MCF-B
Multi Signal Conditioner

  • CE
  • TEDS

MCF-B

Summary

• Easy operation greatly reduces the working hours.
• 4-digit CAL switch.
• Hardly affected by noise.
• Various fixing fixtures and handles.
• You can set and control the MCF-B from your PC.
• Input Open Detection Function (DPM-91A/92A)

Features

The multi signal conditioner MCF-B is a portable amplifier. The MCF-B is able to build a high-performance conditioner system by mounting conditioner cards in the unit base.
The MCF-B is able to keep outputting the calibration values by using the CAL LOCK function.

Contents

■Unit Base
MCF-8B (For measurement of up to 8 conditioner cards)
MCF-16B (For measurement of up to 16 conditioner cards)

■Conditioner Cards*
DPM-91A/91A-I (Strain Amplifier Card, carrier frequency 5 kHz)
DPM-92A/92A-I (Strain Amplifier Card, carrier frequency 12 kHz)
 (The suffix “-I” means with feature of robustness against inverter noise.)
CDV-90A (Signal Conditioner Card)
CTA-90A (Thermocouple Card)
CCA-90A (Charge Amplifier Card)
CFV-90A (F/V Converter Card)

* The carrier frequencies are different.
 Please be sure DPM-91A and DPM-92A are not mounted in the same unit base.
* When using strain gages, use bridge boxes.
* When mounting the CTA-90A and an empty channel exists, be sure to mount the dummy card.

Products table

All display

MCF-8B
Multi Signal Conditioner, 8 channels

MCF-B
  • CE
  • TEDS

Specifications

Number of Conditioner Cards 8
Applicable Conditioner Cards Card Model ----------------Compatible Firmware
DPM-91A ------------------ Ver. 03.00 or later
DPM-91A-I ----------------- Ver. 03.00 or later
DPM-92A ------------------ Ver. 03.00 or later
DPM-92A-I ---------------- Ver. 03.00 or later
CDV-90A ------------------ Ver. 03.00 or later
CTA-90A ------------------ Ver. 01.02 or later
CFV-90A ------------------ Ver. 03.01 or later
CCA-90A ------------------ Ver. 03.00 or later
Monitor Meter Indicate output voltage of selected any channel by 1-digit sign and 4-digit value.
When error occurs, error No. and message are indicated.
Control Switches (Front)
CH select: Switching channels that indicate on the monitor meter.
BAL: Balance adjustment is executed all channels simultaneously.
+CAL, -CAL: Calibration output is executed all channels simultaneously.
KEY LOCK: When set to ON, no operation switches are available.

(Rear)
OSC select: To select the oscillator signal internal or external.
COM change-over: To change open or short between COM terminal and GND terminal.
Channel Indication CH LED lights up when the channel is monitored.
Key-lock Indication KEYLOCK LED lights up when Key-lock set to ON.
PHYSICAL QUANTITY Indication PHYSICAL QUANTITY LED lights up when the conditioner card in TEDS mode.
Master Indication OSC INT LED lights up when use DPM card and when OSC select switch sets to INT as internal oscillator.
Other Function Automatically select carrier frequency according to the types of the connected DPM card.
Command Control Function
Signaling System RS-485 half duplex system
Communication Speed 9600 bps
Character Length 8 bits
Parity None
Stop Bit 1 bit
Delimiter CR: Command transmission to the MCF-B.
CR+LF: Data transmission from the MCF-B.
Device ID 0 to F (up to 16 units can be controlled by one computer)

