Conditioner Cards for EDX Series

Summary
· Conditioner cards for EDX-200A and EDX-5000A.
Features
The cards are for measuring strain, voltage, acceleration, and temperature. CAN data is supported. Isolated amplifier cards and carrier-wave amplifier cards also available for countermeasure against noise.
Products table
All display
CDV-40B
Strain/Voltage Measurement Card

Specifications
Measuring Targets | [Strain Measurement] Strain gages Strain-gage transducers [Voltage Measurement] Voltage |
---|---|
Channels | 8 |
Input Modes | [Strain Measurement] Balanced differential [Voltage Measurement] Unbalanced |
Input Resistance | [Strain Measurement] Within (10 MΩ + 10 MΩ) ±10% [Voltage Measurement] Within 1 MΩ±10% |
Coupling | DC/AC |
Frequency Response | [DC coupling] DC to 50 kHz (Deviation: +1 dB to -3 dB) [DC cut (AC coupling)] 0.2, 1 to 50 kHz (See the HPF) |
Gage Factor | [Strain Measurement] 2.00 [Voltage Measurement] - |
Bridge Excitation | [Strain Measurement] 2 VDC ±2% [Voltage Measurement] - |
Compatible Bridge Resistance | [Strain Measurement] 120 to 1000 Ω [Voltage Measurement] - |
Balance Adjustment | [Strain Measurement] ±2.4% or more (±12000 ×10-6 strain) [Voltage Measurement] - |
Balance Adjustment Methods | [Strain Measurement] Auto balance (Saved in nonvolatile memory) Accuracy: Within ±(0.1% FS +2 ×10-6 strain) [Voltage Measurement] - |
Measuring Range | [Strain Measurement] 8 steps of 500, 1 k, 2 k, 5 k, 10 k, 20 k, 50 k (×10-6 strain), OFF [Voltage Measurement] 8 steps of 0.1, 0.2, 0.5, 1, 2, 5, 10 V, OFF |
Range Accuracy | Within ±0.2% FS |
Calibration | ±100%, ±50%, each range Accuracy: Within ±0.3% FS |
Nonlinearity | Within ±0.1% FS |
LPF | Transfer characteristics : 2nd order Butterworth Cutoff frequencies : 8 steps of 10, 30, 100, 300, 1 k, 3 k, 10 k [Hz] and FLAT Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
HPF (DC cut) | Cutoff frequencies: 0.2, 1 Hz Attenuation: -6 dB/oct. |
AD Converter | 16 bits |
Sampling Frequencies | 200 kHz (Max.) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Optional Accessories | 8-channel input cable U-38 to U-48 Conversion adapter FV-1A |
---|---|
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. *Models with output are available, inquiries are welcome. |
CDV-40B-F
Strain/Voltage Measurement Card, with Antialiasing LPF

Specifications
Measuring Targets | [Strain Measurement] Strain gages Strain-gage transducers [Voltage Measurement] Voltage |
---|---|
Channels | 8 |
Input Modes | [Strain Measurement] Balanced differential [Voltage Measurement] Unbalanced |
Input Resistance | [Strain Measurement] Within (10 MΩ + 10 MΩ) ±10% [Voltage Measurement] Within 1 MΩ±10% |
Coupling | DC/AC |
Frequency Response | [DC coupling] DC to 50 kHz (Deviation: 1 dB to -3 dB) [DC cut (AC coupling)] 0.2, 1 to 50 kHz (See the HPF) |
Gage Factor | [Strain Measurement] 2.00 [Voltage Measurement] - |
Bridge Excitation | [Strain Measurement] 2 VDC ±2% [Voltage Measurement] - |
Compatible Bridge Resistance | [Strain Measurement] 120 to 1000 Ω [Voltage Measurement] - |
Balance Adjustment | [Strain Measurement] ±2.4% or more (±12000 ×10-6 strain) [Voltage Measurement] - |
Balance Adjustment Methods | [Strain Measurement] Auto balance (Saved in nonvolatile memory) Accuracy: Within ±(0.1% FS +2 ×10-6 strain) [Voltage Measurement] - |
Measuring Range | [Strain Measurement] 8 steps of 500, 1 k, 2 k, 5 k, 10 k, 20 k, 50 k (×10-6 strain), OFF [Voltage Measurement] 8 steps of 0.1, 0.2, 0.5, 1, 2, 5, 10 V, OFF |
Range Accuracy | Within ±0.2% FS |
Calibration | ±100%, ±50%, each range Accuracy: Within ±0.3% FS |
Nonlinearity | Within ±0.1% FS |
LPF | Transfer characteristics : 2nd order Butterworth Cutoff frequencies : 8 steps of 10, 30, 100, 300, 1 k, 3 k, 10 k [Hz] and FLAT Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
HPF (DC cut) | Cutoff frequencies: 0.2, 1 Hz Attenuation: -6 dB/oct. |
Antialiasing LPF | Available when the LPF is set to AUTO on the DCS-100A. Transfer characteristics: 8th order Butterworth Cutoff frequencies = Automatically set at sampling frequencies x 0.25 Attenuation: -48 dB ±5 dB (At sampling frequency x 0.5) Note: Available when the sampling frequency is 100 Hz or more. |
AD Converter | 16 bits |
Sampling Frequencies | 200 kHz (Max.) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Optional Accessories | 8-channel input cable U-38 to U-48 Conversion adapter FV-1A |
---|---|
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
CVM-41A
Strain/Voltage/Acceleration Measurement Card

