EDX-10 Series
Compact Recording System

Summary
· Compact & lightweight
· With stacked connection, no synchronization cable is required, therefore wiring-saving.
· Max. sampling frequency 20 k Hz for 4 channels of a single measuring unit in sync.
· Sensors are easily connected with one-touch input cables or input adapters.
Features
The EDX-10 Series is measuring instrument, which can measure strain, voltage, and temperature (Thermocouples) simply by connecting a PC via USB interface.
Up to 4 measuring units for 16 channels can be powered by USB interface, no separated power supply is required.
Products table
All display
EDX-10B
Control Unit

Specifications
Interfaces | USB2.0 compliant Connector configuration: USB standard B receptacle |
---|---|
Installed Measuring Units | Max. 4 (16 channels) |
Sampling Frequencies | 1 Hz to 20 k Hz (1 to 4 channels) 1 Hz to 10 k Hz (1 to 8 channels) 1 Hz to 5 k Hz (1 to 16 channels) |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 90% (Non-condensing) |
Power Supply | 5 VDC by USB bus power or an optional accessory AC adapter Notes: 1. When using USB bus-power, please connect directly to the USB port of the PC without using a USB hub. 2. When using AC adapter, the maximum connection units is 4 units in any combination of EDX-11A, 12A, 13A, 14A and 15A. |
Current Consumption | 140 mA or less (5 VDC) |
Weight | Approx. 170 g |
Dimensions | 84.0 W x 26.6 H x 84.0 D mm (Excluding protrusions) |
Control Software | DCS-100A |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | USB cable N-38 (1 m) Ground wire P-72 (5 m) DVD (Dynamic data acquisition software DCS-100A) |
---|---|
Optional Accessories | AC adapter UN310-0515 (For U.S.A.: UN312-0520) |
USB Ports | The number of measuring units at USB bus-power operation, refer to the following table. 【USB3.0】-------------------------------------------------------------------------------- EDX-11A units Number of EDX-12A/13A/14A/15A Connection units 0 Up to 4 units in any combination Max. 4 1 Until either 1 unit Max. 2 1 0 Max. 2 【USB2.0】-------------------------------------------------------------------------------- EDX-11A units Number of EDX-12A/13A/14A/15A Connection units 0 Up to 2 units in any combination Max. 2 1 0 Max. 1 |
Dimensions

EDX-14A
Strain Measuring Unit

Specifications
Measuring Targets | Strain-gage transducers, strain gages *Bridge boxes or input adapters are required for strain measurement. |
---|---|
Channels | 4 |
Measuring Range | 10 k, 50 k ×10-6 strain (2 steps) |
Compatible Bridge Resistance | 120 Ω to 1 kΩ |
Bridge Excitation | 1 VDC |
Gage Factor | 2.00 fixed |
Range Accuracy | Within ±0.3%FS |
Balance Adjustment | Within ±1/2 F.S. of setting range |
Nonlinearity | Within ±0.1%FS |
AD Converter | 24 bits |
Frequency Response | DC to 2 kHz |
LPF | Cutoff frequencies : 100 Hz, 2 k Hz Transfer characteristic : 2nd order Butterworth |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 90% (Non-condensing) |
Input Connectors | D-sub 37-pin connector |
Power Supply | 5 VDC supplied by control unit |
Current Consumption | 140 mA or less (120 Ω load with all channels connected, at power supply 5 VDC) |
Weight | Approx. 150 g |
Dimensions | 84.0 W × 26.6 H × 84.0 D mm (Excluding protrusions) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Strain input cable U-124 (30 cm) |
---|---|
Optional Accessories | Bridge box connection cable U-126 (50 cm) Input connector set EDX10-DSUB Input adapter UI-51A One-touch type input adapter UI-52A Bridge adapter for quarter bridge system UI-53B-120/350 Bridge adapter for quarter bridge system UI-54B-120/350 One-touch type input adapter UI-55A |
Dimensions

EDX-15A
Strain/Voltage Measuring Unit

Specifications
Measuring Targets | Strain-gage transducers, strain gages *Bridge boxes or input adapters are required for strain measurement. Voltage |
---|---|
Channels | 4 |
Measuring Range | Strain-gage transducers, strain gages: 10 k, 50 k ×10-6 strain (2 steps) Voltage: 10, 50 V |
Compatible Bridge Resistance | 120 Ω to 1 kΩ |
Bridge Excitation | 1 VDC |
Gage Factor | 2.00 fixed |
Range Accuracy | Within ±0.3%FS |
Balance Adjustment | Within ±1/2 F.S. of setting range |
Nonlinearity | Within ±0.1%FS |
AD Converter | 24 bits |
Frequency Response | DC to 2 kHz |
LPF | Cutoff frequencies : 100 Hz, 2 k Hz Transfer characteristic : 2nd order Butterworth |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 90% (Non-condensing) |
Input Connectors | D-sub 37-pin connector |
Power Supply | 5 VDC supplied by control unit |
Current Consumption | 150 mA or less (120 Ω load with all channels connected, at power supply 5 VDC) |
Weight | Approx. 150 g |
Dimensions | 84.0 W × 26.6 H × 84.0 D mm (Excluding protrusions) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Strain input cable U-124 (30 cm) Conversion adapter FV-1A x 4 |
---|---|
Optional Accessories | Bridge box connection cable U-126 (50 cm) Input connector set EDX10-DSUB Input adapter UI-51A One-touch type input adapter UI-52A Bridge adapter for quarter bridge system UI-53B-120/350 Bridge adapter for quarter bridge system UI-54B-120/350 One-touch type input adapter UI-55A BNC input cable U-125 (30 cm) |
Dimensions