*Device IDs are set using the DIP switches in the monitor display section.
Communication Range Up to 200 m
Command Start command control
End command control
Select monitor display channel
Get card type
Get connected models
Get version
Get error number
Get voltage
Execute balance adjustment (DPM-90A series, CDV-90A)
Set calibration output
 (DPM-90A series, CDV-90A, CTA-90A, CFV-90A, CCA-90A)
Set internal gain
 (DPM-90A series, CDV-90A, CTA-90A, CFV-90A, CCA-90A)
Perform zero-point adjustment
 (DPM-90A series, CDV-90A, CFV-90A, CCA-90A)
Perform internal gain adjustment
 (DPM-90A series, CDV-90A, CTA-90A, CCA-90)
Set bridge voltage (CDV-90A)
Switch TEDS mode on/off (DPM-90A series, CDV-90A, CCA-90A)
Load TEDS information (DPM-90A series, CDV-90A, CCA-90A)
Set low-pass filter cutoff frequency (DPM-90A series, CDV-90A, CCA-90A)
Set high-pass filter cutoff frequency (DPM-90A series, CDV-90A)
Select thermocouple type (CTA-90A)
Switch measurement range (CFV-90A)
Set input coupling (CFV-90A)
Set trigger level (CFV-90A)
Switch Status in Command Control Mode Switch operations are locked (except for [CH SELECT] switches).
Notification in Command Control Mode [KEY LOCK] LED flashes.
Exiting Command Control Mode Canceled by command or using switch.
Settings After Exiting Command Control Mode Retains most recent settings.
Vibration Resistance 49.03 m/s2 (5 G) 5 to 55 Hz
15 cycles each in X, Y, Z directions, 1 min/cycle
Operating Temperature -10 to 50°C
Operating Humidity 20 to 85% (Non-condensing)
Storage Temperature -20 to 60°C
Power Supply 10 to 30 VDC
Current Consumption 1.5 A or less
(Fully equipped with CDV-90A cards, bridge excitation = 2 V, bridge resistance 60 Ω, power supply 15 VDC.)
Connector Centralized output connector (MDR connector, pincount 26, female)
BNC connector
Dimensions 246.4 W x 101 H x 170 D mm (Excluding protrusions)
Weight Approx. 3.5 kg, unit base only

Approx. 5 kg
(With 8 DPM-91A cards mounted.)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories DC power cable P-76 (Approx. 1.8 m)
Ground wire P-72 (Approx. 5 m)
Synchronous cable U-59 (Approx. 1.5 m)
Connector for concentralized output cable
Spare fuse
Instruction manual (CD-R)
Optional Accessories AC adapter UEA360-1540
1 channel dummy card MCF-DUMMY2
RS-485 extension cable for MCF N-136 (Approx. 1 m)
RS-485 communication Y cable for MCF N-137 (Approx. 0.3 m)
Independent left/right rack fixture MCF-BKT
MCF handle fixture MCF-HANDLE

Dimensions

MCF-B

MCF-16B
Multi Signal Conditioner, 16 channels

MCF-B
  • CE
  • TEDS

Specifications

Number of Conditioner Cards 16
Applicable Conditioner Cards Card Model ----------------Compatible Firmware
DPM-91A ------------------ Ver. 03.00 or later
DPM-91A-I ----------------- Ver. 03.00 or later
DPM-92A ------------------ Ver. 03.00 or later
DPM-92A-I ---------------- Ver. 03.00 or later
CDV-90A ------------------ Ver. 03.00 or later
CTA-90A ------------------ Ver. 01.02 or later
CFV-90A ------------------ Ver. 03.01 or later
CCA-90A ------------------ Ver. 03.00 or later
Monitor Meter Indicate output voltage of selected any channel by 1-digit sign and 4-digit value.
When error occurs, error No. and message are indicated.
Control Switches (Front)
CH select: Switching channels that indicate on the monitor meter.
BAL: Balance adjustment is executed all channels simultaneously.
+CAL, -CAL: Calibration output is executed all channels simultaneously.
KEY LOCK: When set to ON, no operation switches are available.

