UCAM-550A
Fast Data Logger

  • TEDS

Summary

· Synchronous sampling of all channels
· Synchronously sample at 50 Hz for 1000 channels
· 5 types of measuring units available

Features

UCAM-550A is a fast data logger that repeatedly measures a maximum of 1000 channels at an interval of 0.02 s.
Because it is capable of high-speed synchronous measurement, this unit measures a wide range of phenomena, from static to dynamic phenomena.

Block diagrams

UCAM-550A

Products table

All display

UCAM-550A
Fast Data Logger

Specifications

Models UCAM-550A: With DCS-100A
UCAM-550A-0: Without DCS-100A
Channels Maximum of 50 channels/unit
(Possible up to 5 units of the measuring unit)
(Each measuring unit measures 10 channels.)

Measurement is possible of up to 1000 channels at maximum by adding an optional software DCS-106A.
*The public command corresponds up to 20 units (Max. 1000 channels).
*DCS-100A corresponds up to 6 units (Max. 300 channels).
Sampling Method Synchronous sampling of all channels
Sampling Frequencies 1, 2, 10, 20, and 50 Hz

Response frequency depends on the measuring unit.
  USM-51B/52B*, USS-51B, USV-51B, UST-51B: DC to 7.8 Hz
  Deviation:0.5 to -3.5 dB
*For temperature measurement with USM-51B/52B using scanning mode, the updating rate is approx. 1 s.
Measuring Functions Original value measurement
Measure value measurement
Interfaces 10 BASE-T
100BASE-TX

[Between PC and UCAM]
 LAN cable (Straight)
 Max. 100 m

[Between UCAM and UCAM]
 STP straight cable (See Note.)
 Max. 100 m
Note: "STP" is the initials of Shield Twisted Pair, and an STP cable is a shielded LAN cable.
Display LCD (20 digits x 2 lines)

Status display LED:
 POWER (When power ON, lit green)
 MASTER (When master, lit green, when slave, not lit)
 TRANSFER (When communications, flashing green)
Operation Keys UP
Down
Left
Right
Data Storage Measurement data is saved on a PC. (No internal storage)
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 85% (Non-condensing)
Power Supply 100 to 240 VAC
Approx. 50 VA
(With five USS-51B strain units installed, and 120 Ω load on all channels connected)
Dimensions 426 W × 132.5 H ×305 D mm (Excluding protrusions)
Weight Approx. 7 kg (With five USS-51B strain units installed)
Standard Accessories AC power cable P-18 (With a 2-pin conversion plug CM-52)
Ground wire P-72
DVD (DCS-100A, instruction manual)
Dedicated Optional Accessories Strain/Voltage/Thermocouple Unit USM-51B/USM-52B
Strain Unit USS-51B
Voltage Unit USV-51B
Thermocouple Unit UST-51B
Connection Cable U-17 to 20
Isolation Transformer UPT-300BR
One-touch Terminal Block JT-1A
Dummy Panel UD-50A

Dimensions

UCAM-550A

USM-51B
Strain/Voltage/Thermocouple Unit

USM-51B
  • TEDS

Specifications

Input Terminals NDIS 4102 (7 pins) connectors, and screw-soldering terminal blocks
Channels 10
Measuring Targets Strain gages, strain-gage transducers, potentiometer-type sensors,
voltage, and thermocouples
Bridge Excitation 2 VDC
Power Supply to Sensors 2 VDC, for potentiometer-type sensors
Gage Factor 2.00 fixed
Frequency Response DC to 7.8 Hz, deviation: 0.5, -3.5 dB
(Except temperature measurement)
Burnout Check Performing burnout when checking
TEDS Reads information from TEDS-installed sensor.
Strain, Potentiometers, and Voltage [Strain]
Mode: L
Measuring Range:0 to ±19 k×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±0.08%FS

[Strain]
Mode: H
Measuring Range: 0 to ±300 k×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±0.08%FS

[Potentiometers]
Measuring Range: -50% to 50%
Resolution: 0.01%
Accuracy: ±0.1%FS

[Voltage]
Measuring Range: -20 to 20 V
Resolution: 1 mV
Accuracy: ±0.08%FS
Thermocouples [Types K]
Range:-200.0 to 1200.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 1200.0ºC ±(0.2% of reading +0.6ºC)

[Types T]
Range: -200.0 to 350.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 350.0ºC ±(0.2% of reading +0.6ºC)


[Types E]
Range: -200.0 to 800.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 800.0ºC ±(0.2% of reading +0.6ºC)


[Types J]
Range:-200.0 to 750.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 750.0ºC ±(0.2% of reading +0.6ºC)


[Types R]
Range: 0.0 to 1600.0ºC
Accuracy:
 0.0 to below 100.0ºC ±(0.6% of reading +1.2ºC)
 100.0 to 1600.0ºC ±(0.5% of reading +1.0ºC)


[Types N]
Range: -200.0 to 1250.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -200.0 to 1250.0ºC ±(0.2% of reading +0.6ºC)

* Accuracy of the Internal Reference-junction Compensator
Within ±1.0°C
 (When temperature balanced at input terminals)
 (The ambient temperature is 25 ±10°C)
Within ±2.0°C
 (When temperature balanced at input terminals)
 (The ambient temperature is other than mentioned above.)
Standard Accessories Terminal cover UM-51B

