SME-30A/31A/100A/101A
Handy Data Logger

  • TEDS

Summary

· Built-in bridge box for easy to connection strain gages
· Wide measuring range ±300 k μm/m
· Power supply AA batteries

Features

The SME Series are handy data loggers for connecting strain gages, strain-gage transducers, and civil engineering transducers with thermal sensor.

Products table

All display

SME-30A
Handy Data Logger

  • TEDS

Specifications

Channels 1
Sampling Period Approx. 0.5 s: 0 to ± 30 k ×10-6 strain
Approx. 1 s: ± 30 k ×10-6 strain or more
Approx. 1 s: Civil engineering transducers with a thermal sensor
Measuring Functions RELATIVE measurement (Relative value measurement):
Each value is obtained by subtracting the ZERO value.
*ZERO value is equivalent to the initial unbalance value. Capable of obtaining the ZERO value at arbitrary timing
Arithmetic Operations Calculation using coefficients
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor
Quarter bridge: 120, 240, and 350 Ω
Half/full bridge: 120 to 1000 Ω
Bridge Excitation Constant-voltage bridge excitation: Approx. 2 VDC
Constant-current bridge excitation: Approx. 5.6 mA
(Bridge resistance 350 Ω)
Measuring Range At strain measurement
0 to ± 300 k ×10-6 strain (Constant-voltage bridge excitation)
0 to ± 20 k ×10-6 strain (Constant-current bridge excitation)
When measuring temperature with civil engineering transducers with a thermal sensor: -30.0°C to 70.0°C
Resolution At strain measurement
0 to ± 30 k ×10-6 strain: 1 ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: 10 ×10-6 strain
When measuring temperature with civil engineering
transducer with a thermal sensor: 0.1°C
Accuracy At full bridge strain measurement
0 to ± 30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain
When measuring temperature with civil engineering transducers with a thermal sensor: ± 0.5°C
Check Functions Insulation resistance measurement: 2 M to 100 M Ω
Resistance measurement: 0 to 20 kΩ
Interval Measurement 1 minute to 99 hours 59 minutes in 1-minute steps
Starting time: year/month/day/hour/minute
Storage SD card (Option)
Applicable Cards 256 MB, 512 MB, 1 GB, 2 GB (FAT16)
(SDHC and SDXC not supported)
Display Monochrome LCD, 128 × 64
TEDS Reads information from TEDS-installed sensors.
Channel name writing
(Kyowa ID only in up to 10 characters)
Operating Temperature -10 to 50ºC
Operating Humidity 20 to 85% (Non-condensing)
Power Supply 2 AA alkaline batteries
Consecutive Operation Time Approx. 10 h (With alkaline batteries)
* Nickel metal hydride batteries is also used.
Auto Power Off Power is automatically turned off if no key operation is detected for 5 minutes. In interval measuring mode with an interval of 3 minutes or longer, power is automatically turned off during standby period and
turned on again 1 minute before the next measurement is started. (ON/OFF of Auto Power Off is settable.)
Dimensions 188 W × 41 H x 108.4 D mm (Excluding protrusions)
Weight Approx. 450 g (Excluding batteries)
Standard Accessories Input cable U-119
AA alkali battery × 2
Shoulder belt
Hand strap
Instruction manual (CD-R)

Dimensions

SME-30A

SME-31A
Handy Data Logger

  • TEDS

Specifications

Channels 1
Sampling Period Approx. 0.5 s: 0 to ± 30 k ×10-6 strain
Approx. 1 s: ± 30 k ×10-6 strain or more
Approx. 1 s: Civil engineering transducers with a thermal sensor
Measuring Functions RELATIVE measurement (Relative value measurement):
Each value is obtained by subtracting the ZERO value.*ZERO value is equivalent to the initial unbalance value. Capable of obtaining the ZERO value at arbitrary timing
Arithmetic Operations Calculation using coefficients
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor
Quarter bridge: 120, 240, and 350 Ω
Half/full bridge: 120 to 1000 Ω
Bridge Excitation Constant-voltage bridge excitation: Approx. 2 VDC
Constant-current bridge excitation: Approx. 5.6 mA
(Bridge resistance 350 Ω)
Measuring Range At strain measurement
0 to ± 300 k ×10-6 strain (Constant-voltage bridge excitation)
0 to ± 20 k ×10-6 strain (Constant-current bridge excitation)
When measuring temperature with civil engineering transducers with a thermal sensor: -30.0°C to 70.0°C
Resolution At strain measurement
0 to ± 30 k ×10-6 strain: 1 ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: 10 ×10-6 strain
When measuring temperature with civil engineering
transducer with a thermal sensor: 0.1°C
Accuracy At full bridge strain measurement
0 to ± 30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain
When measuring temperature with civil engineering transducers with a thermal sensor: ± 0.5°C
Check Functions Insulation resistance measurement: 2 M to 100 M Ω
Resistance measurement: 0 to 20 kΩ
Interval Measurement 1 minute to 99 hours 59 minutes in 1-minute steps
Starting time: year/month/day/hour/minute
Storage SD card (Option)
Applicable Cards 256 MB, 512 MB, 1 GB, 2 GB (FAT16)
(SDHC and SDXC not supported)
Display Monochrome LCD, 128 × 64
TEDS Reads information from TEDS-installed sensors.
Channel name writing
(Kyowa ID only in up to 10 characters)
Operating Temperature -10 to 50ºC
Operating Humidity 20 to 85% (Non-condensing)
Power Supply 2 AA alkaline batteries
Consecutive Operation Time Approx. 10 h (With alkaline batteries)
* Nickel metal hydride batteries is also used.
* An AC adapter (Optional, SW-0522E) is provided for SME-31A.
Auto Power Off Power is automatically turned off if no key operation is detected for 5 minutes. In interval measuring mode with an interval of 3 minutes or longer, power is automatically turned off during standby period and
turned on again 1 minute before the next measurement is started. (ON/OFF of Auto Power Off is settable.)
Dimensions 188 W × 41 H x 108.4 D mm (Excluding protrusions)
Weight Approx. 450 g (Excluding batteries)
Standard Accessories Input cable U-119
AA alkali battery × 2
Shoulder belt
Hand strap
Instruction manual (CD-R)
Optional Accessories AC adapter SW-0522E

