UCAM-60 Series
Data Logger

UCAM-60B/65B

Summary

· High resolution 0.1μm/m, max measuring channels 1000
· All-in-one static strain Data Logger UCAM-60B
· Online static Data Logger UCAM-65B

Features

The UCAM-60B is an all-in-one Data Logger, which has easy to operate keys, easy to visual display and a printer for immediate confirmation of measurement results.

Products table

All display

UCAM-60B
All-in-one Data Logger

UCAM-60B

Specifications

Models UCAM-60B-AC AC only
UCAM-60B-DC DC only
Measuring Targets Strain gages, strain-gage transducers, DC voltage-output or DC currentoutput instruments, civil engineering transducers with a thermal sensor, potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Channels Max. 30 with dedicated scanners
Max. 1000 with external scanners connected
Max. 1000 with dedicated scanners + external scanners connected.
Scanning Speed 50 ms/channel (Standard mode)
280 ms/channel (High-resolution mode)
* Individually switchable for desired channels.
20 ms/channel (High-speed mode)
* Collectively switchable for all channels of dedicated scanners.
Note : Scanning speeds stated above are standard maximum speeds in respective modes. Besides these, the following speeds are set for
each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s
Switching Terminal Semiconductor relay
Operating Modes Real-time, monitor, and automatic
Measurement Functions
Initial Initial values are measured and stored in internal memory
(Except for temperatures measured by civil engineering transducers with a thermal sensor).
Original Raw values are measured without subtraction of initial values.
Measure Initial values are subtracted from original values
(Except for temperatures measured by civil engineering transducers with a thermal sensor).
Easy Measure Auto zero balancing function is activated.
* The selected function is applied to all channels.
Calculation Functions Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.
Engineering Units 59 units
Automatic Measurement Functions
Trigger Measurement A relative value (Certain changing quantity) or an absolute value triggers measurement. In addition to the usual trigger function, a variable trigger function is provided with which the trigger value changes at each step during measurement.
With this special function, a trigger value and the measurement times (Repeat times) under the trigger condition is registered for each step to perform a series of automatic measurements in the order of steps.
The maximum steps available for setting is 15 and the repeat times may be a value selected from a range of 1 to 9999 or infinite.
Trigger channels: 1 desired channel
Trigger value: A desired real number of 6 effective figures or less
Reference value: Amount of level shift to determine the first trigger value
(Selected from the same range as for the trigger values)
Repeat times: 1 to 9999 (0 for infinite times)
Measuring steps: Max. 15
Interval Measurement Measurement is automatically performed at preset time intervals.
Starting time setting: Year, month, day, hour, minute, and second
Interval setting: Days and hours: minutes: seconds
(In a range of 00 and 00:00:01 to 99 and 12:59:59)
Repeat times: 1 to 9999 (0 for infinite times)
Max. steps: 15
Trigger Interval Measurement Combination of trigger measurement and interval measurement.
Trigger value: Absolute value.
Measuring times: Max. 9999
Interval time: Available in a range of
1 second to 99 days 23 hours:59 minutes:59 seconds
Storage Internal memory, approx. 7 MB
Flash ATA card (Optional); the capacity depends on the card.
Strain Measurement (Standard Mode)
Bridge Excitation
Constant Voltage Excitation Approx. 2 or 5 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω)
(Up to 5 km with a 4-conductor (0.5 mm2) shielded cable)
Approx. 16.7 mA (Bridge resistance 120 Ω)
(Up to 2 km with a 4-conductor (0.5 mm2) shielded cable) 
Scanning Speed 50 ms/channel
Gage Factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±50 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.05% of reading + 1) ×10-6 strain

Measuring Range: ±50 k to 500 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±(0.05% of reading + 10) ×10-6 strain
Strain Measurement (High-resolution Mode)
Bridge Excitation
Constant Voltage Excitation Approx. 5 VDC
Constant Current Excitation Approx. 16.7 mA (Bridge resistance 350 Ω)
(Up to 2 km with a 4-conductor (0.5 mm2) shielded cable)
Scanning Speed 280 ms/channel
Gage Factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±20 k ×10-6 strain
Resolution: 0.1 ×10-6 strain
Accuracy: ±(0.05% of reading + 0.3)×10-6 strain

