ASIN Arc sine (Return value: Radian)
 ACOS Arc cosine (Return value: Radian)
 ATAN Arc tangent (Return value: Radian)
 DSIN (Return value: Angle)
 DCOS (Return value: Angle)
 DTAN (Return value: Angle)
 LOG Common logarithm
 LN Natural logarithm
 EXP Exponent
 HMX Max. principal strain
 HMN Min. principal strain
 HSM Max. shearing strain
 SMX Max. principal stress
 SMN Min. principal stress
 SSM Max. shearing stress
 DEG Principal strain direction
FFT Analysis
Analysis Types Linear spectrum
Power spectrum
Cross spectrum
Auto-correlation
Cross-correlation
Coherence
Transfer function
Analyzed Data Points 256, 512, 1024, 2048, 4096, 8192, 16384, and 32768
Window Functions OFF
Hamming
Hanning
Fejer
Blackman
Gaussian
LPF 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000 Hz and FLAT, 12 steps
Integration Times 0 (None) to 2
Average Times 0 or more (0: whole waveform)
Shift Data Points 2 or more
Analysis Types and Graphs [Linear spectrum]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: Phase

[Power spectrum]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: ---

[Cross-spectrum]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: Phase

[Auto-correlation]
Y-time graph: Yes
Analysis graph 1: Correlation
Analysis graph 2: ---

[Cross-correlation]
Y-time graph: Yes
Analysis graph 1: Correlation
Analysis graph 2: ---

[Coherence]
Y-time graph: Yes
Analysis graph 1: Coherence
Analysis graph 2: ---

[Transfer function]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: Phase

Analyzed results are saved in CSV files.
Histogram Analysis
Algorithm Types Peak/valley
Maxima/minima
1D rainflow
Amplitude
1D time at level
1D rainflow + peak/valley
1D rainflow + maxima/minima
2D rainflow
Slices 1D algorithm:
 Even numbers from 10 (±5) to 256 (±128)

2D algorithm:
 Even numbers from 10 to 50
 Slice width, hysteresis, offset (For maxima/minima), etc. to be specified.

Results:
 Tabular or drawing display (3D display for 2D rainflow)
Life Prediction Processing Predicts life from the result of histogram analysis of 1D rainflow method, 2D rainflow method, or amplitude method.
The life prediction data is shown and saved as a file.
(Allows S-N data files to be read.)
Filtering Digital IIR filters, 4th order Butterworth
 (Cutoff: -6dB, no phase delay)
HPF & LPF: FLAT to 500 k Hz
 (Effective up to one-half of the sampling frequency)
Mirroring

Analysis results are saved as an additional format
Differentiation/Integration Differential/integration times: 0 (None) to 2

Analysis results are saved as additional format
Utility
Multiple Files Conversion Conversion to CSV, XLS, XLSX, or RPCIII format
File Coupling Multiple files (Host and guests) acquired in synchronized operation are coupled into one file.
Reverse Conversion Data files converted by this software into CSV format are converted back to KS2 format.
File Division Extracts the specified block No. data of the data file into single file.
Batch Analysis Analysis of multiple files under same conditions at one time.
Histogram analysis, filtering and differentiation/integration are available.
Overwriting of Multiple Files Up to 16 data files are displayed and overlapped as Y-time data.
Allows start point of each file to be set.
Multiple Files Conversion Converts multiple KS2 files acquired in the interval measurement into a CSV file.

Dimensions

EDX-5000A

EDX-5000A-80
10 slots, 80 channels

EDX-5000A
  • CE
  • TEDS

Specifications

Model EDX-5000A-80-HE: Hard disk drive (HDD) , English
EDX-5000A-80-SE: Solid state drive (SSD) , English
Compatible Conditioner Cards CDV-40B (-F)
CVM-41A
DPM-42B (-F)
DPM-42B-I (-F)
CTA-40A
CFV-40A
CCA-40A (-F)
CDV-44AS
CDA-44AS
CDA-45AS
AD-40AS (-F)

*The number of CFV-40A that able to be mounted is up to 2 units.
Channels Max. 80 (CDV-40B × 10)
Analog Input See spec of respective conditioner cards.
Digital I/O
I/O Points Max. 8
I/O Settings Sets every bit to digital input, digital output, or remote-control input with the same common.