Specifications
Applicable Recorders | EDX-100A EDX-200A EDX-5000A |
---|---|
Channels | 8 |
Measuring Targets | [Strain measurement] Strain gages, strain-gage transducers * For strain measurement, use bridge boxes [Voltage measurement] Voltage Voltage-output sensors [Acceleration measurement (Piezoelectric)] Piezoelectric accelerometers (Built-in amplifier) |
Input Modes | [Strain measurement] Balanced differential input [Voltage measurement] Balanced differential input *When using the Conversion Adapter FV-1A, this becomes unbalanced input *Common mode input voltage range ±20 VDC, absolute input voltage range ±50 V [Acceleration measurement (Piezoelectric)] Unbalanced input *Conversion Adapter FV-1A usage possible |
Input Impedance | [Strain Measurement] - [Voltage measurement] 1 MΩ + 1 MΩ Within ±10% *When using Conversion Adapter FV-1A (At unbalanced input), within 1 MΩ ±10% [Acceleration measurement (Piezoelectric)] ー |
Bridge Excitation or Power Supply to Sensors | [Strain measurement] Const. voltage output BV2V: 2 VDC BV5V: 5 VDC BV10V: 10 VDC [Voltage measurement] Const. voltage output BV2V: 2 VDC (±1 V) BV5V: 5 VDC (±2.5 V) BV10V: 10 VDC (±5 V) or OFF 20 mA/channels or less [Acceleration measurement (Piezoelectric)] Const. current output (Approx. 4 mA) (Excitation voltage: Approx. 23 VDC, Load: 1 kΩ or less) *Each channel settable. The max. channels of CVM-41A in EDX-100A is 3 times of units of CVM |
Gage Factor | [Strain measurement] 2.00 fixed [Voltage measurement] - [Acceleration measurement (Piezoelectric)] - |
Compatible Bridge Resistance | [Strain measurement] BV2V: 120 to 1000 Ω BV5V: 350 to 1000 Ω BV10V: 500 to 1000 Ω [Voltage measurement] - [Acceleration measurement (Piezoelectric)] - |
Balance Operation Settings(Zero suppression) | [Strain measurement] -------Autobalance enabled Cancel the unbalanced bridge portion in the analog circuit, and zero the measurement value. -------Autobalance disabled Do not cancel the unbalanced bridge portion (The initial unbalanced value in the bridge circuit can be confirmed) [Voltage measurement] -------Zero suppression enabled Cancel the input voltage in the analog circuit, and zero the measurement value -------Zero suppression disabled Do not cancel the input voltage in the analog circuit (Display the input voltage as is) [Acceleration measurement (Piezoelectric)] - |
Balance Adjustment Range | [Strain measurement] BV2V: Resistance ±10% (±50k ×10-6 strain) BV5V: Resistance ±4% (±20k ×10-6 strain) BV10V: Resistance ±2% (±10k ×10-6 strain) [Voltage measurement] ±5 V [Acceleration measurement (Piezoelectric)] - |
Measuring Range | [Strain measurement] BV2V: 5 k, 10 k, 20 k, 50 k, 100 k, 200 k, 500 k ×10-6 strain BV5V: 5 k, 10 k, 20 k, 50 k, 100 k, 200 k ×10-6 strain BV10V: 2 k, 5 k, 10 k, 20 k, 50 k, 100 k ×10-6 strain [Voltage measurement] 1, 2, 5, 10, 20, and 50 V [Acceleration measurement (Piezoelectric)] 100, 200, 500, 1000, 2000, and 5000 mV |
Range Accuracy | [Strain measurement] Within ±0.2%FS [Voltage measurement] Within ±0.2%FS [Acceleration measurement (Piezoelectric)] Within ±1.0%FS |
Nonlinearity | [Strain measurement] Within ±0.1%FS [Voltage measurement] Within ±0.1%FS [Acceleration measurement (Piezoelectric)] Within ±0.2%FS |
Calibration (CAL)SHUNT CAL | [Strain measurement] ±100% and ±50% of each range and SHUNT *When SHUNT CAL has 350 Ω load connected, approx. 257 ×10-6 strain output. [Voltage measurement] ±100% and ±50% of each range [Acceleration measurement (Piezoelectric)] ±100% and ±50% of each range |
Frequency Response | [Strain measurement] DC coupling: DC to 5 kHz, deviation +1dB, -3dB AC coupling: 0.2, 1 Hz to 5 kHz (See the HPF.) [Voltage measurement] DC coupling: DC to 5 kHz, deviation +1dB, -3dB AC coupling: 0.2, 1 Hz to 5 kHz (See the HPF.) [Acceleration measurement (Piezoelectric)] 0.5 Hz to 5 kHz, deviation +1dB, -3dB |
LPF | Transfer characteristics : 5 Butterworth type Cutoff frequencies : 30, 100, 300, 1 k, 3 k Hz, FLAT and AUTO *With AUTO settings, the cutoff frequency is set to 1/4 of the sampling frequency. Cutoff accuracy : Within -3 ±1 dB Attenuation : -30 (+3, -7) dB /oct. |
HPF | [Strain measurement] Cutoff frequencies: 0.2 Hz, 1 Hz Attenuation: -6 dB/oct. [Voltage measurement] Cutoff frequencies: 0.2 Hz, 1 Hz Attenuation: -6 dB/oct. [Acceleration measurement (Piezoelectric)] - |
Resolution | 24 bits *When installed in EDX-100A, its resolution becomes 16 bits. |
Distortion Rate | [Strain measurement] - [Voltage measurement] - [Acceleration measurement (Piezoelectric)] 1% or less |
Monitor Output | Accuracy: Within ±5 V ±0.5% (With ±FS) Nonlinearity: Within ±0.5%FS |
Dimensions | 22 W x 119 H x 213 D mm (Excluding protrusions) |
Weight | Approx. 400 g |
TEDS | Reads information from TEDS-installed sensors. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | 2 cross recessed binding head screw M3x6 |
---|---|
Optional Accessories | CCA input cable U-111 CVM input cable U-121 to U-123 CVM input integrated cable N-121 Integrated output cable U-62 Conversion adapter FV-1A Voltage input box VI-8A (-T) Bridge box for quarter bridge system DBS-120B-8 (C) (T) DBS-350B-8 (C) (T) One-touch type bridge box DBV-120A-8 (C) DBV-350A-8 (C) |
DPM-42B
Dynamic Strain Amplifier Card

