NTB-100/200 Series
Network Terminal Box

Summary
· Network output conforms with to CAN, enables a single wire connection
· Digitized data transmission thereby strong against noise
· Compact & lightweight, so easy to install on site
Features
The NTB-100/200 Series is a data logger, which digitizes data from the sensor and can be decentralized arrangement.
A single unit measures 4 channels, and allows up to 99 units to be connected, so measurement up to 396 channels is possible.
Products table
All display
NTB-100B-120
Constant-voltage, one-touch terminal, 120 Ω

Specifications
Measuring Targets | Strain gages, strain-gage transducers Quarter bridge 120 Ω Half-bridge 120 to 1000 Ω Active-active system Full bridge 120 to 1000 Ω Full bridge |
---|---|
Channels | 4 |
Scanning Speed | Approx. 0.5 s/channel: 0 to ±30 k ×10-6 strain Approx. 1 s/channel: ±30 k ×10-6 strain or more |
Bridge Excitation | Approx. 2 VDC |
Measuring Range | 0 to ±300 k ×10-6 strain |
Resolution | 0 to ±30 k ×10-6 strain: 1 ×10-6 strain ±30 k to ±300 k ×10-6 strain: 10 ×10-6 strain |
Accuracy | 0 to ±30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain ±30 k to ±300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain |
TEDS | Reads information from TEDS-installed sensors. Channel name writing (Kyowa ID only) |
Power Save Mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interfaces | Dedicated interface conforming to CAN, cable extension up to 1 km |
Operating Temperature | -10 to 50 °C |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 100 mA or less Standby: 60 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W × 28 H × 55 D mm (Excluding protrusions) |
Weight | Approx. 310 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 Wire connection seals Rubber feet Screwdriver Control software NTB-10A *NTB-10A is optional for "NTB-100B-120-0". Instruction manual (CD-R) |
---|---|
Optional Accessories | Y cable N-103 (10 cm) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) *Please contact us for communication cables other than those listed above. AC adapter SA-10A-EDS (100 to 240 VAC) (For U.S.A.: UNI318-1215-EDS) Connection board/clip CN-1A DIN rail mounting plate Terminal resistor CANTERM120 USB/CAN converter LEAFLIGHT HS V2 |
Dimensions

NTB-100B-350
Constant-voltage, one-touch terminal, 350 Ω

Specifications
Measuring Targets | Strain gages, strain-gage transducers Quarter bridge 350 Ω Half-bridge 120 to 1000 Ω Active-active system Full bridge 120 to 1000 Ω Full bridge |
---|---|
Channels | 4 |
Scanning Speed | Approx. 0.5 s/channel: 0 to ±30 k ×10-6 strain Approx. 1 s/channel: ±30 k ×10-6 strain or more |
Bridge Excitation | Approx. 2 VDC |
Measuring Range | 0 to ±300 k ×10-6 strain |
Resolution | 0 to ±30 k ×10-6 strain: 1 ×10-6 strain ±30 k to ±300 k ×10-6 strain: 10 ×10-6 strain |
Accuracy | 0 to ±30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain ±30 k to ±300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain |
TEDS | Reads information from TEDS-installed sensors. Channel name writing (Kyowa ID only) |
Power Save Mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interfaces | Dedicated interface conforming to CAN, cable extension up to 1 km |
Operating Temperature | -10 to 50 °C |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 100 mA or less Standby: 60 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W × 28 H × 55 D mm (Excluding protrusions) |
Weight | Approx. 310 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 Wire connection seals Rubber feet Screwdriver Control software NTB-10A *NTB-10A is optional for "NTB-100B-350-0". Instruction manual (CD-R) |
---|---|
Optional Accessories | Y cable N-103 (10 cm) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) *Please contact us for communication cables other than those listed above. AC adapter SA-10A-EDS (100 to 240 VAC) (For U.S.A.: UNI318-1215-EDS) Connection board/clip CN-1A DIN rail mounting plate Terminal resistor CANTERM120 USB/CAN converter LEAFLIGHT HS V2 |
Dimensions