*Remote-control input:
 Start and stop measuring
 Execute BAL, etc.
Input Modes Isolated
TTL level input
Input Voltage Max. 5 VDC
Isolation Modes Digital isolation
Output Modes Isolated open-collector output
(With 10 kΩ built-in pull-up resistors)
Output Voltage 5 VDC
Output Current Max. 25 mA (per point)
Voice Input 1 channel (To be recorded together with measuring data.)
An optional remote control unit RCU-42 is required.
Sampling Frequencies
Sampling System Simultaneous sampling of all channels
Sampling Frequencies 1-2-5 series
1 Hz to 200 kHz: When recording data up to 32 channels (16 channels)
1 Hz to 100 kHz: When recording data up to 64 channels (32 channels)
1 Hz to 50 kHz: When recording data up to 80 channels (64 channels)
1 Hz to 20 kHz: When recording data up to 80 channels (80 channels)
1 Hz to 10 kHz: When real-time simultaneous process is enabled
2n series
2 Hz to 131.072 kHz: When recording data up to 32 channels (16 channels)
2 Hz to 65.536 kHz: When recording data up to 64 channels (32 channels)
2 Hz to 32.768 kHz: When recording data up to 80 channels (64 channels)
2 Hz to 16.384 kHz: When recording data up to 80 channels (80 channels)
2 Hz to 8.192 kHz: When real-time simultaneous process is enabled
[Note] AD data format: In 16-bit mode
[Note] In the 24-bit mode, the number of channels mentioned in parentheses.
Data Storage Device Hard Disk Drive (HDD) Capacity: 300 GB or more
Solid State Drive (SSD) Capacity: 30 GB or more
Display Channel status LED (OVER value of every channel is settable)
REC, BUSY, BATTERY, POWER LED
With 12.1-inch wide touch panel mounted display
Operation Keys Touch panel and front mounted keys(REC, STOP, BAL., OPT).
Operation with external keyboard and mouse is also possible.
External Control Connectors CONT IN, CONT OUT (remote control,synchronous operation)
External I/O Connectors External triggers: TRG IN, TRG OUT
External clock: CLK IN, CLK OUT
(Output at any frequency division ratio)
External Interfaces USB type keyboard
USB type mouse
VGA connector for monitor

USB I/F: 3 USB 2.0 in front, 2 USB 3.0 in rear
*The USB 2.0 port in the front is limited to about 100 mA that can be supplied.

LAN I/F: 100/1000BASE-T
Power Supply 100 to 240 VAC and 10 to 30 VDC, dual supply
Built-in batteries against momentary power failure
Current Consumption Approx. 2.0 A
(For 100 VAC, CDV-40B x 10)
Operating Temperature 0 to 40 ºC
Operating Humidity 20 to 80% (Non-condensing)
Storage Temperature -20 to 60 ºC
Vibration Resistant [EDX-5000A-80-HE, EDX-5000A-80-SE]
49.0 m/s2 (5 G), 5 to 55 Hz (When not operating)
29.4 m/s2 (3 G), 5 to 55 Hz (When operating)

[EDX-5000A-80-HE]
9.8 m/s2 (1 G), 10 to 200 Hz (When operating)

[EDX-5000A-80-SE]
19.6 m/s2 (2 G), 10 to 200 Hz (When operating)
Impact Resistant 196.1 m/s2 (20 G), 11 ms
Dimensions 410 W × 159 H × 300 D mm (Excluding protrusions)
Weight Approx. 12.5 kg (Mainframe only)
Synchronous Operation Records data synchronously by connecting multiple EDX-5000A units.
Able to connect EDX-5000A and EDX-3000B, EDX-5000A and EDX-200A.