(Rear)
OSC select: To select the oscillator signal internal or external.
COM change-over: To change open or short between COM terminal and GND terminal.
Channel Indication CH LED lights up when the channel is monitored.
Key-lock Indication KEYLOCK LED lights up when Key-lock set to ON.
PHYSICAL QUANTITY Indication PHYSICAL QUANTITY LED lights up when the conditioner card in TEDS mode.
Master Indication OSC INT LED lights up when use DPM card and when OSC select switch sets to INT as internal oscillator.
Other Function Automatically select carrier frequency according to the types of the connected DPM card.
Command Control Function
Signaling System RS-485 half duplex system
Communication Speed 9600 bps
Character Length 8 bits
Parity None
Stop Bit 1 bit
Delimiter CR: Command transmission to the MCF-B.
CR+LF: Data transmission from the MCF-B.
Device ID 0 to F (up to 16 units can be controlled by one computer)

*Device IDs are set using the DIP switches in the monitor display section.
Communication Range Up to 200 m
Command Start command control
End command control
Select monitor display channel
Get card type
Get connected models
Get version
Get error number
Get voltage
Execute balance adjustment (DPM-90A series, CDV-90A)
Set calibration output
 (DPM-90A series, CDV-90A, CTA-90A, CFV-90A, CCA-90A)
Set internal gain
 (DPM-90A series, CDV-90A, CTA-90A, CFV-90A, CCA-90A)
Perform zero-point adjustment
 (DPM-90A series, CDV-90A, CFV-90A, CCA-90A)
Perform internal gain adjustment
 (DPM-90A series, CDV-90A, CTA-90A, CCA-90)
Set bridge voltage (CDV-90A)
Switch TEDS mode on/off (DPM-90A series, CDV-90A, CCA-90A)
Load TEDS information (DPM-90A series, CDV-90A, CCA-90A)
Set low-pass filter cutoff frequency (DPM-90A series, CDV-90A, CCA-90A)
Set high-pass filter cutoff frequency (DPM-90A series, CDV-90A)
Select thermocouple type (CTA-90A)
Switch measurement range (CFV-90A)
Set input coupling (CFV-90A)
Set trigger level (CFV-90A)
Switch Status in Command Control Mode Switch operations are locked (except for [CH SELECT] switches).
Notification in Command Control Mode [KEY LOCK] LED flashes.
Exiting Command Control Mode Canceled by command or using switch.
Settings After Exiting Command Control Mode Retains most recent settings.
Vibration Resistance 49.03 m/s2 (5 G) 5 to 55 Hz
15 cycles each in X, Y, Z directions, 1 min/cycle
Operating Temperature -10 to 50°C
Operating Humidity 20 to 85% (Non-condensing)
Storage Temperature -20 to 60°C
Power Supply 10 to 30 VDC
Current Consumption 3 A or less
(Fully equipped with CDV-90A cards, bridge excitation = 2 V, bridge resistance 60 Ω, power supply 15 VDC.)
Connector Centralized output connector (MDR connector, pincount 26, female)
BNC connector
Dimensions 440 W x 101 H x 170 D mm (Excluding protrusions)
Weight Approx. 5 kg, unit base only

Approx. 8 kg
(With 16 DPM-91A cards mounted.)
Compliance Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories DC power cable P-76 (Approx. 1.8 m)
Ground wire P-72 (Approx. 5 m)
Synchronous cable U-59 (Approx. 1.5 m)
Connector for concentralized output cable
Spare fuse
Instruction manual (CD-R)
Optional Accessories AC adapter UEA360-1540
1 channel dummy card MCF-DUMMY2
RS-485 extension cable for MCF N-136 (Approx. 1 m)
RS-485 communication Y cable for MCF N-137 (Approx. 0.3 m)
Independent left/right rack fixture MCF-BKT
MCF handle fixture MCF-HANDLE
Fixture for JIS rack for MCF-16A/B MCF-JIS
Fixture for DIN rack for MCF-16A/B MCF-DIN

Dimensions

MCF-B

DPM-91A
Strain Amplifier Cards

DPM-91A
  • TEDS

Specifications

Measuring Targets Strain gages
Strain-gage transducers (A separate bridge box is necessary.)
Channels 1
Compatible Bridge Resistance 60 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Excitation 2 Vrms
Balance Adjustment Resistance: Within ±2% (±10 k ×10-6 strain)
Capacitance: Within 2000 pF (Bridge resistance 120 Ω)
Balance Adjustment Methods Resistance: Auto balance
Accuracy: Within ±0.5 ×10-6 strain (When 5 V output to 100 ×10-6 strain input)
Capacitance: CST (Capacitance self-tracking)
Frequency Response DC to 2.5 kHz (Deviation: ±10%)
Carrier Frequencies 5 kHz
Nonlinearity Within ±0.2% FS
Calibration (CAL) ±(1 to 9999 ×10-6 strain)
Accuracy: Within ±(0.5% +0.5 ×10-6 strain)
Sensitivity Adjustment Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.

OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
INPUT switch range: 100 to 9999 ×10-6 strain in 1 ×10-6 strain step
Accuracy: Within ±(0.5% +5 mV)
Range: ×200 to ×50000
Fine Sensitivity Adjustment ×0.4 to ×1
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 10, 30, 100, 300, 1k Hz and FLAT (6 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct. (Except set at 1 kHz.)
HPF Cutoff frequencies : 0.2 Hz, OFF (2 steps)
SN Ratio Conditions: When 200 ×10-6 strain input and 5 V output, 120 Ω short.
DPM-91A: ≥ 49 dBp-p
Stability Temperature
Zero point: Within ±0.1×10-6 strain per ºC
Sensitivity: Within ±0.05% / ºC

Time
Zero point: Within ±1 ×10-6 strain per 8 h
Sensitivity: Within ±0.3% / 8 h
(When 100 ×10-6 strain input and 5 V output, zero point is measured by 120 Ω short, sensitivity is measured by 120 Ω bridge.)
Output Dual output, both BNC connector and integrated connector output the same signal

Voltage output: ±5 V, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
TEDS Reads the sensor's TEDS information then sets the rated output to OUTPUT switch as output voltage.
Input Open Detection Function Saturates output in the negative direction when input is open.
Connector Type Input connector : NDIS4102 (7 pins) connector
Output connector : BNC connector
Dimensions 24 W × 96 H ×170 D mm (Excluding protrusions)
Weight Approx. 190 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Output cable U-59

DPM-91A-I
Strain Amplifier Cards
(The suffix “-I” means with feature of robustness against inverter noise.)

DPM-91A-I
  • TEDS

Specifications

Measuring Targets Strain gages
Strain-gage transducers (A separate bridge box is necessary.)
Channels 1
Compatible Bridge Resistance 60 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Excitation 2 Vrms
Balance Adjustment Resistance: Within ±2% ( ±10 k ×10-6 strain)
Capacitance: Within 2000 pF (Bridge resistance 120 Ω)
Balance Adjustment Methods Resistance: Auto balance
Accuracy: Within ±0.5 ×10-6 strain (When 5 V output to 100 ×10-6 strain input)
Capacitance: CST (Capacitance self-tracking)
Frequency Response DC to 2.5 kHz (Deviation: ±10%)
Carrier Frequencies 5 kHz
Nonlinearity Within ±0.2% FS
Calibration (CAL) ±(1 to 9999 ×10-6 strain)
Accuracy: Within ±(0.5% + 0.5 ×10-6 strain)
Sensitivity Adjustment Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.

OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
INPUT switch range: 100 to 9999 ×10-6 strain in 1×10-6 strain step
Accuracy: Within ±(0.5% +5 mV)
Range: ×200 to ×50000
Fine Sensitivity Adjustment ×0.4 to ×1
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 10, 30, 100, 300, 1 k Hz and FLAT (6 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct. (Except set at 1 kHz.)
HPF Cutoff frequencies : 0.2 Hz, OFF (2 steps)
SN Ratio Conditions: When 200 ×10-6 strain input and 5 V output, 120 Ω short.
DPM-91A-I: ≥ 44 dBp-p
Stability Temperature
Zero point: Within ±0.1×10-6 strain per ºC
Sensitivity: Within ±0.05% / ºC

Time
Zero point: Within ±1 ×10-6 strain per 8 h
Sensitivity: Within ±0.3% / 8 h

(When 100 ×10-6 strain input and 5 V output, zero point is measured by 120 Ω short, sensitivity is measured by 120 Ω bridge.)
Output Dual output, both BNC connector and integrated connector output the same signal