NTB-101A-120
Constant-voltage, screw soldering terminal, 120 Ω

Specifications
Measuring Targets | Strain gages, strain-gage transducers Quarter bridge 120 Ω Half-bridge 120 to 1000 Ω Active-active system Full bridge 120 to 1000 Ω Full bridge |
---|---|
Channels | 4 |
Scanning Speed | Approx. 0.5 s/channel: 0 to ±30 k ×10-6 strain Approx. 1 s/channel: ±30 k ×10-6 strain or more |
Bridge Excitation | Approx. 2 VDC |
Measuring Range | 0 to ±300 k ×10-6 strain |
Resolution | 0 to ±30 k ×10-6 strain: 1 ×10-6 strain ±30 k to ±300 k ×10-6 strain: 10 ×10-6 strain |
Accuracy | 0 to ±30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain ±30 k to ±300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain |
TEDS | Reads information from TEDS-installed sensors. Channel name writing (Kyowa ID only) |
Power Save Mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interfaces | Dedicated interface conforming to CAN, cable extension up to 1 km |
Operating Temperature | -10 to 50 °C |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 100 mA or less Standby: 60 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W × 28 H × 110 D mm (Excluding protrusions) |
Weight | Approx. 650 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 Wire connection seals Rubber feet Terminal block Control software NTB-10A *NTB-10A is optional for "NTB-101A-120-0". Instruction manual (CD-R) |
---|---|
Optional Accessories | Y cable N-103 (10 cm) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) *Please contact us for communication cables other than those listed above. AC adapter SA-10A-EDS (100 to 240 VAC) (For U.S.A.: UNI318-1215-EDS) Connection board/clip CN-1A DIN rail mounting plate Terminal resistor CANTERM120 USB/CAN converter LEAFLIGHT HS V2 |
Dimensions

NTB-101A-350
Constant-voltage, screw soldering terminal, 350 Ω

Specifications
Measuring Targets | Strain gages, strain-gage transducers Quarter bridge 350 Ω Half-bridge 120 to 1000 Ω Active-active system Full bridge 120 to 1000 Ω Full bridge |
---|---|
Channels | 4 |
Scanning Speed | Approx. 0.5 s/channel: 0 to ±30 k ×10-6 strain Approx. 1 s/channel: ±30 k ×10-6 strain or more |
Bridge Excitation | Approx. 2 VDC |
Measuring Range | 0 to ±300 k ×10-6 strain |
Resolution | 0 to ±30 k ×10-6 strain: 1 ×10-6 strain ±30 k to ±300 k ×10-6 strain: 10 ×10-6 strain |
Accuracy | 0 to ±30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain ±30 k to ±300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain |
TEDS | Reads information from TEDS-installed sensors. Channel name writing (Kyowa ID only) |
Power Save Mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interfaces | Dedicated interface conforming to CAN, cable extension up to 1 km |
Operating Temperature | -10 to 50 °C |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 100 mA or less Standby: 60 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W × 28 H × 110 D mm (Excluding protrusions) |
Weight | Approx. 650 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 Wire connection seals Rubber feet Terminal block Control software NTB-10A *NTB-10A is optional for "NTB-101A-350-0". Instruction manual (CD-R) |
---|---|
Optional Accessories | Y cable N-103 (10 cm) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) *Please contact us for communication cables other than those listed above. AC adapter SA-10A-EDS (100 to 240 VAC) (For U.S.A.: UNI318-1215-EDS) Connection board/clip CN-1A DIN rail mounting plate Terminal resistor CANTERM120 USB/CAN converter LEAFLIGHT HS V2 |
Dimensions