Connectable units: 10
(When connecting EDX-5000A-80 units only: Max. 800 channels)

When connecting the EDX-200A, the connectable units is max. 8.
Optional Functions CAN data acquisition (Input channels: Max. 512)
GPS data acquisition (Tracing data acquisition, time synchronous function)
Online Control Controlled by DCS-100A
Compliance Compliance Directive 2014/30/EU (EMC)
Directive 2014/35/EU (LVD) (AC model only)
Directive 2011/65/EU, (EU)2015/863
(10 restricted substances) (RoHS)
Standard Accessories AC power cable P-18 (With 2-pin conversion plug CM-52)
DC power cable P-70
Ground wire P-72
Simplified instruction manual
CD-R (Instruction manual)
EDX accessory bag
AC fuse
DC fuse
Soldering plug for DIO
Shell case for DIO plug
Optional Accessories Synchronous cable N-128
Remote control unit RCU-42A
Belt hook
Dummy panel for EDX EDX1P-DUMMY
EDX2000-DUMMY

Software ETA-55A Specifications

Measuring Condition Settings
Measuring Channel Conditions Measurement ON/OFF
Measuring modes
Range
HPF
LPF
Digital HPF
Digital LPF
Balance ON/OFF
CAL
Calibration coefficient
Offset
Offset zero ON/OFF
Unit
Channel name
Measuring range
Rated capacity
Rated output
Upper-limit check value
Lower-limit check value
Numeric display digits (Selects any items to display)
Inner sensitivity register ON/OFF
Input cable
Cable compensation value
Recordable Data Up to remaining disk space of built-in memory
Manual Measurement Recording from REC to STOP, or preset data starts from REC
Interval Measurement Automatic recording according to designated starting time and recording interval
Trigger Measurement Recording and stopping according to preset trigger conditions

• Common trigger conditions
End trigger: Settable
Pre-trigger: Max. 2097152 points per channel
Post-trigger: Max. 2097152 points per channel
 (When AD is set to 24 bits)
Trigger data correlates with the number of channels and the AD data.

• Analog trigger conditions
Trigger channels: Any 1 analog channel
Trigger level: Sets in physical quantity.
Trigger slope: Up, down

• Digital trigger conditions
Trigger bit: Any 1 bit
Trigger level: 0, 1

• External trigger conditions
Trigger slope: Up, down

• Complex trigger conditions
Trigger source:
Selection of any 4 analog/digital channels, an external trigger channel, or a manual trigger channel
AND/OR:
Be used for analog trigger, digital trigger and external trigger.
Trigger level:
An physical quantity is set for the analog channel, and 0 or 1 for the digital channel.
Trigger slope: Up, down
TEDS Reads sensor's information and sets to channel condition automatically.
Load/Save Measurement Conditions File Load ME5 file and configure settings according to loaded measurement conditions.
Save the measurement conditions in ME5 file.

*You can also load/save ME5 file using DCS-100A and EDX-3000B COMPATIBLE SOFTWARE.
Measuring Operations
  Monitor measurement, recording start, pause, stop, balancing, CAL output, etc.
Real-time Processing Simultaneous monitoring and recording of data

The sampling frequencies up to 10 kHz are available.
Video Data Acquisition with a Web Camera
Camera DirectX compatible Web camera
(Recognized by the OS as an imaging device)
Number of Cameras 1
Resolution Max. 640 x 480
Frame Rate Max. 30 fps
Saving File Format AVI format

*Resolution and frame rate depend on the camera.
*The Web camera is optional.
Measuring Conditions During Recording Video Manual mode
Manual mode (Data points preset)
Arithmetic Operation
Digital Filters Butterworth filters (IIR)
Type of filters: LPF, HPF
Order of filter: 1 to 8
Amplitude ratio at cut off point: -3 dB
Attenuation: -6 x N dB/oct. (N is the filtering orders)

*Simultaneous use with built-in LPF of conditioner is possible.
Applicable Operators and Constants Calculation channels: Max. 64