Voltage output: ±5 V, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
TEDS Reads the sensor's TEDS information then sets the rated output to OUTPUT switch as output voltage.
Input Open Detection Function Saturates output in the negative direction when input is open.
Connector Type Input connector : NDIS4102 (7 pins) connector
Output connector : BNC connector
Dimensions 24 W × 96 H ×170 D mm (Excluding protrusions)
Weight Approx. 190 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Output cable U-59

DPM-92A
Strain Amplifier Cards

DPM-92A
  • TEDS

Specifications

Measuring Targets Strain gages
Strain-gage transducers (A separate bridge box is necessary.)
Channels 1
Compatible Bridge Resistance 60 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Excitation 2 Vrms
Balance Adjustment Resistance: Within ±2% (±10 k ×10-6 strain)
Capacitance: Within 2000 pF (Bridge resistance 120 Ω)
Balance Adjustment Methods Resistance: Auto balance
Accuracy: Within ±0.5 ×10-6 strain (When 5 V output to 100 ×10-6 strain input)
Capacitance: CST (Capacitance self-tracking)
Frequency Response DC to 5 kHz (Deviation: ±10%)
Carrier Frequencies 12 kHz
Nonlinearity Within ±0.2% FS
Calibration (CAL) ±(1 to 9999 ×10-6 strain)
Accuracy: Within ±(0.5% +0.5 ×10-6 strain)
Sensitivity Adjustment Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.

OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
INPUT switch range: 100 to 9999 ×10-6 strain in 1×10-6 strain step
Accuracy: Within ±(0.5% +5 mV)
Range: ×200 to ×50000
Fine Sensitivity Adjustment ×0.4 to ×1
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 10, 30, 100, 300, 1 k Hz and FLAT (6 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct. (Except set at 1 kHz.)
HPF Cutoff frequencies : 0.2 Hz, OFF (2 steps)
SN Ratio Conditions: When 200 ×10-6 strain input and 5 V output, 120 Ω short.
DPM-92A: ≥ 45 dBp-p
Stability Temperature
Zero point: Within ±0.1×10-6 strain per ºC
Sensitivity: Within ±0.05% / ºC

Time
Zero point: Within ±1 ×10-6 strain per 8 h
Sensitivity: Within ±0.3% / 8 h

(When 100 ×10-6 strain input and 5 V output, zero point is measured by 120 Ω short, sensitivity is measured by 120 Ω bridge.)
Output Dual output, both BNC connector and integrated connector output the same signal

Voltage output: ±5 V, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
TEDS Reads the sensor's TEDS information then sets the rated output to OUTPUT switch as output voltage.
Input Open Detection Function Saturates output in the negative direction when input is open.
Connector Type Input connector : NDIS4102 (7 pins) connector
Output connector : BNC connector
Dimensions 24 W × 96 H ×170 D mm (Excluding protrusions)
Weight Approx. 190 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Output cable U-59

DPM-92A-I
Strain Amplifier Cards
(The suffix “-I” means with feature of robustness against inverter noise.)

DPM-92A
  • TEDS

Specifications

Measuring Targets Strain gages
Strain-gage transducers (A separate bridge box is necessary.)
Channels 1
Compatible Bridge Resistance 60 to 1000 Ω
Gage Factor 2.00 fixed
Bridge Excitation 2 Vrms
Balance Adjustment Resistance: Within ±2% (±10 k ×10-6 strain)
Capacitance: Within 2000 pF (Bridge resistance 120 Ω)
Balance Adjustment Methods Resistance: Auto balance
Accuracy: Within ±0.5 ×10-6 strain (When 5 V output to 100 ×10-6 strain input)
Capacitance: CST (Capacitance self-tracking)
Frequency Response DC to 5 kHz (Deviation: ±10%)
Carrier Frequencies 12 kHz
Nonlinearity Within ±0.2% FS
Calibration (CAL) ±(1 to 9999 ×10-6 strain)
Accuracy: Within ±(0.5% + 0.5 ×10-6 strain)
Sensitivity Adjustment Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.

OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
INPUT switch range: 100 to 9999 ×10-6 strain in 1×10-6 strain step
Accuracy: Within ±(0.5% +5 mV)
Range: ×200 to ×50000
Fine Sensitivity Adjustment ×0.4 to ×1
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 10, 30, 100, 300, 1k Hz and FLAT (6 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct. (Except set at 1 kHz.)
HPF Cutoff frequencies : 0.2 Hz, OFF (2 steps)
SN Ratio Conditions: When 200 ×10-6 strain input and 5 V output, 120 Ω short.
DPM-92A-I: ≥ 40 dBp-p
Stability Temperature
Zero point: Within ±0.1×10-6 strain per ºC
Sensitivity: Within ±0.05% / ºC

Time
Zero point: Within ±1 ×10-6 strain per 8 h
Sensitivity: Within ±0.3% / 8 h

(When 100 ×10-6 strain input and 5 V output, zero point is measured by 120 Ω short, sensitivity is measured by 120 Ω bridge.)
Output Dual output, both BNC connector and integrated connector output the same signal

Voltage output: ±5 V, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
TEDS Reads the sensor's TEDS information then sets the rated output to OUTPUT switch as output voltage.
Input Open Detection Function Saturates output in the negative direction when input is open.
Connector Type Input connector : NDIS4102 (7 pins) connector
Output connector : BNC connector
Dimensions 24 W × 96 H ×170 D mm (Excluding protrusions)
Weight Approx. 190 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Output cable U-59

CDV-90A
Signal Conditioner Card

CDV-90A
  • TEDS

Specifications

Measuring Targets Strain gages, strain-gage transducers
Channels 1
Compatible Bridge Resistance 300 Ω to 1 kΩ (With excitation 10 V)
60 Ω to 1 kΩ (With excitation 2 V)
Gage Factor 2.00 fixed
Bridge Excitation 2 or 10 VDC, switchable
Balance Adjustment Resistance: Within ±2% (±10 k ×10-6 strain )
Balance Adjustment Methods Resistance: Auto balance
(Compensated value stored in nonvolatile memory)

Accuracy: Within ±1 ×10-6 strain
(With excitation 10 V, 5 V output to 200 ×10-6 strain input)
Frequency Response DC to 50 kHz (Deviation +0.5, -3 dB)
Nonlinearity Within ±0.05% FS
Calibration (CAL) ±(1 to 9999 ×10-6 strain)
Accuracy: Within ±(0.3% +1 ×10-6 strain)
Sensitivity Adjustment Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.

OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
INPUT switch range: 200 to 9999 ×10-6 strain in 1 ×10-6 strain step
Accuracy: Within ±(0.5% +5 mV)
Range: ×200 to ×5000
Fine Sensitivity Adjustment ×0.4 to ×1
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 10, 30, 100, 300, 1 k, 3 k, 10 k Hz and FLAT (8 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct.
HPF Cutoff frequencies : 0.2 Hz, OFF (2 steps)
Noise 10 ×10-6 strain p-p
(With excitation 10 V, 5 V output to 200 ×10-6 strain input)
Stability Temperature
Zero point: Within ±1 ×10-6 strain per °C
Sensitivity: Within ± 0.02%/°C

Time
Zero point: Within ±10 ×10-6 strain per 8 h
Sensitivity: Within±0.1%/8 h
(With excitation 10 V, 5 V output to 1000 ×10-6 strain input)
Output Dual output, both BNC connector and integrated connector output the same signal.