NTB-110B-350
Constant-current, one-touch terminal

Specifications
Measuring Targets | Civil engineering transducers Full bridge 350 Ω Full bridge Full bridge 350 Ω Civil engineering transducer with thermal sensors |
---|---|
Channels | 4 |
Scanning Speed | Approx. 0.5 s/channel: 0 to ±30 k ×10-6 strain Approx. 1 s/channel: ±30 k ×10-6 strain or more Approx. 1 s/channel: With civil engineering transducers with a thermal sensor |
Bridge Excitation | Approx. 5.6 mA (At bridge resistance 350 Ω) |
Measuring Range | [Strain measurement] 0 to ±30 k ×10-6 strain [Temperature measurement with civil engineering transducers with a thermal sensor] -30.0 to 70.0°C |
Resolution | [Strain measurement] 0 to ±30 k ×10-6 strain: 1 ×10-6 strain ±30 k to ±300 k ×10-6 strain: 10 ×10-6 strain [Temperature measurement with civil engineering transducers with a thermal sensor] 0.1°C |
Accuracy | [Strain measurement] 0 to ±30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain ±30 k to ±300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain [Temperature measurement with civil engineering transducers with a thermal sensor] ±0.5°C |
TEDS | Reads information from TEDS-installed sensors. Channel name writing (Kyowa ID only) |
Power Save Mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interfaces | Dedicated interface conforming to CAN, cable extension up to 1 km |
Operating Temperature | -10 to 50 °C |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 70 mA or less Standby: 60 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W × 28 H × 55 D mm (Excluding protrusions) |
Weight | Approx. 310 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 Wire connection seals Rubber feet Screwdriver Control software NTB-10A *NTB-10A is optional for "NTB-110B-350-0". Instruction manual (CD-R) |
---|---|
Optional Accessories | Y cable N-103 (10 cm) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) *Please contact us for communication cables other than those listed above. AC adapter SA-10A-EDS (100 to 240 VAC) (For U.S.A.: UNI318-1215-EDS) Connection board/clip CN-1A DIN rail mounting plate Terminal resistor CANTERM120 USB/CAN converter LEAFLIGHT HS V2 |
NTB-111A-350
Constant-current, screw soldering terminal

Specifications
Measuring Targets | Civil engineering transducers Full bridge 350 Ω Full bridge Full bridge 350 Ω Civil engineering transducer with thermal sensors |
---|---|
Channels | 4 |
Scanning Speed | Approx. 0.5 s/channel: 0 to ±30 k ×10-6 strain Approx. 1 s/channel: ±30 k ×10-6 strain or more Approx. 1 s/channel: With civil engineering transducers with a thermal sensor |
Bridge Excitation | Approx. 5.6 mA (At bridge resistance 350 Ω) |
Measuring Range | [Strain measurement] 0 to ±30 k ×10-6 strain (Constant-current bridge excitation) [Temperature measurement with civil engineering transducers with a thermal sensor] -30.0 to 70.0°C |
Resolution | [Strain measurement] 0 to ±30 k ×10-6 strain: 1 ×10-6 strain ±30 k to ±300 k ×10-6 strain: 10 ×10-6 strain [Temperature measurement with civil engineering transducers with a thermal sensor] 0.1°C |
Accuracy | [Strain measurement] 0 to ±30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain ±30 k to ±300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain [Temperature measurement with civil engineering transducers with a thermal sensor] ±0.5°C |
TEDS | Reads information from TEDS-installed sensors. Channel name writing (Kyowa ID only) |
Power Save Mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interfaces | Dedicated interface conforming to CAN, cable extension up to 1 km |
Operating Temperature | -10 to 50 °C |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 70 mA or less Standby: 60 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W × 28 H × 110 D mm (Excluding protrusions) |
Weight | Approx. 650 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 Wire connection seals Rubber feet Terminal block Control software NTB-10A *NTB-10A is optional for "NTB-111A-350-0". Instruction manual (CD-R) |
---|---|
Optional Accessories | Y cable N-103 (10 cm) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) *Please contact us for communication cables other than those listed above. AC adapter SA-10A-EDS (100 to 240 VAC) (For U.S.A.: UNI318-1215-EDS) Connection board/clip CN-1A DIN rail mounting plate Terminal resistor CANTERM120 USB/CAN converter LEAFLIGHT HS V2 |
NTB-201A
DC Voltage-output, thermocouples