EDX-12A
Voltage Measuring Unit

Specifications
Measuring Targets | Voltage |
---|---|
Channels | 4 (Single end) |
Measuring Range | 10 V, 50 V (2 steps) |
Range Accuracy | Each range within ±0.3%FS |
Balance Adjustment | Within ±1/2 FS of setting range |
Nonlinearity | Within ±0.1%FS |
AD Converter | 24 bits |
Frequency Response | DC to 2 kHz |
LPF | Cutoff frequencies : 100 Hz, 2 k Hz Transfer characteristic : 2nd order Butterworth |
Operating Temperature | 0 to 40ºC |
Input Connectors | D-sub 37-pin connector |
Power Supply | 5 VDC supplied by control unit |
Current Consumption | 110 mA or less (5 VDC) |
Weight | Approx. 150 g |
Dimensions | 84.0 W × 26.6 H × 84.0 D mm (Excluding protrusions) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Input adapter UI-51A |
---|---|
Optional Accessories | BNC input cable U-125 (30 cm) Bridge box connection cable U-126 (50 cm) Input connector set EDX10-DSUB One-touch type input adapter UI-52A |
Dimensions

EDX-13A
Thermocouple Measuring Unit

Specifications
Measuring Targets | Thermocouples |
---|---|
Channels | 4 |
Input Connectors | Screw type terminal box |
Applicable Thermocouples | K, T, J, N (Resistance of thermocouple: 1 kΩ or less) |
Range | K: -200.0 to 1370.0ºC T: -200.0 to 400.0ºC J: -200.0 to 1200.0ºC N: -200.0 to 1300.0ºC |
Resolution | 0.1°C |
Accuracy | [K] −200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC) −100.0 to 1370.0°C: ± (0.1% of reading + 0.4ºC) [T] −200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC) −100.0 to 400.0°C: ± (0.1% of reading + 0.4ºC) [J] −200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC) −100.0 to 1200.0°C: ± (0.1% of reading+0.4ºC) [N] −200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC) −100.0 to 1300.0°C: ± (0.1% of reading+0.4ºC) Note: The measurement accuracy does not include the accuracy of the internal reference junction compensator and thermocouples. |
Accuracy of internal reference junction compensator | ±1.0°C 〔Input terminal temperature at equilibrium〕 〔Ambient temperature: 25 ±10°C〕 Mount the EDX-13A on the bottom when using it with measuring units other than the EDX-13A. ±2.0°C 〔Input terminal temperature in equilibrium〕 For temperatures other than those in the ambient temperature and operating temperature described above |
Check Functions | Burnout check |
AD Converter | 24 bits |
Sampling System | Scanning |
Inside Samping Frequencies | Approx. 0.5 Hz, approx. 2.0 Hz |
Operating Temperature | 0 to 40℃ |
Operating Humidity | 20 to 90% (Non-condensing) |
Input Connector | Screw type terminal box |
Power Supply | 5 VDC supplied by control unit |
Current Consumption | 120 mA or less (5 VDC) |
Weight | Approx. 130 g |
Dimensions | 84.0 W × 26.6 H × 84.0 D mm (Excluding protrusions) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Terminal box 1 piece Screwdriver 1 piece |
---|
Dimensions

EDX-11A
Strain Measuring Unit

Specifications
Measuring Targets | Strain-gage transducers, strain gage *Bridge boxes or input adapters are required for strain measurement. |
---|---|
Channels | 4 |
Measuring Range | 10 k, 50 k ×10-6 strain (2 steps) |
Balance Adjustment | Within ±1/2 F.S. of setting range |
Compatible Bridge Resistance | 120 Ω to 1 kΩ |
Bridge Excitation | 2 VDC |
Gage Factor | 2.00 fixed |
Range Accuracy | Each range within ±0.3%FS |
Nonlinearity | Within ±0.1%FS |
AD Converter | 24 bits |
Frequency Response | DC to 2 kHz |
LPF | Cutoff frequencies : Lo (100 Hz), Hi (2 k Hz) Transfer characteristic : 2nd order Butterworth |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 90% (Non-condensing) |
Input Connectors | D-sub 37-pin connector |
Power Supply | 5 VDC supplied by control unit |
Current Consumption | 180 mA or less (120 Ω load with all channels connected, at power supply 5 VDC) |
Weight | Approx. 150 g |
Dimensions | 84.0 W × 26.6 H × 84.0 D mm (Excluding protrusions) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Strain input cable U-124 (30 cm) |
---|---|
Optional Accessories | Bridge box connection cable U-126 (50 cm) Input connector set EDX10-DSUB Input adapter UI-51A One-touch type input adapter UI-52A, UI-55A Bridge adapter for quarter bridge system UI-53B-120/350 Bridge adapter for quarter bridge system UI-54B-120/350 |
Dimensions