USM-52B
Strain/Voltage/Thermocouple Unit

USM-51B
  • TEDS

Specifications

Input Terminals NDIS4102 (7 pins) connectors, and one-touch terminal blocks
Channels 10
Measuring Targets Strain gages, strain-gage transducers, potentiometer-type sensors,
voltage, and thermocouples
Bridge Excitation 2 VDC
Power Supply to Sensors 2 VDC, for potentiometer-type sensors
Gage Factor 2.00 fixed
Frequency Response DC to 7.8 Hz, deviation: 0.5, -3.5 dB
(Except temperature measurement)
Burnout Check Performing burnout when checking
TEDS Reads information from TEDS-installed sensor.
Strain, Potentiometers, and Voltage [Strain]
Mode: L
Measuring Range:0 to ±19 k×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±0.08%FS

[Strain]
Mode: H
Measuring Range: 0 to ±300 k×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±0.08%FS

[Potentiometers]
Measuring Range: -50% to 50%
Resolution: 0.01%
Accuracy: ±0.1%FS

[Voltage]
Measuring Range: -20 to 20 V
Resolution: 1 mV
Accuracy: ±0.08%FS
Thermocouples [Types K]
Range:-200.0 to 1200.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 1200.0ºC ±(0.2% of reading +0.6ºC)

[Types T]
Range: -200.0 to 350.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 350.0ºC ±(0.2% of reading +0.6ºC)


[Types E]
Range: -200.0 to 800.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 800.0ºC ±(0.2% of reading +0.6ºC)


[Types J]
Range:-200.0 to 750.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -100.0 to 750.0ºC ±(0.2% of reading +0.6ºC)


[Types R]
Range: 0.0 to 1600.0ºC
Accuracy:
 0.0 to below 100.0ºC ±(0.6% of reading +1.2ºC)
 100.0 to 1600.0ºC ±(0.5% of reading +1.0ºC)


[Types N]
Range: -200.0 to 1250.0ºC
Accuracy:
 -200.0 to below -100.0ºC ±(0.3% of reading +0.8ºC)
 -200.0 to 1250.0ºC ±(0.2% of reading +0.6ºC)

* Accuracy of the Internal Reference-junction Compensator
Within ±1.0°C
 (When temperature balanced at input terminals)
 (The ambient temperature is 25 ±10°C)
Within ±2.0°C
 (When temperature balanced at input terminals)
 (The ambient temperature is other than mentioned above.)
Standard Accessories Terminal cover UM-51B

USS-51B
Strain Unit

USS-51B

Specifications

Channels 10
Measuring Targets Strain gage, strain-gage transducers, potentiometer-type sensors
Bridge Excitation 2 VDC constant voltage (Applied constantly)
Power Supply to Sensors 2 VDC constant voltage (Applied constantly)
Gage Factor 2.00 fixed
(Correction is possible at 2.00/Ks with the engineering value conversion function)
Measuring Range, Resolution, Accuracy (In static (DC) Inputting) [Strain]
Mode: L
Measuring Range: 0 to ±19 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±0.05% FS

[Strain]
Mode: H
Measuring Range: 0 to ±200 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±0.05% FS

[Potentiometers]
Measuring Range: 0 to ±50%
Resolution: 0.01%
Accuracy: ±0.1% FS

Note:
Measuring range is indicated when the initial measurement and the original value measurement are performed. In the case of a measure value measurement, the value of the initial measurement is subtracted in advance from the original measurement value.
Optional Accessories Terminal cover UT-50A

UST-51B
Thermocouple Unit

UST-51B

Specifications

Channels 10
Measuring Targets Temperature (Thermocouples)
Measuring Range, Resolution, Accuracy (In static (DC) Inputting) [Type K]
Mode: L
Measuring Range: -200.0 to 437.0ºC
Accuracy: ±0.8ºC

[Type K]
Mode: H
Measuring Range: -200.0 to 1200.0ºC
Accuracy: ±2.8ºC

[Type T]
Measuring Range: -200.0 to 350.0ºC
Accuracy: ±0.7ºC

[Type E]
Mode: L
Measuring Range: -200.0 to 260.0ºC
Accuracy: ±0.5ºC

[Type E]
Mode: H
Measuring Range: -200.0 to 800.0ºC
Accuracy: ±1.7ºC

[Type J]
Mode: L
Measuring Range: 0 to 330.0ºC
Accuracy: ±0.6ºC

[Type J]
Mode: H
Measuring Range: 0 to 750.0ºC
Accuracy: ±2.0ºC

[Type R]
Measuring Range: 0 to 1600.0ºC
Accuracy: ±2.2ºC

[Type N]
Mode: L
Measuring Range: -200.0 to below -100.0°C
Accuracy: ±(0.4% of reading +1.0ºC)
Measuring Range: -100 to 530.0°C
Accuracy: ±(0.3% of reading +0.8ºC)

[Type N]
Mode: H
Measuring Range: -200.0 to below -100.0°C
Accuracy: ±(0.4% of reading +1.2ºC)
Measuring Range: -100 to 1250.0°C
Accuracy: ±(0.3% of reading +1.0ºC)

*When temperature balanced at input terminals, and the ambient temperature is 25 ±10ºC.
Type K, T, E, J, and R: Within ±0.5ºC
Type N: Within ±1.0ºC

Note:
Accuracy does not include internal reference junction accuracy.
Switching between internal and external standard connect compensators is possible.
Thermocouple resistance 300 Ω or less (K type).

Standard Accessories Terminal cover UT-50A

USV-51B
Voltage Unit

USV-51B

Specifications

Channels 10
Measuring Targets DC voltage, voltage output type sensors
Measuring Range, Resolution, Accuracy (In static (DC) Inputting) Measuring Range: 0 to ±20.000 V
Resolution: 1 mV
Accuracy: ±0.05% FS
Signal Source Resistance: 50 Ω or less
Standard Accessories Terminal cover UT-50A

Page Top

To Ensure Safe Usage

DCS-100A, standard accessory, can measure up to 300 channels.
Measurement up to 1000 channels requires an optional software DCS-106A.