Calculation conditions:
 ON/OFF
 Operators (Max. 200 characters)
 Unit
 Numeric display digits
 Channel name (Max. 40 characters)
 Calculation zero
 Calculation zero value

Applicable operators and constants:
+, -, *, /, ^ [power], PI [π], ()

Functions:
 SQR Square root
 ABS Absolute value
 SIN Sine
 COS Cosine
 TAN Tangent
 ASIN Arc sine (Return value: Radian)
 ACOS Arc cosine (Return value: Radian)
 ATAN Arc tangent (Return value: Radian)
 DSIN (Return value: Angle)
 DCOS (Return value: Angle)
 DTAN (Return value: Angle)
 LOG Common logarithm
 LN Natural logarithm
 EXP Exponent
 HMX Max. principal strain
 HMN Min. principal strain
 HSM Max. shearing strain
 SMX Max. principal stress
 SMN Min. principal stress
 SSM Max. shearing stress
 DEG Principal strain direction
FFT Analysis
Analysis Types Linear spectrum
Power spectrum
Cross spectrum
Auto-correlation
Cross-correlation
Window Functions OFF
Hamming
Hanning
Fejer
Blackman
Gaussian
Analyzed Data Points 256, 512, 1024, 2048, 4096, and 8192
Analysis Channels 4 channels/window
Saving File Format Kyowa standard file format (KS2)
KS2 file version: 01.06
Monitor
Y-time Graph X axis indicates the time, and Y axis the physical values of measurement for a maximum of 8 channels.
X-Y Graph X and Y axes indicate the measured data, max. 4 channels
Bar Graph Up to 80 channels are contained in a graph.
Displays max. and min. values are possible.
Circle Meter Any one channel is displayed in a circle meter.
Numeric Display Up to 80 channels are listed.
Displays max. and min. values of each channel are possible.
FFT Graph Up to 4 channels of analyzed results are displayed.
Web Camera Displays captured images.
Over Input Indication Displays the over-input values, background, etc. in red.
Graph Scale Capable of displaying auto-scale value on the Y-time graph (Y axis), X-Y graph (X, Y axes) and bar graph (Y axis).
Cursors Available on Y-time graph, X-Y graph, physical values appear at the cursor positions
Windows Up to 12 different graph windows are available at the same time.
Saving Images Saves each graph as an image file.
Environmental Settings
Synchronous Operation Standalone, sync host, or sync guest is selected
Others
  Inner, external oscillators switching, operating beep, balance reference value, AD bits switching, and optional switches selection
GPS/Multi-channel CAN Optional Functions
CAN Data Recording Up to 512 channels of CAN data can be recorded for each unit.
(CAN data is saved in E4A file format.)

*Use the ELA-55A for CAN condition settings.
CAN Trigger Start/stop recording data by using analog triggers, digital triggers, and complex triggers.

*Use the ELA-55A for CAN condition settings.
Graph Display Display numerical and analog data in graphs simultaneously.
Point ZERO (Manual) Measurement Able to start manual measurement at zero second (at the moment when ms is zero) using the time data received from GPS satellites.
Interval (GPS Synchronization) Measurement Able to start interval measurement at zero second (at the moment when ms is zero) using the time data received from GPS satellites.
GPS Data Recording Able to monitor and record GPS data (latitude, longitude, speed, bearing, etc.).
GPS data is saved in NMEA file format.

Software ELA-55A Specifications

Measuring Condition Settings
Measuring Channel Conditions Measurement ON/OFF
Measuring modes
Range
HPF
LPF
Digital HPF
Digital LPF
Balance ON/OFF
CAL
Calibration coefficient
Offset
Offset zero ON/OFF
Unit
Channel name
Measuring range
Rated capacity
Rated output
Upper-limit check value
Lower-limit check value
Numeric display digits (Selects any items to display)
Inner sensitivity register ON/OFF
Input cable
Cable compensation value
Recordable Data Up to remaining disk space of built-in memory
Manual Measurement Recording from REC to STOP, or preset data starts from REC
Interval Measurement Automatic recording according to designated starting time and recording interval
Trigger Measurement • Common trigger conditions
End trigger: Settable
Pre-trigger or post-trigger:
 Max. 2097152 points per channel when AD is set to 24 bits.
Trigger data correlates with the number of channels and the AD data.

