EDX-10 Series
Compact Recording System

  • CE

EDX-10series

Summary

· Compact & lightweight
· With stacked connection, no synchronization cable is required, therefore wiring-saving.
· Max. sampling frequency 20 k Hz for 4 channels of a single measuring unit in sync.
· Sensors are easily connected with one-touch input cables or input adapters.

Features

The EDX-10 Series is measuring instrument, which can measure strain, voltage, and temperature (Thermocouples) simply by connecting a PC via USB interface.
Up to 4 measuring units for 16 channels can be powered by USB interface, no separated power supply is required.

Products table

All display

EDX-10B
Control Unit

EDX-10B
  • CE

Specifications

Interfaces USB2.0 compliant
Connector configuration: USB standard B receptacle
Installed Measuring Units Max. 4 (16 channels)
Sampling Frequencies 1 Hz to 20 k Hz (1 to 4 channels)
1 Hz to 10 k Hz (1 to 8 channels)
1 Hz to 5 k Hz (1 to 16 channels)
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 90% (Non-condensing)
Power Supply 5 VDC by USB bus power or an optional accessory AC adapter

Notes:
1. When using USB bus-power, please connect directly to the USB port of the PC without using a USB hub.
2. When using AC adapter, the maximum connection units is 4 units in any
combination of EDX-11A, 12A, 13A, 14A and 15A.
Current Consumption 140 mA or less (5 VDC)
Weight Approx. 170 g
Dimensions 84.0 W x 26.6 H x 84.0 D mm (Excluding protrusions)
Control Software DCS-100A
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories USB cable N-38 (1 m)
Ground wire P-72 (5 m)
DVD (Dynamic data acquisition software DCS-100A)
Optional Accessories AC adapter UN310-0515 (For U.S.A.: UN312-0520)
USB Ports The number of measuring units at USB bus-power operation, refer to the following table.

【USB3.0】--------------------------------------------------------------------------------
EDX-11A units   Number of EDX-12A/13A/14A/15A  Connection units
    0      Up to 4 units in any combination    Max. 4
    1      Until either 1 unit           Max. 2
    1      0                  Max. 2

【USB2.0】--------------------------------------------------------------------------------
EDX-11A units   Number of EDX-12A/13A/14A/15A  Connection units
    0      Up to 2 units in any combination    Max. 2
    1      0                  Max. 1

Dimensions

EDX-10B

EDX-14A
Strain Measuring Unit

EDX-14A
  • CE

Specifications

Measuring Targets Strain-gage transducers, strain gages
*Bridge boxes or input adapters are required for strain measurement.
Channels 4
Measuring Range 10 k, 50 k ×10-6 strain (2 steps)
Compatible Bridge Resistance 120 Ω to 1 kΩ
Bridge Excitation 1 VDC
Gage Factor 2.00 fixed
Range Accuracy Within ±0.3%FS
Balance Adjustment Within ±1/2 F.S. of setting range
Nonlinearity Within ±0.1%FS
AD Converter 24 bits
Frequency Response DC to 2 kHz
LPF Cutoff frequencies : 100 Hz, 2 k Hz
Transfer characteristic : 2nd order Butterworth
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 90% (Non-condensing)
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 140 mA or less
(120 Ω load with all channels connected, at power supply 5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Strain input cable U-124 (30 cm)
Optional Accessories Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A
Bridge adapter for quarter bridge system UI-53B-120/350
Bridge adapter for quarter bridge system UI-54B-120/350
One-touch type input adapter UI-55A

Dimensions

EDX-14A

EDX-15A
Strain/Voltage Measuring Unit

EDX-15A
  • CE

Specifications

Measuring Targets Strain-gage transducers, strain gages
*Bridge boxes or input adapters are required for strain measurement.

Voltage
Channels 4
Measuring Range Strain-gage transducers, strain gages:
 10 k, 50 k ×10-6 strain (2 steps)

Voltage: 10, 50 V
Compatible Bridge Resistance 120 Ω to 1 kΩ
Bridge Excitation 1 VDC
Gage Factor 2.00 fixed
Range Accuracy Within ±0.3%FS
Balance Adjustment Within ±1/2 F.S. of setting range
Nonlinearity Within ±0.1%FS
AD Converter 24 bits
Frequency Response DC to 2 kHz
LPF Cutoff frequencies : 100 Hz, 2 k Hz
Transfer characteristic : 2nd order Butterworth
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 90% (Non-condensing)
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 150 mA or less
(120 Ω load with all channels connected, at power supply 5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Strain input cable U-124 (30 cm)
Conversion adapter FV-1A x 4
Optional Accessories Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A
Bridge adapter for quarter bridge system UI-53B-120/350
Bridge adapter for quarter bridge system UI-54B-120/350
One-touch type input adapter UI-55A
BNC input cable U-125 (30 cm)