Specifications
Measuring Targets | Strain gages Strain-gage transducers |
---|---|
Channels | 4 |
Frequency Response | DC to 5 kHz (Deviation: Within ±10%) |
Compatible Bridge Resistance | 120 to 1000 Ω |
Gage Factor | 2.00 fixed |
Bridge Excitation | 2, 0.5 Vrms selectable 12 kHz sine wave |
Balance Adjustment Range | |
Resistance | Within ±2.4% (±12000 ×10-6 strain) |
Capacity | Within 2000 pF |
Balance Adjustment Methods | |
Resistance | Auto balance |
Capacity | CST method (Capacitance self-tracking) |
Measuring Range | [With bridge excitation 2 Vrms] 200, 500, 1k, 2k, 5k, 10k, 20k ×10-6 strain and OFF - 8 steps [With bridge excitation 0.5 Vrms] 1k, 2k, 5k, 10k, 20k, 50k ×10-6 strain and OFF - 7 steps Accuracy: Within ±0.5% |
Calibration Values (CAL) | Output at ±100% and ±50% of each range |
Nonlinearity | Within ±0.2% FS |
SN Ratio | 50 dBp-p or more (Range: 500 ×10-6 strain) |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 kHz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
Resolution | 16 bits |
Additional Functions | Input check function, TEDS |
Monitor Output | Accuracy: ±5 V ±0.5% (At ±FS) Nonlinearity: Within ±0.5% FS |
Withstand Voltage | Between input and output: 250 VAC, 1 min. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Optional Accessories | Monitor output cable U-64 |
---|---|
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
DPM-42B-F
Dynamic Strain Amplifier Card, with Antialiasing LPF