Dimensions

SME-30A

SME-100A
Handy Data Logger

  • TEDS

Specifications

Channels 1 (In independent use of the logger)
Max. 33 channels with NTBs connected
Sampling Frequencies (In independent use, or NTB-dependent when
connected to NTBs)
Approx. 0.5 s: 0 to ± 30 k ×10-6 strain
Approx. 1 s: ± 30 k ×10-6 strain or more
Approx. 1 s: When measuring temperature with civil engineering transducers with a thermal sensor
Measuring Modes RELATIVE measurement (Relative value measurement):
Each value is obtained by subtracting the ZERO value.
*ZERO value is equivalent to the initial unbalance value.
Capable of obtaining the ZERO value at arbitrary timing
Arithmetic Operations Calculation using coefficients
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor
Quarter bridge: 120, 240, and 350 Ω
Half/full bridge: 120 to 1000 Ω
Bridge Excitation Constant-voltage bridge excitation: Approx. 2 VDC
Constant-current bridge excitation: Approx. 5.6 mA
(Bridge resistance 350 Ω)
Measuring Range At strain measurement
0 to ± 300 k ×10-6 strain (Constant-voltage bridge excitation)
0 to ± 20 k ×10-6 strain (Constant-current bridge excitation)
When measuring temperature with civil engineering
transducers with a thermal sensor: -30.0°C to 70.0°C
Resolution At strain measurement
0 to ± 30 k ×10-6 strain: 1 ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: 10 ×10-6 strain
When measuring temperature with civil engineering
transducers with a thermal sensor: 0.1°C
Accuracy At full bridge strain measurement
0 to ± 30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain
When measuring temperature with civil engineering
transducers with a thermal sensor: ± 0.5°C
Check Functions Insulation resistance measurement: 2 M to 100 M Ω
Resistance measurement: 0 to 20 kΩ
Interval Measurement 1 minute to 99 hours 59 minutes in 1-minute steps
Starting time: year/month/day/hour/minute
Storage SD card (Option)
SD Cards 256 MB, 512 MB, 1 GB, 2 GB (FAT16)
(SDHC and SDXC not supported)
Display Monochrome LCD, 128 × 64
TEDS Reads information from TEDS-installed sensors
Channel name writing
(Kyowa ID only within 10 characters)
Operating Temperature -10 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Power Supply AA battery × 2 Consecutive operation time:
Approx. 10 hours (With alkaline batteries, NTB not connected)
Nickel hydride batteries are also used.
Auto Power Off Power is automatically turned off if no key operation is detected for 5 minutes. In interval measuring mode with an interval of 3 minutes or longer, power is automatically turned off during standby period and turned on again 1 minute before the next measurement is started. (ON/OFF of Auto Power Off is settable.)
Dimensions 188 W × 41 H x 108.4 D mm (Excluding protrusions)
Weight Approx. 450 g (Excluding batteries)
Standard Accessories Input cable U-119
Communication cable N-102
AA alkali battery × 2
Shoulder belt
Hand strap
Instruction manual (CD-R)

