EDX-200A-2H/4H/4T Option Cards

Summary
· Measures CAN data and GPS information.
· With 8-bit digital I/O
Features
The ECAN-40A, ETIM-40A, and EGPC-40A are optional cards for EDX-200A. The cards can measure CAN data or GPS information without reducing the number of analog channels and sampling frequencies.
Products table
All display
ECAN-40A
Multichannel CAN Card

Specifications
Applicable Models | EDX-200A-4H EDX-200A-2H (Installed in option slot) *For EDX-200A-4T, please use ECAN-40A M72. |
---|---|
CAN Ports | 2 (High-speed CAN/low-speed CAN selectable) |
Channels | Up to 512 (Total for 2 ports) |
Compatible CAN Input Channels | CAN 2.0 A/B(Conforming to ISO-11898, ISO-11519-2) |
Baud Rates | High speed CAN 1000, 800, 500, 250, 125, 100, 83.3, 62.5, 50, 33.3, 25, 20, and 10 [kbps] Low speed CAN 125, 100, 83.3, 62.5, 50, 33.3, 25, 20, and 10 [kbps] |
CAN Data Output | Output at start: Output any given CAN data when AD conversion starts Output at stop: Output any given CAN data when AD conversion stops Manual output: Output any given CAN data at an arbitrary timing. Interval output: Output any given CAN data in a predetermined fixed cycle. |
Digital I/O
I/O Points | Max. 8 |
---|---|
I/O Settings | Switch among digital input, digital output and remote-controlled input for each bit (Common applied to all). *Remote control input: Start and stop measuring, execute BAL, etc. |
Input Type | Isolated type, TTL level input |
Input Voltage | Max. 5 VDC |
Isolation Methods | Digital isolator |
Output Type | Isolated type, open collector type output (With 10 kΩ internal pull-up resistors) |
Output Voltage | 5 VDC |
Output Current | 25 mA max. (Per point) |
Isolation Methods | Digital isolator |
Connector Type | CAN port: D-sub connector (Male) 9-pin Digital I/O port: MDR connector (Female) 14-pin |
Operating Temperature | 0 to 50°C, ECAN-40A M72 is -20 to 65°C |
---|---|
Operating Humidity | 20 to 90% (Non-condensing) |
Storage Temperature | -20 to 60°C, ECAN-40A M72 is -30 to 70°C |
Dimensions | 22.0 W x 128.0 H x 221.5 D mm |
Weight | Approx. 170 g |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) |
Note: When using EDX-200A and DCS-101A for arithmetic operation, no CAN measurement is possible. |
Standard Accessories | Connector plug for digital I/O ports Shell case CANdb File Read Optional Software DCS-105A |
---|
ETIM-40A
Time Synchronization Card

Specifications
Applicable Models | EDX-200A-4H EDX-200A-2H (Installed in option slot) *For EDX-200A-4T, please use ETIM-40A M72. |
---|---|
Synchronization | Synchronizes the recording start time of remote EDX-200A units by using the time data received from GPS satellites. |
GPS Data | Latitude, longitude, elevation, course over ground,speed, time, receiving conditions, the number of satellites in use, etc. Saving format: NMEA format |
Digital I/O
I/O Points | Max. 8 |
---|---|
I/O Settings | Switch among digital input, digital output and remote-controlled input for each bit (Common applied to all). *Remote control input: Start and stop measuring, execute BAL, etc. |
Input Type | Isolated type, TTL level input |
Input Voltage | Max. 5 VDC |
Output Type | Isolated type, open collector type output (With 10 kΩ internal pull-up resistors) |
Output Voltage | 5 VDC |
Output Current | 25 mA max. (Per point) |
Isolation Methods | Digital isolator |
Connector Type | GPS sensor port: D-sub connector (Male) 9-pin Digital I/O port: MDR connector (Female) 14-pin |
Operating Temperature | 0 to 50°C ETIM-40A M72 is −20 to 65°C |
---|---|
Operating Humidity | 20 to 90% (Non-condensing) |
Storage Temperature | −20 to 60ºC ETIM-40A M72 is −30 to 70ºC |
Dimensions | 22.0 W x 128.0 H x 221.5 D mm |
Weight | Approx. 160 g |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) |
Standard Accessories | Connector plug for digital I/O ports GPS sensor (Cable length: 5 m) Shell case |
---|
EGPC-40A
GPS/Multichannel CAN Card

Specifications
Applicable Models | EDX-200A-4H EDX-200A-2H (Installed in optional slot) *For EDX-200A-4T, please use EGPC-40A M72. |
---|---|
I/O Ports | GPS/CAN Shared Port 1 Dsub connector 9 pins (male) CAN only port 1 Dsub connector 9 pins (male) Digital I/O port 1 MDR connector 14 pins (female) |
Channels | Up to 512 |
Compatible CAN Versions | CAN 2.0A/B (Conforming to ISO-11898, ISO-11519-2) |
Baud Rates | High speed CAN: 1000, 800, 500, 250, 125, 100, 83.3, 62.5, 50, 33.3, 25, 20, and 10 [kbps] Low speed CAN: 125, 100, 83.3, 62.5, 50, 33.3, 25, 20, and 10 [kbps] |
CAN Data Output | Output at start: Output any given CAN data when AD conversion starts Output at stop: Output any given CAN data when AD conversion stops Manual output: Output any given CAN data at an arbitrary timing. Interval output: Output any given CAN data in a predetermined fixed cycle. |
GPS Data | Latitude, longitude, elevation, course over ground,speed, time, receiving conditions, the number of satellites in use, etc. Saving format: NMEA format |
Synchronization | Synchronizes the recording start time of remote EDX-200A units by using the time data received from GPS satellites. |
Digital I/O
I/O Points | Max. 8 |
---|---|
I/O Settings | Switch among digital input, digital output and remote-controlled input for each bit (Common applied to all). *Remote-controlled input: Start and stop measuring, execute BAL, etc. |
Input Type | Isolated type (Digital isolator), TTL level input |
Input Voltage | Max. 5 VDC |
Output Type | Isolated type (Digital isolator) Open collector type output (With 10 kΩ internal pull-up resistors) |
Output Voltage | 5 VDC |
Output Current | 25 mA max. (Per point) |
Operating Temperature | 0 to 50ºC EGPC-40A M72 is -20 to 65ºC |
---|---|
Operating Humidity | 20 to 90% (Non-condensing) |
Storage Temperature | -20 to 60ºC EGPC-40A M72 is -30 to 70ºC |
Dimensions | 22.0 W x 128.0 H x 221.5 D mm |
Weight | Approx. 170 g |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) |
When using EDX-200A and DCS-101A for arithmetic operation, no CAN measurement is possible. |
Standard Accessories | Connector plug for digital I/O ports GPS sensor (Cable length: 5 m) Shell case CANdb File Read Optional Software DCS-105A |
---|