Specifications
Measuring Targets | Strain gages Strain-gage transducers |
---|---|
Channels | 4 |
Frequency Response | DC to 5 kHz (Deviation: Within ±10%) |
Compatible Bridge Resistance | 120 to 1000 Ω |
Gage Factor | 2.00 fixed |
Bridge Excitation | 2, 0.5 Vrms selectable 12 kHz sine wave |
Balance Adjustment Range | |
Resistance | Within ±2.4% (±12000 ×10-6 strain) |
Capacity | Within 2000 pF |
Balance Adjustment Methods | |
Resistance | Auto balance |
Capacity | CST method (Capacitance self-tracking) |
Measuring Range | [With bridge excitation 2 Vrms] 200, 500, 1k, 2k, 5k, 10k, 20k ×10-6 strain and OFF – 8 steps [With bridge excitation 0.5 Vrms] 1k, 2k, 5k, 10k, 20k, 50k ×10-6 strain and OFF – 7 steps Accuracy: Within ±0.5% |
Calibration Values (CAL) | Output at ±100% and ±50% of each range |
Nonlinearity | Within ±0.2% FS |
SN Ratio | 50 dBp-p or more (Range: 500 ×10-6 strain) |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 kHz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
Antialiasing LPF | Transfer characteristics : 8th order Butterworth type Cutoff frequencies : Automatic setting at x0.25 sampling frequency Attenuation : -48 ±5 dB (When ×0.5 sampling frequency) Note: Enabled when "AUTO" set in LPF settings. |
Resolution | 16 bits |
Additional Functions | Input check function, TEDS |
Monitor Output | Accuracy: ±5 V ±0.5% (At ±FS) Nonlinearity: Within ±0.5% FS |
Withstand Voltage | Between input and output: 250 VAC, 1 min. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Optional Accessories | Monitor output cable U-64 |
---|---|
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
DPM-42B-I
Dynamic Strain Amplifier Card, Low Inverter Noise Type

Specifications
Measuring Targets | Strain gages Strain-gage transducers |
---|---|
Channels | 4 |
Frequency Response | DC to 5 kHz (Deviation: Within ±10%) |
Compatible Bridge Resistance | 120 to 1000 Ω |
Gage Factor | 2.00 fixed |
Bridge Excitation | 2, 0.5 Vrms selectable 12 kHz sine wave |
Balance Adjustment Range | |
Resistance | Within ±2.4% (±12000 ×10-6 strain) |
Capacity | Within 2000 pF |
Balance Adjustment Methods | |
Resistance | Auto balance |
Capacity | CST method (Capacitance self-tracking) |
Measuring Range | [With bridge excitation 2 Vrms] 200, 500, 1k, 2k, 5k, 10k, 20k ×10-6 strain and OFF – 8 steps [With bridge excitation 0.5 Vrms] 1k, 2k, 5k, 10k, 20k, 50k ×10-6 strain and OFF – 7 steps Accuracy: Within ±0.5% |
Calibration Values (CAL) | Output at ±100% and ±50% of each range |
Nonlinearity | Within ±0.2% FS |
SN Ratio | 44 dBp-p or more (Range: 500 ×10-6 strain) |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 kHz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
Resolution | 16 bits |
Additional Functions | Input check function, TEDS |
Monitor Output | Accuracy: ±5 V ±0.5% (At ±FS) Nonlinearity: Within ±0.5% FS |
Withstand Voltage | Between input and output: 250 VAC, 1 min. |
Noise Resistant | Low inverter noise |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Optional Accessories | Monitor output cable U-64 |
---|---|
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
DPM-42B-I-F
Dynamic Strain Amplifier Card, with Antialiasing LPF, Low Inverter Noise Type

Specifications
Measuring Targets | Strain gages Strain-gage transducers |
---|---|
Channels | 4 |
Frequency Response | DC to 5 kHz (Deviation: Within ±10%) |
Compatible Gage Resistance | 120 to 1000 Ω |
Gage Factor | 2.00 fixed |
Bridge Excitation | 2, 0.5 Vrms selectable 12 kHz sine wave |
Balance Adjustment Range | |
Resistance | Within ±2.4% (±12000 ×10-6 strain) |
Capacity | Within 2000 pF |
Balance Adjustment Methods | |
Resistance | Auto balance |
Capacity | CST method (Capacitance self-tracking) |
Measuring Range | [With bridge excitation 2 Vrms] 200, 500, 1k, 2k, 5k, 10k, 20k ×10-6 strain and OFF – 8 steps [With bridge excitation 0.5 Vrms] 1k, 2k, 5k, 10k, 20k, 50k ×10-6 strain and OFF – 7 steps Accuracy: Within ±0.5% |
Calibration Values (CAL) | Output at ±100% and ±50% of each range |
Nonlinearity | Within ±0.2% FS |
SN Ratio | 44 dBp-p or more (Range: 500 ×10-6 strain) |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 k Hz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
Antialiasing LPF | Transfer Characteristic : 8th order Butterworth type Cutoff frequencies : Automatic setting at x0.25 sampling frequency Attenuation : -48 ±5 dB (When ×0.5 sampling frequency) Notes: Enabled when "AUTO" set in LPF settings. |
Resolution | 16 bits |
Additional Functions | Input check function, TEDS |
Monitor Output | Accuracy: ±5 V ±0.5% (At ±FS) Nonlinearity: Within ±0.5% FS |
Withstand Voltage | Between input and output: 250 VAC, 1 min. |
Noise Resistant | Low inverter noise |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Optional Accessories | Monitor output cable U-64 |
---|---|
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
CTA-40A
Thermocouple Card