Specifications
Channels | 4 |
---|---|
Scanning Speed | Approx. 0.5 s/channel |
Measuring Targets | DC voltage-output, thermocouples |
Voltage-output Measurement | [Range 10 V] Measuring range: 0 to ±10.0000 V Resolution: 100 μV Accuracy: ±(0.1% of reading +0.0003 V) Input resistance: Approx. 1 MΩ [Range 50 V] Measuring range: 0 to ±50.000 V Resolution: 1 mV Accuracy: ±(0.1% of reading +0.003 V) Input resistance: Approx. 1 MΩ |
Thermocouples | [Type K] Measuring range: -200.0 to 1230.0°C Resolution: 0.1ºC Accuracy: ±(0.2% of reading +0.3°C) (-200.0 to below -100°C) ±(0.1% of reading +0.2°C) (-100.0 to 1230.0°C) Accuracy of Internal Reference-junction Compensator: Within ±0.5ºC, when temperature balanced at input terminals, and the ambient temperature is 0 to 50ºC. Within ±1.0ºC, when temperature balanced at input terminals, and the ambient temperature is -10 to 0ºC. [Type T] Measuring range: -200.0 to 400.0°C Resolution: 0.1ºC Accuracy: ±(0.2% of reading +0.3°C) (-200.0 to below -100°C) ±(0.1% of reading +0.2°C) (-100.0 to 400.0 °C) Accuracy of Internal Reference-junction Compensator: Within ±0.5ºC, when temperature balanced at input terminals, and the ambient temperature is 0 to 50ºC. Within ±1.0ºC, when temperature balanced at input terminals, and the ambient temperature is -10 to 0ºC. [Type E] Measuring range: -200.0 to 660.0°C Resolution: 0.1ºC Accuracy: ±(0.2% of reading +0.3°C) (-200.0 to below -100°C) ±(0.1% of reading +0.2°C) (-100.0 to 660.0ºC) Accuracy of Internal Reference-junction Compensator: Within ±0.5ºC, when temperature balanced at input terminals, and the ambient temperature is 0 to 50ºC. Within ±1.0ºC, when temperature balanced at input terminals, and the ambient temperature is -10 to 0ºC. [Type J] Measuring range: -200.0 to 870.0°C Resolution: 0.1ºC Accuracy: ±(0.2% of reading +0.3°C) (-200.0 to below -100ºC) ±(0.1% of reading +0.2°C) (-100.0 to 870.0ºC) Accuracy of Internal Reference-junction Compensator: Within ±0.5ºC, when temperature balanced at input terminals, and the ambient temperature is 0 to 50ºC. Within ±1.0ºC, when temperature balanced at input terminals, and the ambient temperature is -10 to 0ºC. [Type R] Measuring range: 0.0 to 1760.0°C Resolution: 0.1ºC Accuracy: ±(0.2% of reading +0.8°C) (0.0 to below 100°C) ±(0.125% of reading +0.6°C) (100.0 to 1760.0°C) Accuracy of Internal Reference-junction Compensator: Within ±0.5ºC, when temperature balanced at input terminals, and the ambient temperature is 0 to 50ºC. Within ±1.0ºC, when temperature balanced at input terminals, and the ambient temperature is -10 to 0ºC. [Type N] Measuring range: -200.0 to 1300.0°C Resolution: 0.1ºC Accuracy: ±(0.2% of reading +0.3°C) (-200.0 to below -100°C) ±(0.1% of reading +0.2°C) (-100.0 to 1300.0°C) Accuracy of Internal Reference-junction Compensator: Within ±0.5ºC, when temperature balanced at input terminals, and the ambient temperature is 0 to 50ºC. Within ±1.0ºC, when temperature balanced at input terminals, and the ambient temperature is -10 to 0ºC. Notes: 1. Accuracies do not include the accuracies of the internal reference junction compensator and the sensors. 2. The reference junction compensator is switchable between internal and external. 3. The thermocouple resistance should be 1 kΩ or less. |
Check Functions | Burnout check |
Power Save Mode | Provided ON/OFF using “OPT.3" DIP switch. |
Interfaces | Dedicated interface conforming to CAN. |
Operating Temperature | -10 to 50ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply | 11 to 16 VDC |
Current Consumption | (At 12 VDC) Operation: 100 mA or less Standby (In power save mode): 40 mA or less |
Dimensions | 150 W x 28 H x 55 D mm (Excluding protrusions) |
Weight | Approx. 320 g |
Standard Accessories | DC power cable P-76 Ground wire P-72 |
---|
Dimensions