• Analog trigger conditions
Trigger channels: Any 1 analog channel
Trigger level: Sets in physical quantity.
Trigger slope: Up, down

• Digital trigger conditions
Trigger bit: Any 1 bit
Trigger level: 0, 1

• External trigger conditions
Trigger slope: Up, down

• Complex trigger conditions
Trigger source:
Selection of any 4 analog/digital channels, an external trigger channel, or a manual trigger channel
AND/OR:
Be used for analog trigger, digital trigger and external trigger.
Trigger level:
An physical quantity is set for the analog channel, and 0 or 1 for the digital channel.
Trigger slope: Up, down
Reading/Saving Measuring Condition Files Capable of reading/saving measuring condition files
Reading/Saving Sensors Files Capable of reading/saving sensor's CSV files by channel condition
Digital I/O Settings I/O points: Max. 8

I/O settings:
 Sets every bit to digital input, digital output, or remote-control
TEDS Reads sensor's information and sets to channel condition automatically.
Load/Save ME5 File Load ME5 file and configure settings according to loaded measurement conditions.
Save the measurement conditions in ME5 file.

*You can also load/save ME5 file using DCS-100A and MAIN SOFTWARE.
Measuring Operations
  Monitor measurement, recording start, pause, stop, balancing, CAL output, etc.
Real-time Processing Simultaneous monitoring and recording of data

The sampling frequencies up to 10 kHz are available.
Video Data Acquisition with a Web Camera
Camera DirectX compatible Web camera
(Recognized by the OS as an imaging device)
Number of Cameras 1
Resolution Max. 640 x 480
Frame Rate Max. 30 fps
Saving File Format AVI format

*Resolution and frame rate depend on the camera.
*The Web camera is optional.
Measuring Conditions During Recording Video Manual mode
Manual mode (Data points preset)
Arithmetic Operation
Digital Filters Butterworth filters (IIR)
Type of filters: LPF, HPF
Order of filter: 1 to 8
Amplitude ratio at cut off point: -3 dB
Attenuation: -6 x N dB/oct. (N is the filtering orders)

*Simultaneous use with built-in LPF of conditioner is possible.
Applicable Operators and Constants Calculation channels: Max. 64

Calculation conditions:
 ON/OFF
 Operators (Max. 200 characters)
 Unit
 Numeric display digits
 Channel name (Max. 40 characters)
 Calculation zero
 Calculation zero value

Applicable operators and constants:
+, -, *, /, ^ [power], PI [π], ()

Functions:
 SQR Square root
 ABS Absolute value
 SIN Sine
 COS Cosine
 TAN Tangent
 ASIN Arc sine (Return value: Radian)
 ACOS Arc cosine (Return value: Radian)
 ATAN Arc tangent (Return value: Radian)
 DSIN (Return value: Angle)
 DCOS (Return value: Angle)
 DTAN (Return value: Angle)
 LOG Common logarithm
 LN Natural logarithm
 EXP Exponent
 HMX Max. principal strain
 HMN Min. principal strain
 HSM Max. shearing strain
 SMX Max. principal stress
 SMN Min. principal stress
 SSM Max. shearing stress
 DEG Principal strain direction
FFT Analysis
Analysis Types Linear spectrum
Power spectrum
Cross spectrum
Auto-correlation
Cross-correlation
Window Functions OFF
Hamming
Hanning
Fejer
Blackman
Gaussian
Analyzed Data Points 256, 512, 1024, 2048, 4096, and 8192
Analysis Channels 4 channels/window
Max. 8 windows
Saving Analyzed results are saved in the FFT analyzed result file format (CSV).
Saving File Format Kyowa standard file format (KS2)
KS2 file version: 01.06
Monitor
Y-time Graph X axis indicates the time, and Y axis the physical amount of measurement for a maximum of 16 channels.
1 to 10 graphs per window is possible.
Y-time (All channel) Graph X axis indicates the time, and Y axis the physical amount of measurement.
The line color is the same for all channels.
Y-time (DIV) Graph X axis indicates the time, and Y axis the physical amount of measurement for a maximum of 16 channels possible.
Allows channel's zero position to be set on the Y axis.
X-Y Graph Any combination of 8 channels is plotted on X and Y axis
Bar Graph Up to 32 channels are contained in a graph.
1 to 4 graphs are indicated on a window.
Peak hold ON/OFF
Digital Graph X axis indicates the time, and Y axis bit data of a digital channel (Up to 16 bits).
1 to 4 graphs are displayed on a window.
Circle Meter Any one channel is displayed in a circle meter.
Bar Meter Any one channel is displayed in a horizontal or vertical bar meter.
Numeric Display Any one channel, 16 channels or all channels are listed.