Specifications
Measuring Targets | Thermocouples |
---|---|
Channels | 8 |
Applicable Thermocouples | K (CA) T (CC) |
Thermocouple Resistance | 200 Ω or less (Burnout ON) 1000 Ω or less (Burnout OFF) |
Measuring Range | K1230, K480, K240, T400, T210, OFF K1230 (-200 to 1230ºC) K480 (-200 to 480ºC) K240 (-200 to 240ºC) T400 (-200 to 400ºC) T210 (-200 to 210ºC) |
General Accuracy | Within ±(0.5% of reading +1)°C (At ambient temp. 20 ±3°C) Within ±(0.5% of reading +2)°C (At ambient temp. 0 to 40°C) |
Calibration (CAL) | Output at 100% and 50% and 0ºC as absolute value of each range |
Frequency Response | DC to 10 Hz |
Resolution | 16 bits |
Burnout | Built-in: At burnout [Burnout display], with ON/OFF |
Monitor Output | Accuracy: Within ±5 V ±0.5% (At + FS) Nonlinearity: Within ±0.5%FS |
Isolation | Between input and output, and between channels: 50 MΩ or more (500 VDC) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | 8-channel input cable U-104 Temperature measuring adapter CT-2A x8 |
---|---|
Optional Accessories | Integrated output cable U-62 |
CFV-40A
F/V Converter Card

Specifications
Measuring Targets | Alternating signal output sensors |
---|---|
Channels | 4 |
Input Signals | AC (Zero cross), TTL level (Including open collector signals) |
Input Voltage | ± (0.5 V to 50 V): High hysteresis ± (0.1 V to 50 V): Low hysteresis |
Measuring Range | 50, 100, 500, 1 k, 2 k, 5 k, 10 k, 20 kHz and OFF - 9 steps Accuracy: Within ±0.1%FS |
Calibration (CAL) | Output at 100%, 50% (added), and 0% (Absolute value) of each range |
Response Time | 10 μs (Continuous pules input ) or less (2 cycles of input pules + 50 μs) or less (Input pules are broken) |
Resolution | 16 bits |
Sensor Power supply | 12 VDC: Within ±10% (Each channel 50 mA or less) |
Monitor Output | Accuracy: Within 5 V±0.5% (At + FS) Nonlinearity: Within ±0.1%FS |
Isolation | Between input and output, and between channels: 50 MΩ or more (500 VDC) |
Number of Cards Installed | Up to 2 cards [EDX-2000A/B-64, EDX-3000A/B] When two F/V cards are installed: Up to 4 other cards When one F/V card is installed: Up to 6 other cards [EDX-100A-1, EDX-200A-1, EDX-200A-2H] One F/V card can be installed [EDX-100A-2, EDX-100A-4, EDX-200A-4H, EDX-5000A-64/80] Two F/V cards can be installed |
Standard Accessories | Conversion adapter FV-1A x 4 |
---|---|
Optional Accessories | Input cable U-12 Input cable U-13 Monitor output cable U-64 |
CCA-40A
Charge Amplifier Card

Specifications
Measuring Targets | Piezoelectric accelerometers |
---|---|
Applied Accelerometers | Built-in amplifier (Voltage output) type |
Channels | 8 |
Power Supply to Sensors | Constant current power (Current: 4 mA, excitation voltage: approx. 24 VDC, load 1 kΩ or less) |
Frequency Response | 1 to 20 kHz (Deviation: +1 dB, -3 dB) |
Measuring Range | 20, 50, 100, 200, 500, 1000, 2000, 5000 mV and OFF - 9 steps Accuracy: Within ±1%FS |
Calibration | [DC CAL] ±100% and ±50% of each range Accuracy: Within ±0.2%FS [AC CAL] 100% and 50% of each range Accuracy: Within ±1%FS Frequency accuracy: Within 100 Hz ±5% |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 300, 1 k, 3 k, 10 k Hz, and FLAT (5 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
Distortion Factor | 1% or less |
Resolution | 16 bits |
Monitor Output | Accuracy: Within ±5 V ±1% (At + FS) |
TEDS | Reads information from TEDS-installed sensors. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Input cable U-111 |
---|---|
Optional Accessories | Integrated output cable U-62 Conversion adapter (BNC-miniature) BNCP-C25J-A Conversion adapter (Miniature-Tajimi) CCA-1B Conversion adapter (BNC-Tajimi) CCA-2B |
CCA-40A-F
Charge Amplifier Card, with Antialiasing LPF