Dimensions

SME-101A
Handy Data Logger

  • TEDS

Specifications

Channels 1 (In independent use of the logger)
Max. 33 channels with NTBs connected
Sampling Frequencies (In independent use, or NTB-dependent when
connected to NTBs)
Approx. 0.5 s: 0 to ± 30 k ×10-6 strain
Approx. 1 s: ± 30 k ×10-6 strain or more
Approx. 1 s: When measuring temperature with civil engineering transducers with a thermal sensor
Measuring Modes RELATIVE measurement (Relative value measurement):
Each value is obtained by subtracting the ZERO value.
*ZERO value is equivalent to the initial unbalance value.
Capable of obtaining the ZERO value at arbitrary timing
Arithmetic Operations Calculation using coefficients
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor
Quarter bridge: 120, 240, and 350 Ω
Half/full bridge: 120 to 1000 Ω
Bridge Excitation Constant-voltage bridge excitation: Approx. 2 VDC
Constant-current bridge excitation: Approx. 5.6 mA
(Bridge resistance 350 Ω)
Measuring Range At strain measurement
0 to ± 300 k ×10-6 strain (Constant-voltage bridge excitation)
0 to ± 20 k ×10-6 strain (Constant-current bridge excitation)
When measuring temperature with civil engineering
transducers with a thermal sensor: -30.0°C to 70.0°C
Resolution At strain measurement
0 to ± 30 k ×10-6 strain: 1 ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: 10 ×10-6 strain
When measuring temperature with civil engineering
transducers with a thermal sensor: 0.1°C
Accuracy At full bridge strain measurement
0 to ± 30 k ×10-6 strain: ± (0.05% of reading + 2) ×10-6 strain
± 30 k to ± 300 k ×10-6 strain: ± (0.1% of reading + 20) ×10-6 strain
When measuring temperature with civil engineering
transducers with a thermal sensor: ± 0.5°C
Check Functions Insulation resistance measurement: 2 to 100 M Ω
Resistance measurement: 0 to 20 kΩ
Interval Measurement 1 minute to 99 hours 59 minutes in 1-minute steps
Starting time: year/month/day/hour/minute
Storage SD card (Option)
SD Cards 256 MB, 512 MB, 1 GB, 2 GB (FAT16)
(SDHC and SDXC not supported)
Display Monochrome LCD, 128 × 64
TEDS Reads information from TEDS-installed sensors
Channel name writing
(Kyowa ID only within 10 characters)
Operating Temperature -10 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Power Supply AA battery × 2
Consecutive operation time:
Approx. 10 hours (With alkaline batteries, NTB not connected)
Nickel hydride batteries is also used.
An AC adapter (Optional, SW-0522E) is provided for SME-101A
Auto Power Off Power is automatically turned off if no key operation is detected for 5 minutes. In interval measuring mode with an interval of 3 minutes or longer, power is automatically turned off during standby period and
turned on again 1 minute before the next measurement is started. (ON/OFF of Auto Power Off is settable.)
Dimensions 188 W × 41 H x 108.4 D mm (Excluding protrusions)
Weight Approx. 450 g (Excluding batteries)
Standard Accessories Input cable U-119
Communication cable N-102
AA alkali battery × 2
Shoulder belt
Hand strap
Instruction manual (CD-R)
Optional Accessories AC adapter SW-0522E

Dimensions

SMET-1A
SME Thermocouple Adapter

Specifications

Measuring Targets Thermocouples (K, T)
Channels 1
Sampling Frequencies Approx. 0.5 s
Input Terminal block
Applicable wires
Solid wire: UL AWG14 to 28
Twisted wire: UL AWG20 to 24
Applicable Models Handy data logger SME-30A/31A
Handy data logger SME-100A/101A
Check Functions Burn out check (Operation on SME possible)
Operating Temperature -10 to 50ºC
Operating Humidity 20 to 85% (Non-condensing)
Dimensions 42 W x 33 H x 29.5 D mm (Excluding protrusions)
Weight Approx. 35 g
Measuring Range, Accuracy, Resolution K :
Temperature range: -200.0 to below -100.0°C
Accuracy with external standard junction compensation:±(0.2% of specified value + 0.6)°C
Accuracy under ambient temperature with internal standard junction compensation (25 ± 10)ºC: ±(0.2% of specified value + 2.6)°C
Resolution: 0.1°C

Temperature range: -100.0 to 1230.0°C or less
Accuracy with external standard junction compensation: ±(0.1% of specified value + 0.4)°C
Accuracy under ambient temperature with internal standard junction compensation (25 ± 10)ºC: ±(0.1% of specified value + 1.4)°C
Resolution: 0.1°C

T :
Temperature range: -200.0 to below -100.0°C
Accuracy with external standard junction compensation: ±(0.2% of specified value + 0.6)°C
Accuracy under ambient temperature with internal standard junction compensation25 ± 10)ºC: ±(0.2% of specified value + 2.6)°C
Resolution: 0.1°C

Temperature range: -100.0 to 400.0°C
Accuracy with external standard junction compensation: ±(0.1% of specified value + 0.4)°C
Accuracy under ambient temperature with internal standard junction compensation (25 ± 10)ºC: ±(0.1% of specified value + 1.4)°C
Resolution: 0.1°C
Notes : 1. Accuracy does not include the accuracy of the sensor.
2. For the standard junction compensator, switching between internal and external is possible using the SME.
3. Thermocouple resistance is 1 kΩ or less.
Standard Accessories Mounting screw×2
Instruction manual (CD-R)