The max. and min. values of each channel are possible.
FFT Graph Up to 4 channels of analyzed results are displayed
Web Camera Displays captured images
Over Input Indication Capable of displaying the excessive channel values in red, except FFT graphs.
Graph Scale Capable of displaying auto-scale and full scale values on the Y-time graph (Y axis), X-Y graph (X, Y axes) and bar graph (Y axis).
Y axis of Y-time graph are switchable among 1-axis, 2-axis, and channel-by-channel
Display Color Any color is selectable per window.
Title, Labels Any title or X and Y axis labels are specified.
Windows 8 numeric windows
8 graph windows
(Including an FFT window)
Auxiliary Lines Capable of displaying the desired auxiliary lines on the Y-time Graphs (X axis and Y axis), X-Y Graphs (X axis and Y axis), and Bar Graphs (X axis and Y axis).
(Up to 4 auxiliary lines each for both X axis and Y axis.)
Comparative Data Displays the comparative data (existing KS2 format data) on the Y-time graphs, neither Y-time (All channels) graphs nor Y-time (DIV) graphs, and X-Y graphs for comparing the monitor data.

The size of the data file is maximum 10 MB.
If the file size exceeds 10 MB, displays the 10 MB data from its head.
Configuration
Operation Modes Standalone, sync host, sync guest setting
Control Modes Off-line, online switching
Automatic File Conversion Automatic file conversion upon measurement
(CSV, XLS, XLSX, or RPCIII format)
Others
  Inner, external oscillators switching, operating beep, balance reference value, AD bits switching, and optional switches selection
GPS/Multi-channel CAN Optional Functions
Recording CAN Data Capable of recording up to 512-channel CAN data per one unit.
(In the EDX-5000A data drive, as the E4A file).
CAN Trigger • Analog trigger conditions
Trigger channel: Any CAN channel
Trigger level: Set in CAN data
Trigger slope: Up, down

• Digital trigger conditions
Trigger channel: Any 1 bit of CAN channel
Trigger level: 0, 1

• Complex trigger conditions
Trigger source: Any 4 channels
AND/OR
 Analog trigger: Logic is possible.
 Digital trigger: Logic is possible.
Trigger level
 Analog channel: CAN data
 Digital channel: 0, 1
 Trigger slope: Up, down
Communications Conditions Communication standards
Communication speed
Terminal resistance

*Sample points, sampling frequency, resynchronization jump width are set automatically.
Measuring ID Conditions Port No
Format
Frame ID
Data length
Measuring Channel Conditions Start bit
Bit length
Data type
Endian
Calibration coefficient
Offset
(Conditions for conversion of extracting CAN data to physical quantity)
Graph Display Simultaneous display of graph
Numerical value
Analog data
Read CANdb Files File: DBC format
The number of “Network Node”: Maximum 1024
The number of “Message”: Maximum 2048
The number of “Signal”: Maximum 4096
Point ZERO (Manual Measurement) Starts the manual measurement at 0 ms by using the time data, received from the GPS satellite.
Interval (GPS Synchronization Measurement) Starts the interval measurement at 0 ms by using the time data, received from the GPS satellite.
Recording GPS Data Monitors/records the GPS data (latitude, longitude, speed, azimuth, etc.).
The ELA-55A saves the GPS data in the EDX-5000A data drive (in the NMEA format).