Specifications
Measuring Targets | Piezoelectric accelerometers |
---|---|
Applied Accelerometers | Built-in amplifier (Voltage output) type |
Channels | 8 |
Power Supply to Sensors | Constant current power (Current: 4 mA, excitation voltage: approx. 24 VDC, load 1 kΩ or less) |
Frequency Response | 1 to 20 kHz (Deviation: +1 dB, -3 dB) |
Measuring Range | 20, 50, 100, 200, 500, 1000, 2000, 5000 mV and OFF - 9 steps Accuracy: Within ±1%FS |
Calibration | [DC CAL] ±100% and ±50% of each range Accuracy: Within ±0.2%FS [AC CAL] 100% and 50% of each range Accuracy: Within ±1%FS Frequency accuracy: Within 100 Hz ±5% |
LPF | Transfer characteristics : 2nd order Butterworth Cutoff frequencies : 300, 1 k, 3 k, 10 k Hz, and FLAT (5 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
Antialiasing LPF | Transfer characteristic : 8th order Butterworth type Cutoff frequencies : Automatic setting at × 0.25 sampling frequency Attenuation : -48 ±5dB (When × 0.5 sampling frequency) Note: Available when the LPF is set to AUTO. |
Distortion Factor | 1% or less |
Resolution | 16 bits |
Monitor Output | Accuracy: Within ±5 V ±1% (At +FS) |
TEDS | Reads information from TEDS-installed sensors. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Input cable U-111 |
---|---|
Optional Accessories | Integrated output cable U-62 Conversion adapter (BNC-miniature) BNCP-C25J-A Conversion adapter (Miniature-Tajimi) CCA-1B Conversion adapter (BNC-Tajimi) CCA-2B |
CAN-41A
CAN Card

Specifications
CAN Ports | 2 |
---|---|
Supported CAN Version | Bosch2.0B active support (ISO-11898 specifications-compliant) High-speed CAN/low-speed CAN selectable |
Connector Shape | Dsub 9-pin (male) |
Measurement ID | Max. 32 |
CAN Controller Operation Clock | 40 MHz,32 MHz |
Communication Speed(kbps) | high-speed CAN 10, 25, 33.3, 50, 62.5, 83.3, 100, 125,250, 500, 800, 1000 low-speed CAN 10, 25, 33.3, 50, 62.5, 83.3, 100, 125 |
Communication Conditions | Sample points, sampling frequency, resynchronization jump width selection. |
Measuring Channel Conditions | Start bit, bit length, data type, calibration coefficient (Conditions for conversion of extracting CAN data to physical quantity) |
Graph Display | Simultaneous display of graph, numerical value, frame, and analog data |
Others | Only one CAN-41A can be mounted in the last slot of the EDX-100A, or EDX-200A. CAN-41A can not be mounted in the EDX-5000A. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
CDA-44AS
Constant Current Amplifier Card, Compatible Bridge Resistance 120 Ω

Specifications
Measuring Targets | Strain gages (Full bridge system) Strain-gage transducers Voltage |
---|---|
Channels | 4 |
Input Resistance | Approx. 10 MΩ + 10 MΩ (Strain mode) Approx. 1 MΩ (Voltage mode) |
Input Format | Balanced differential input (Strain mode) Unbalanced input (Voltage mode) |
IMRR | 120 dB (When 500 ×10-6 strain range) |
Frequency Response | [DC coupling] DC to 200 Hz, deviation: +1 dB, -3 dB [DC cut (AC coupling)] 0.2 Hz (See the HPF) |
Gage Factor | 2.00 fixed (Strain mode) |
Compatible Bridge Resistance | 120 Ω |
Bridge Excitation | Approx. DC 16.7 mA (Constant current) when gage resistance 120 Ω connected *If sensitivity or temperature resistance is in the transducer bridge excitation lines, then sensitivity and temperature characteristics are not corrected. |
Cable Length | 500 m (At cross section: 0.5 mm2) |
Range Accuracy | Within ±0.3% FS |
Measuring Range | 500, 1 k, 2 k, 5 k, 10 k, 20 k ×10-6 strain, OFF (Strain mode) 1, 2, 5, 10, 20, 50 V, OFF (Voltage mode) |
Balance Adjustment | Within ±2.4% (±12000 ×10-6 strain) (At strain measurement) Within ±5 V (At voltage measurement) |
ZERO Accuracy | Within ±0.3% FS (Voltage OFF mode) |
Nonlinearity | Within ±0.1% FS |
Calibration (CAL) | Output at ±100% and ±50% of each range Accuracy: Within ±0.3%FS |
Monitor Output | Accuracy: Within ±5 V ±0.5% |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 1, 3, 10, 30, 100 Hz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
HPF | Cutoff frequencies : 0.2 Hz Attenuation : Within (-6 ±1) dB/oct. |
AD Converter | 16 bits |
TEDS | Reads information from TEDS-installed sensors. |
Isolation | Between input and case (output): Withstand voltage 500 VDC,1 min. Between channels: Withstand voltage 500 VDC,1 min. |
Standard Accessories | Conversion adapter FV-2A × 4 |
---|---|
Optional Accessories | Monitor output cable U-64 |
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
CDA-45AS
Constant Current Amplifier Card, Compatible Bridge Resistance 350 Ω