NTB-20A
NTB power supply box

Power Save Function (AUTO Mode) | When the PC power supply is OFF, then the power supply of the NTB-20 connected by a USB cable is off. |
---|---|
Power Supply Output Limitation Function | When a power supply exceeding the power supply range for operating the NTB-20A is input, for safety, the power supply output from the serial connector and the OUT connector is turned OFF. |
Specifications
NTB Series Connected Units and Cable Length | [With USB port] NTB series connected units: 1, Cable length: 200 m or less NTB series connected units: 2, Cable length: 100 m or less [When AC adapter or DC power supply used] NTB series connected units: 1, Cable length: 1000 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 2, Cable length: 840 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 3, Cable length: 560 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 4, Cable length: 470 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 5, Cable length: 330 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 6, Cable length: 280 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 7, Cable length: 240 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 8, Cable length: 200 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 9, Cable length: 180 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 10, Cable length: 160 m or less (When SA-10A-EDS or UIA345-12 used) NTB series connected units: 11, Cable length: 150 m or less (When UIA345-12 used) NTB series connected units: 12, Cable length: 130 m or less (When UIA345-12 used) NTB series connected units: 13, Cable length: 120 m or less (When UIA345-12 used) NTB series connected units: 14, Cable length: 120 m or less (When UIA345-12 used) NTB series connected units: 15, Cable length: 110 m or less (When UIA345-12 used) NTB series connected units: 16, Cable length: 100 m or less (When UIA345-12 used) |
---|---|
Power Save Functions | Provided In POWER switch "AUTO" mode (*1) (*1) In "AUTO" mode, turning off the PC power supply automatically turns OFF the power supply to the NTB-20A (NTB power supply box). When using "AUTO" mode, ensure that the PC and NTB-20A (NTB power supply box) are connected using a USB cable. |
Operating Temperature | -10 to 50ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Power Supply Input | USB port: 5 VDC External power supply: 11 to 16 VDC (AC adapter, DC power supply) |
Current Consumption | When using 12 VDC (Using AC adapter) OFF mode: 7.0 mA or less AUTO mode: 7.0 mA ON mode: 30.0 mA or less When using 5 VDC (Using USB port) OFF mode: 5.0 mA or less AUTO mode: 30.0 mA or less ON mode: 30.0 mA or less |
Dimensions | 150 W x 28 H x 55 D mm (Excluding protrusions) |
Weight | Approx. 260 g |
Optional Accessories | Connection cable N-38 (1 m), N-39 (2 m) Communication cable N-102 (1 m) Communication cable H-11681 (3 m) Communication cable H-11682 ( 5 m) Communication cable H-11683 (10 m) |
---|
Dimensions

NTB-21A
NTB relay box

Power Supply Output Limitation Function | When power supply exceeding the range is input into NTB, the power supply output from the serial connector is turned to OFF. |
---|
Specifications
Input Voltage Range | 11 to 16 VDC |
---|---|
Operating Temperature | -10 to 50ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Dimensions | 150 W x 28 H x 29 D mm (Excluding protrusions) |
Weight | Approx. 160 g |
Optional Accessories | Power and communication cable N-114 |
---|
Dimensions

NTB-10A
Network terminal box control software

Specifications
Operating Environment
OS | Windows® 8/8.1, Windows® 10 Japanese/English, 32/64 bits support |
---|---|
CPU | Core 2 Duo, 2 GHz or advanced |
Memory | If 32-bit OS, 2 GB or more If 64-bit OS, 4 GB or more |
Hard Disk | Empty storage 10 MB or more (Excluding the target data file size.) |
Display | 1024×768 pixels or more |
USB/CAN Converter | Manufacture: KVASER Model: LEAFLIGHT HS V2 |
Measuring Units | NTB series: 1 to 99 (Max. 396 channels) |
---|---|
Measuring Functions | RELATIVE measurement (Relative value measurement): Each value is obtained by subtracting the ZERO value. *ZERO value is equivalent to the initial unbalance value. Capable of obtaining the ZERO value at arbitrary timing. |
Channels Conditions | Meas channel ON/OFF CAL coefficient calculation ON/OFF CAL coefficient Offset Unit Dec. digits Resistance at 0℃ CH Name (20 characters) |
Measuring Condition File | Load and save |
Measuring Operations | MONITOR Meas (Measures ZERO value during MONITOR measurement.) One-time measurement Continuous measurements INTERVAL Meas (To be specified the number of measurements) |
Interval Measurement | Interval start time Interval time Repeat 0 to 9999 (0 to infinite) Interval measuring steps: 5 |
Numeric Display | Available windows: 1 Window switch: List only |
Graph Display | Y-time, BAR graph Max. 8 channels/graph |
Measured Data Saving Function | The measured data is saved with the CSV format. |
TEDS | Reads sensor's information and sets to channel condition automatically. Channel name writing |
Data File Destination | PC hard disk |
File Split | No split Split every hour Split every day |