Data Reproduction, Data Analysis Specifications

Data Reproduction
Graph Display 4 patterns of display condition are set for a graph.
Allows 1, 2, or 4 graphs to be displayed
Y-time Graph Up to 16 channels per graph, and allows Y axis to be scaled automatically.
X-Y Graph 1 graph fixed, allows any 4 channels to be plotted on X- and Y-axis.
Both axes to be scaled automatically.
All Data Display Displays all data by every 4 channels.
Numeric Data Display Lists any 16 channels data
Cursor Numeric display of the engineering value of cursor position.
Zoom in data between 2 cursors.
Display max. and min. data between 2 cursors.
Scrolling Scrolls X axis on Y-time graphs.
Data File Editing Extracting of an arbitrary range or arbitrary channel from collected data file and conversion to a CSV format file are possible.
Data file titles, comments, channel conditions display and editing are possible.
Max. and Min. Display Showing max. and min. data of each channel (Max. 5-data) , 400-data around the max. or min. are shown in graphs.
KS2 File Block number support (1 block display, all block display).
Displays and plays back the audio data.
Video Playback Playback
Backward
Repeat
Frame-by-frame forward
Frame-by-frame backward
Jump to the beginning
Jump to the end
Set the beginning position
Set the end position

Playback speed: ×0.1 to ×20

Plays back the measured data, video and GPS data at the same time.
AVI Files Playback frame rate
Start frame No. (time)
Static Measuring Files Reproduces the static measuring files (CSV format files).
Data Analysis
Static Processing A list of maxima, minima, averages and standard deviations in a desired section of the data file.
The results are saved in CSV files.
Arithmetic Operations Inter-channel operations for up to 2 files.
The result is saved in a new file (Up to 320 expressions is possible).
Expression: Up to 200 characters

Applicable operators and constants:
+, -, ×, /, ^ [power], PI [π], ()

Functions:
 SQR Square root
 ABS Absolute value
 SIN Sine
 COS Cosine
 TAN Tangent
 ASIN Arc sine (Return value: Radian)
 ACOS Arc cosine (Return value: Radian)
 ATAN Arc tangent (Return value: Radian)
 DSIN (Return value: Angle)
 DCOS (Return value: Angle)
 DTAN (Return value: Angle)
 LOG Common logarithm
 LN Natural logarithm
 EXP Exponent
 HMX Max. principal strain
 HMN Min. principal strain
 HSM Max. shearing strain
 SMX Max. principal stress
 SMN Min. principal stress
 SSM Max. shearing stress
 DEG Principal strain direction
FFT Analysis
Analysis Types Linear spectrum
Power spectrum
Cross spectrum
Auto-correlation
Cross-correlation
Coherence
Transfer function
Analyzed Data Points 256, 512, 1024, 2048, 4096, 8192, 16384, and 32768
Window Functions OFF
Hamming
Hanning
Fejer
Blackman
Gaussian
LPF 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000 Hz and FLAT, 12 steps
Integration Times 0 (None) to 2
Average Times 0 or more (0: whole waveform)
Shift Data Points 2 or more
Analysis Types and Graphs [Linear spectrum]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: Phase

[Power spectrum]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: ---

[Cross-spectrum]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: Phase

[Auto-correlation]
Y-time graph: Yes
Analysis graph 1: Correlation
Analysis graph 2: ---

[Cross-correlation]
Y-time graph: Yes
Analysis graph 1: Correlation
Analysis graph 2: ---

[Coherence]
Y-time graph: Yes
Analysis graph 1: Coherence
Analysis graph 2: ---

[Transfer function]
Y-time graph: Yes
Analysis graph 1: Amplitude (linear)/amplitude (log)
Analysis graph 2: Phase

Analyzed results are saved in CSV files.
Histogram Analysis
Algorithm Types Peak/valley
Maxima/minima
1D rainflow
Amplitude
1D time at level
1D rainflow + peak/valley
1D rainflow + maxima/minima
2D rainflow
Slices 1D algorithm:
 Even numbers from 10 (±5) to 256 (±128)

2D algorithm:
 Even numbers from 10 to 50
 Slice width, hysteresis, offset (For maxima/minima), etc. to be specified.