Specifications
Measuring Targets | Strain gages (Full bridge system) Strain-gage transducers Voltage |
---|---|
Channels | 4 |
Input Resistance | Approx. 10 MΩ + 10 MΩ (Strain mode) Approx. 1 MΩ (Voltage mode) |
Input Format | Balanced differential input (Strain mode) Unbalanced input (Voltage mode) |
IMRR | 120 dB (When 500 ×10-6 strain range) |
Frequency Response | [DC coupling] DC to 200 Hz, deviation: +1 dB, -3 dB [DC cut (AC coupling)] 0.2 Hz (See the HPF) |
Gage Factor | 2.00 fixed (Strain mode) |
Compatible Bridge Resistance | 350 Ω |
Bridge Excitation | Approx. DC 5.7 mA (Constant current) when gage resistance 350 Ω connected *If sensitivity or temperature resistance is in the transducer bridge excitation lines, then sensitivity and temperature characteristics are not corrected. |
Cable Length | Within 1 km (At cross section: 0.5 mm2) |
Range Accuracy | Within ±0.3% FS |
Measuring Range | 500, 1 k, 2 k, 5 k, 10 k, 20 k ×10-6 strain, OFF (Strain mode) 1, 2, 5, 10, 20, 50 V, OFF (Voltage mode) |
Balance Adjustment | Within ±2.4% (±12000 ×10-6 strain) (At strain measurement) Within ±5 V (At voltage measurement) |
ZERO Accuracy | Within ±0.3% FS (Voltage OFF mode) |
Nonlinearity | Within ±0.1% FS |
Calibration (CAL) | Output at ±100% and ±50% of each range Accuracy: Within ±0.3%FS |
Monitor Output | Accuracy: Within ±5 V ±0.5% |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 1, 3, 10, 30, 100 Hz, and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
HPF | Cutoff frequencies : 0.2 Hz Attenuation : Within (-6 ±1) dB/oct. |
AD Converter | 16 bits |
TEDS | Reads information from TEDS-installed sensors. |
Isolation | Between input and case (output): Withstand voltage 500 VDC,1 min. Between channels: Withstand voltage 500 VDC,1 min. |
Standard Accessories | Conversion adapter FV-2A × 4 |
---|---|
Optional Accessories | Monitor output cable U-64 |
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
CDV-44AS
Strain/Voltage Measurement Isolation Card

Specifications
Measuring Targets | Strain gages (Full bridge system) Strain-gage transducers Voltage |
---|---|
Channels | 4 |
Input Resistance | Approx. 10 MΩ + 10 MΩ (Strain mode) Approx. 1 MΩ (Voltage mode) |
Input Format | Balanced differential input (Strain mode) Unbalanced input (Voltage mode) |
IMRR | 120 dB (When 500 ×10-6 strain range) |
Gage Factor | 2.00 fixed (Strain mode) |
Frequency Response | [With DC coupling] DC to 5 kHz, deviation within +1 dB, -3 dB [DC cut (With AC coupling)] 0.2 Hz (See the HPF) |
Bridge Excitation | Within 2 VDC ±2% (Strain mode) |
Range Accuracy | Within ±0.3% FS |
Compatible Bridge Resistance | 120 to 1000 Ω (Strain mode) |
Measuring Range | 500, 1 k, 2 k, 5 k, 10 k, 20 k ×10-6 strain, and OFF (Strain mode) 1, 2, 5, 10, 20, 50 V, and OFF (Voltage mode) |
Balance Adjustment Range | Within ±2.4% (±12000 ×10-6 strain) (At strain measurement) Within ±5 V (At voltage measurement) |
ZERO Accuracy | Within ±0.3% FS (Voltage OFF mode) |
Nonlinearity | Within ±0.1% FS |
Calibration Values (CAL) | Output at ±100% and ±50% of each range Accuracy: Within ±0.3%FS |
Monitor Output | Accuracy: Within ±5 V ±0.5% (±5 V to full scale of each range) |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 kHz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
HPF | Cutoff frequencies : 0.2Hz Attenuation : Within (-6 ±1) dB/oct. |
AD Converter | 16 bits |
TEDS | Reads information from TEDS-installed sensors. |
Isolation | Between input and case (Output) : Withstand voltage 500 VDC, 1 min. Between channels : Withstand voltage 500 VDC, 1 min. |
Standard Accessories | Conversion adapter FV-2A × 4 |
---|---|
Optional Accessories | Monitor output cable U-64 (2 m) |
Note | If the transducer with a remote-sensing function, a 4-conductor extension cable (N-81 to N-85) enables measurement. But the remote-sensing function will be ineffective. |
CDV-46AS
Strain/Voltage Measurement Isolation Card