Dimensions

EDX-15A

EDX-12A
Voltage Measuring Unit

EDX-12A
  • CE

Specifications

Measuring Targets Voltage
Channels 4 (Single end)
Measuring Range 10 V, 50 V (2 steps)
Range Accuracy Each range within ±0.3%FS
Balance Adjustment Within ±1/2 FS of setting range
Nonlinearity Within ±0.1%FS
AD Converter 24 bits
Frequency Response DC to 2 kHz
LPF Cutoff frequencies : 100 Hz, 2 k Hz
Transfer characteristic : 2nd order Butterworth
Operating Temperature 0 to 40ºC
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 110 mA or less (5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Input adapter UI-51A
Optional Accessories BNC input cable U-125 (30 cm)
Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
One-touch type input adapter UI-52A

Dimensions

EDX-12A

EDX-13A
Thermocouple Measuring Unit

EDX-13A
  • CE

Specifications

Measuring Targets Thermocouples
Channels 4
Input Connectors Screw type terminal box
Applicable Thermocouples K, T, J, N
(Resistance of thermocouple: 1 kΩ or less)
Range K: -200.0 to 1370.0ºC
T: -200.0 to 400.0ºC
J: -200.0 to 1200.0ºC
N: -200.0 to 1300.0ºC
Resolution 0.1°C
Accuracy [K]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1370.0°C: ± (0.1% of reading + 0.4ºC)

[T]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 400.0°C: ± (0.1% of reading + 0.4ºC)

[J]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1200.0°C: ± (0.1% of reading+0.4ºC)

[N]
−200.0 to below -100.0°C: ± (0.2% of reading + 0.6ºC)
−100.0 to 1300.0°C: ± (0.1% of reading+0.4ºC)

Note: The measurement accuracy does not include the accuracy of the internal reference junction compensator and thermocouples.
Accuracy of internal reference junction compensator ±1.0°C
〔Input terminal temperature at equilibrium〕
〔Ambient temperature: 25 ±10°C〕
Mount the EDX-13A on the bottom when using it with measuring units other than the EDX-13A.

±2.0°C
〔Input terminal temperature in equilibrium〕
For temperatures other than those in the ambient temperature and operating temperature described above
Check Functions Burnout check
AD Converter 24 bits
Sampling System Scanning
Inside Samping Frequencies Approx. 0.5 Hz, approx. 2.0 Hz
Operating Temperature 0 to 40℃
Operating Humidity 20 to 90% (Non-condensing)
Input Connector Screw type terminal box
Power Supply 5 VDC supplied by control unit
Current Consumption 120 mA or less (5 VDC)
Weight Approx. 130 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Terminal box 1 piece
Screwdriver 1 piece

Dimensions

EDX-13A

EDX-11A
Strain Measuring Unit

EDX-11A
  • CE

Specifications

Measuring Targets Strain-gage transducers, strain gage
*Bridge boxes or input adapters are required for strain measurement.
Channels 4
Measuring Range 10 k, 50 k ×10-6 strain (2 steps)
Balance Adjustment Within ±1/2 F.S. of setting range
Compatible Bridge Resistance 120 Ω to 1 kΩ
Bridge Excitation 2 VDC
Gage Factor 2.00 fixed
Range Accuracy Each range within ±0.3%FS
Nonlinearity Within ±0.1%FS
AD Converter 24 bits
Frequency Response DC to 2 kHz
LPF Cutoff frequencies : Lo (100 Hz), Hi (2 k Hz)
Transfer characteristic : 2nd order Butterworth
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 90% (Non-condensing)
Input Connectors D-sub 37-pin connector
Power Supply 5 VDC supplied by control unit
Current Consumption 180 mA or less
(120 Ω load with all channels connected, at power supply 5 VDC)
Weight Approx. 150 g
Dimensions 84.0 W × 26.6 H × 84.0 D mm
(Excluding protrusions)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Strain input cable U-124 (30 cm)
Optional Accessories Bridge box connection cable U-126 (50 cm)
Input connector set EDX10-DSUB
Input adapter UI-51A
One-touch type input adapter UI-52A, UI-55A
Bridge adapter for quarter bridge system UI-53B-120/350
Bridge adapter for quarter bridge system UI-54B-120/350

Dimensions

EDX-11A

Page Top

eBook