Results:
 Tabular or drawing display (3D display for 2D rainflow)
Life Prediction Processing Predicts life from the result of histogram analysis of 1D rainflow method, 2D rainflow method, or amplitude method.
The life prediction data is shown and saved as a file.
(Allows S-N data files to be read.)
Filtering Digital IIR filters, 4th order Butterworth
 (Cutoff: -6dB, no phase delay)
HPF & LPF: FLAT to 500 k Hz
 (Effective up to one-half of the sampling frequency)
Mirroring

Analysis results are saved as an additional format
Differentiation/Integration Differential/integration times: 0 (None) to 2

Analysis results are saved as additional format
Utility
Multiple Files Conversion Conversion to CSV, XLS, XLSX, or RPCIII format
File Coupling Multiple files (Host and guests) acquired in synchronized operation are coupled into one file.
Reverse Conversion Data files converted by this software into CSV format are converted back to KS2 format.
File Division Extracts the specified block No. data of the data file into single file.
Batch Analysis Analysis of multiple files under same conditions at one time.
Histogram analysis, filtering and differentiation/integration are available.
Overwriting of Multiple Files Up to 16 data files are displayed and overlapped as Y-time data.
Allows start point of each file to be set.
Multiple Files Conversion Converts multiple KS2 files acquired in the interval measurement into a CSV file.

Dimensions

EDX-5000A

EGPC-50A
GPS/Multi-channel CAN Module

EGPC-50A
  • CE

Specifications

Applicable Model Mounted in the EDX-5000A
I/O Ports
GPS/CAN Shared Port 1
Dsub connector 9 pins (male)
CAN Only Port 1
Dsub connector 9 pins (male)
Input CAN Channels Max. 512
Applicable CAN Versions CAN2.0A/B
(Conforming to ISO-11898, ISO-11519-2)
Communication Speed [High Speed CAN]
1000, 800, 500, 250, 125, 100, 83.3, 62.5, 50,33.3, 25, 20, 10 kbps

[Low Speed CAN]
125, 100, 83.3, 62.5, 50, 33.3, 25, 20, 10 kbps
CAN Data Output
Output Start Cond Output an arbitrary CAN data when starting the AD conversion.
Output Stop Cond Output an arbitrary CAN data when the AD conversion is stopped.
Output Interval Cond Output an arbitrary CAN data at the preset constant frequency.
Output Manual Cond Output an arbitrary CAN data at arbitrary timing.
Measurable GPS Data Latitude
Longitude
Height
Course
Speed
Time
GPS status
Number of satellites, etc.

Saving format: NMEA format
Time Synchronization Function Synchronizes the recording start time of EDX-5000A units by using the time data received from GPS satellites.
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 80% (Non-condensing)
Storage Temperature -20 to 60ºC
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU)2015/863
(10 restricted substances) (RoHS)
Standard Accessories GPS sensor (Cable length: 5 m)

RCU-42A
Remote Control Unit

Object Product Model EDX-200A
EDX-5000A
Control Functions REC/PAUSE (Starts/pauses data acquisition)
STOP (Stops data acquisition)
BAL (Balancing)
OPT. (Optional function)
VOICE MEMO (Recording with the built-in microphone)
Indication Recording, pausing and balancing are indicated with LED.
Buzzer Equivalent to the EDX recorder unit buzzer
Cable Length Approx. 1.5 m
Dimensions 35 W× 125 H×22 D (mm) *Excluding protrusions
Weight Approx. 220 g