Specifications
Measuring Targets | Strain gages (Full-bridge system) Strain-gage transducers Voltage |
---|---|
Channels | 4 |
Input Format | Balanced differential input (Strain mode) Unbalanced input (Voltage mode) |
Input Resistance | Approx. 10 MΩ + 10 MΩ (Strain mode) Approx. 1 MΩ (Voltage mode) |
IMRR | 120 dB (When 2k ×10-6 strain range) |
Frequency Response | [With DC coupling] DC to 5 kHz, deviation within +1 dB, -3 dB [DC cut (With AC coupling)] 0.2 Hz (See the HPF) |
Gage Factor | 2.00 fixed (Strain mode) |
Bridge Excitation | Within 2 VDC ±2% (Strain mode) |
Compatible Bridge Resistance | 120 to 1000 Ω (Strain mode) |
Measuring Range | 2k, 5k, 10k, 20k, 50k, 100k ×10-6 strain, OFF (At strain measurement) 1, 2, 5, 10, 20, 50, OFF (At voltage measurement) |
Range Accuracy | Within ±0.3% FS |
Balance Adjustment Range | Within ±2.4% (±12000 ×10-6 strain) (At strain measurement) Within ±5 V (At voltage measurement) |
ZERO Accuracy | Within ±0.3% FS (Voltage OFF mode) |
Nonlinearity | Within ±0.1% FS |
Calibration Values (CAL) | Output at ±100% and ±50% of each range Accuracy: Within ±0.3% FS |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 kHz and FLAT (6 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
HPF | Cutoff frequencies : 0.2 Hz Attenuation : Within (-6 ±1) dB/oct. |
AD Converter | 24 bits |
Monitor Output | Accuracy: Within ±5 V ±0.5% (±5 V of full scale of each range) |
Isolation | Between input and case (output): Withstand voltage 500 VDC, 1 min. Between channels: Withstand voltage 500 VDC, 1 min. |
TEDS | Reads information from TEDS-installed sensors. |
Max. Sampling Frequency | 100 kHz |
Standard Accessories | Conversion adapter FV-2A × 4 |
---|
AD-40AS
AD Converter Card

Specifications
Channels | 8 |
---|---|
Input Range | ±5 V ±10 V OFF |
Input Modes | Unbalanced (Not balanced differential) |
Input Resistance | Approx. 1 MΩ |
AD Converter Methods | Successive approximation |
AD Converter | Resolution: 16 bits (± 32000 counts/FS) |
Accuracy | Within ±0.2% FS |
Nonlinearity | Within ±0.1% FS |
Input Frequencies | Range: DC to 50 kHz Deviation: 1dB to -3dB |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 k, 3 k, 10 kHz and FLAT (8 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
Power Supply to Sensors | ±2.5 V ±1%, each channel |
TEDS | Reads information from TEDS-installed sensors. |
Optional Accessories | 8-channel input cable: U-127 (1.5 m) Voltage input box: VI-8A with a cable N-121 (1.5 m) |
---|
AD-40AS-F
AD Converter Card, with Antialiasing LPF

Specifications
Channels | 8 |
---|---|
Input Range | ±5 V ±10 V OFF |
Input Modes | Unbalanced (Not balanced differential) |
Input Resistance | Approx. 1 MΩ |
AD Converter Methods | Successive approximation |
AD Converter | Resolution: 16 bits (± 32000 counts/FS) |
Accuracy | Within ±0.2% FS |
Nonlinearity | Within ±0.1% FS |
Input Frequencies | Range: DC to 50 kHz Deviation: 1dB to -3dB |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300, 1 k, 3 k, 10 k Hz and FLAT (8 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
Antialiasing LPF | Transfer characteristics : 8th order Butterworth Cutoff frequencies : A quarter of sampling frequency (auto setting) (Set LPF to [AUTO]) Attenuation : -48 ±5 dB |
Power Supply to Sensors | ±2.5 V ±1%, each channel |
TEDS | Reads information from TEDS-installed sensors. |
Optional Accessories | 8-channel input cable: U-127 (1.5 m) Voltage input box: VI-8A with a cable N-121 (1.5 m) |
---|
Remarks
· EDX-5000A does not support CAN card CAN-41A.
eBook
Download
Manual
- CAN-41A INSTRUCTION MANUAL (984KB)
- CCA-40A INSTRUCTION MANUAL (2.9MB)
- CDA-44AS/45AS INSTRUCTION MANUAL (678KB)
- CDV-40A/B INSTRUCTION MANUAL (586KB)
- CDV-44AS INSTRUCTION MANUAL (486KB)
- CFV-40A INSTRUCTION MANUAL (748KB)
- CTA-40A INSTRUCTION MANUAL (426KB)
- CVM-41A INSTRUCTION MANUAL (1.2MB)
- DPM-42B INSTRUCTION MANUAL (2.8MB)