Voltage output: ±5 V or more, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V or wider
Output Off Function Available
Input Impedance 20 MΩ or more
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
TEDS Reads the sensor's TEDS information then sets the rated output to OUTPUT switch as output voltage.
Input Open Detection Function Saturates output in the negative direction when input is open.
Connector Type Input connector : NDIS4102 (7 pins) connector
Output connector : BNC connector
Dimensions 24 W × 96 H ×170 D mm (Excluding protrusions)
Weight Approx. 150 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Output cable U-59

CTA-90A
Thermocouple Card

CTA-90A

Specifications

Applicable Thermocouples K, T, J, N, E, and R
(Thermocouple resistance is 1 kΩ or less)
Measuring Range K: -200 to 1300°C
J: -200 to 1200°C
E: -200 to 1000°C
T: -200 to 400°C
N: -200 to 1300°C
R: 0 to 1700°C
Channels 1
Reference Junction Compensation ±2.5°C (At -10 to 50°C)
±1°C (At 20 ±3°C )
Frequency Response DC to 10 Hz (Deviation: +0.5, -1 dB)
Linearizer Accuracy Within ±0.5% FS (K, J, N, E, R)
Within ±1% FS (T)
Calibration 100 to 1700°C in every 100°C steps
(Varies with thermocouple models.)

Accuracy: Within ±0.5% FS
Sensitivity Adjustment Sensitivity is set in combination with a 2-digit INPUT switch and a 3-digit OUTPUT switch.

OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
INPUT switch range: 100 to 1700°C in 100°C step
Fine Sensitivity Adjustment ×0.4 to ×1
Noise 30 mVp-p or less (Input short)
Stability Temperature
Zero point: Within ±0.05% FS/°C
Sensitivity: Within ±0.05%/°C

Time
Zero point: Within ±0.5% FS/8 h
Sensitivity: Within ±0.5%/8 h

(When type K, 1300℃ input and 5 V output.
Zero point is measured by the input of voltage with response of 0℃, sensitivity is measured by the input of voltage with response of 1300℃.)
Output Dual output
(Both BNC connector and integrated connector output the same signal)

Voltage output: ±5 V or more (When load resistance 5 kΩ or more)
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
Connector Type Input connector: One-touch type terminal block
Output connector: BNC connector

Applicable wire
Solid wire: φ0.4 mm to φ1.3 mm (UL AWG16 to 26)
Twisted wire: φ0.2 mm2 to φ1.3 mm2 (UL AWG16 to 24)
Dimensions 24 W × 96 H ×170 D mm (Excluding protrusions)
Weight Approx. 140 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Standard Accessories Output cable U-59

CCA-90A
Charge Amplifier Card

CCA-90A
  • TEDS

Specifications

Channels 1
Measuring Targets IEPE accelerometer
Max. ±5000 mV (built-in amplifier type)
*Charge converter is necessary when using a charge output accelerometer.

Recommended item: Fuji ceramics corporation, "CAC1R0"
Input Format: Unbalanced input
Sensor supply: Built-in constant-current supply (constant current: 4mA)
Excitation voltage: Approx.24 V
Frequency Response 0.2 Hz to 50 kHz (Deviation: +1, -3 dB)
Sensitivity Adjustment Sensitivity is set in combination with a 4-digit INPUT switch and a 3-digit OUTPUT switch.

INPUT switch range: 20 to 5000 mV in 1-mV step
OUTPUT switch range: 1.00 to 5.00 V in 0.01-V step
Accuracy: Within ±(0.5% +5 mV)
Range: ×1 to ×250
Fine Sensitivity Adjustment ×0.4 to ×1
Calibration (DC CAL) ±(1 to 5000 mV)
Accuracy: Within ±0.3% FS
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 300, 1k, 3k, 10k Hz and FLAT (5 steps)
Amplitude ratio at cutoff point : -3 ±1 dB
Attenuation : (-12 ±1) dB/oct.
Distortion Rate Within 1% (When ±5 V output)
SN Ratio 48 dBP-P or more
(When 20 mV input and 5 V output)
Stability Temperature
Zero point: Within ±0.5 mV per °C
Sensitivity: Within ±0.05%/°C

Time
Zero point: Within ±5 mV per 8h
Sensitivity: Within ±0.5%/8h
TEDS Reads the sensor's TEDS information then displays it on the monitor meter to the value that converted the output voltage into the physical quantity.
Output Dual output, both BNC connector and integrated connector output the same signal.

Voltage output: ±5 V or more, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V or wider
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
Connector Type Input connector: BNC
Output connector: BNC
Dimensions 24 W ×96 H ×170 D mm (Excluding protrusions)
Weight Approx. 140 g
Compliance Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS)
Note The specifications does not include the accuracy of the optional charge converter.
Standard Accessories Output cable U-59

CFV-90A
F/V Converter Card

CFV-90A

Specifications

Channels 1
Measuring Targets AC signal
Pulse signal (including open collector signal)
Frequency Range 0.2 Hz to 100 kHz
*When set to 1.0 V output and 20 kHz range, 100 kHz input can measure as
5 V.
Input Voltage ±0.5 to ±50 V
Input Coupling AC or DC
Select by switch operation.
Trigger Level (Input Detection Level) 0.0 to 5.0 V
Trigger level is set from the 2-digit thumbwheel switch in 0.1-V step.
*When trigger level set to 5.1 V or more, trigger level saturates at 5.0 V.
Nonlinearity Within ±0.1% FS
Sensitivity Adjustment (RANGE) 500, 1k, 2k, 5k, 10k, 20k Hz (6 steps)

Full scale of the range is output with the voltage set with the OUTPUT switch (3-digit thumbwheel switch).

Accuracy: Within ±(0.1% + 5 mV)

[Setting Example]
When "SET" is executed with RANGE set to 20 kHz and OUTPUT switch set to 1.0 V, 1.0 V output at 20 kHz input, 5.0 V output at 100 kHz input.
Calibration (CAL) 100% or 50% of each RANGE
Accuracy: Within ±0.5%
Response Time When frequency continuous:
Less than one cycle of the input signal + 25 μs or less
(At the start of output voltage rising)

When frequency cutoff :
Less than two cycle of the input signal + 25 μs or less
(At the start of output voltage falling)
Power Supply to Sensors 12 VDC ±10% (At load current is 50 mA or less.)
Noise 30 mVP-P or less
(When input the signal: Square, frequency 100 kHz, amplitude: ±0.5 V)
Stability Temperature
Zero point: Within ±0.01% FS per ºC
Sensitivity: Within ±0.01%/ºC

Time
Zero point: Within ±0.05% FS per 8 h
Sensitivity: Within ±0.05%/8 h
Output Dual output, both BNC connector and integrated connector output the same signal.

Voltage output: +5 V or more, when load resistance 5 kΩ or more
Zero adjustment: -0.1 to 0.1 V or wider
Output Off Function Available
Output Impedance 2 Ω or less
Withstand Voltage 500 VAC for 1 min, between the following two positions respectively:
 Input and output
 Input and case
 Output and case
Over Input Indication LED lights up.
Connector Type Input connector: NDIS4109 (Small round 9 pins)
Output connector: BNC
Dimensions 24 W ×96 H ×170 D mm (Excluding protrusions)
Weight Approx. 140 g
Note When using MCF-16A, CFV-90A maximum installed amount are 12 units. (When CFV-90A only installed.)

When using mixed the other conditioner cards, CFV-90A maximum installed amount are 8 units.

When using MCF-8A, no restriction on the number of units to be installed.
Standard Accessories Output cable U-59
Input connector plug 4109P

Mounting Fixture

MCF-DIN
DIN Rack Mounting Bracket for MCF-16A

MCF-DIN
Object Product Model MCF-16A
MCF-16B

MCF-HANDLE
MCF Handle

MCF-HANDLE
Object Product Model MCF-8A
MCF-16A
MCF-8B
MCF-16B

MCF-JIS
JIS Rack Mounting Bracket for MCF-16A

MCF-JIS
Object Product Model MCF-16A
MCF-16B

Page Top

Remarks

* CFV-90A: RoHS compliant models are available. Inquiries are welcome.
* The software for the command control functions should prepared by yourself. We have released the control commands.
* To use the command control functions, you may require to update the firmware of conditioner cards. Note that the update should be handled by Kyowa (with charge).

eBook