Measuring Range: ±20 k to 200 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.05% of reading + 3)×10-6 strain
Note : 1. Available only with full bridge system (120 to 1000 Ω)
2. Bridge resistance should be 350 Ω for bridge excitation with constant current.
3. Measuring range is 0 to 15 k ×10-6 strain for bridge excitation with constant current.
4. Available only with dedicated scanners.
Strain Measurement (High-speed Mode)
Bridge Excitation
Constant Voltage Excitation Approx. 2 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω )
(Up to 5 km with a 4-conductor (0.5 mm2) shielded cable)
Approx. 16.7 mA (Bridge resistance 120 Ω )
(Up to 2 km with a 4-conductor (0.5 mm2) shielded cable) 
Scanning Speed 20 ms/channel
Gage Factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±50 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.08% of reading + 3) ×10-6 strain
Measuring Range: ±50 k to 500 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±(0.08% of reading + 30) ×10-6 strain
Notes : 1. Available only with full bridge system
2. Available only with dedicated scanners.
Voltage Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Range Mode:V/500 mV
Measuring Range:0 to ±50.000 mV
Resolution:1 μV
Accuracy:±(0.05% of reading + 3) mV
Input Resistance:10 M Ω or more
Range Mode:V/500 mV
Measuring Range:±50.00 to 500.00 mV
Resolution:10 μV
Accuracy:±(0.05% of reading + 3) mV
Input Resistance:10 M Ω or more
Range Mode:V/50 V
Measuring Range:0 to ±5.0000 V
Resolution:100 μV
Accuracy±(0.05% of reading + 2) V
Input Resistance:1 M Ω or more
Range Mode:V/50 V
Measuring Range:±5.000 to 50.000 V
Resolution:1 mV
Accuracy±(0.05% of reading + 2) V
Input Resistance:1 M Ω or more
Voltage Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Range Mode:V/500 mV
Measuring Range:0 to ±50.000 mV
Resolution:1 μV
Accuracy:±(0.08% of reading + 6) mV
Input Resistance:10 M Ω or more
Range Mode:V/500 mV
Measuring Range:±50.00 to 500.00 mV
Resolution:10 μV
Accuracy:±(0.08% of reading + 6) mV
Input Resistance:10 M Ω or more
Range Mode:V/50 V
Measuring Range:0 to ±5.0000 V
Resolution:100 μV
Accuracy±(0.08% of reading + 6) V
Input Resistance:1 M Ω or more
Range Mode:V/50 V
Measuring Range:±5.000 to 50.000 V
Resolution:1 mV
Accuracy±(0.08% of reading + 6) V
Input Resistance:1 M Ω or more
Current Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode:1/50 mA
Measuring Range:0 to ±50.00 mA
Resolution:10 μA
Accuracy:±(0.05% of reading + 0.01) mA
Notes 1. External shunt resistor (High-accuracy 250 Ω ) is required.
2. Stated accuracy does not include the external shunt resistor.
Current Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode:1/50 mA
Measuring Range: 0 to ±50.00 mA
Resolution: 10 μA
Accuracy: ±(0.08% of reading + 0.01) mA
Notes : 1. Available only with dedicated scanners.
2. External shunt resistor (High-accuracy 250 Ω) is required.
3. Stated accuracy does not include the external shunt resistor accuracy.
Temperature Measurement with Thermocouples (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy K
Measuring Range:-200.0 to 1230.0 °C
Resolution: 0.1 ºC
Accuracy: ±0.7°C
T
Measuring Range: -200.0 to 400.0 °C
Resolution: 0.1 ºC
Accuracy: ±0.7°C
E
Measuring Range: -200.0 to 660.0 ºC
Resolution: 0.1 ºC
Accuracy: ±0.5°C
J
Measuring Range:-200.0 to 870.0 °C
Resolution: 0.1 ºC
Accuracy: ±0.6°C
R
Measuring Range: -0 to 1760.0 °C
Resolution: 0.1 ºC
Accuracy: ±2.2°C
Notes : 1. Accuracies do not include the internal reference junction
compensator accuracy.
2. The reference junction compensator is switchable between internal and external.
3. Thermocouple resistance should be 1 kΩ or less.
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy Measuring Range:-50.0 to 200.0°C
Resolution:0.1°C
Accuracy:±0.5°C
Notes 1. Target physical quantity and temperature are measured in a single channel.
2. Strain measuring ranges are the same as in strain measurement in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy Type:Pt100
Measuring Range: -200.0 to 660.0 ºC
Resolution: 0.1 ºC
Accuracy: ±0.3 ºC
Type:JPt100
Measuring Range: -200.0 to 510.0 ºC
Resolution: 0.1 ºC
Accuracy: ±0.3 ºC
Note : Connection is 3-wire system
Measurement with Potentiometer Sensor
Scanning Speed 50 ms/channel (Standard mode)
20 ms/channel (High-speed mode)
Initial Value Memory Range Same as measuring range
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 kΩ
Measuring Range, Resolution and Accuracy Channel Mode: POT.
Measuring Range: 0 to ±50.00%
Resolution: 0.01%
Accuracy: ±0.1% FS
Clock Real-time clock is built in (Battery backup 5 years).
Display Fluorescent display tube 128 x 64
Printing Thermal
Paper width 58 mm (24 characters/line),UCAM-60A-RP
Printing speed 60 mm/s
PC Card Slot Conforms with PCMCIA Ver. 4.2.
Accepts a commercially available flash ATA card or CF card (ATA card adapter required).
Interfaces RS-232C and LAN (10BASE-T/100BASE-TX)
File Conversion CSV conversion
Self-diagnosis Functions Checks display, printer, bridge excitation, disconnection, I/O resistance, insulation resistance, mode, etc. Checking of I/O resistance and mode is available only for dedicated scanners.
TEDS Interfaces: IEEE 1451.4 Mixed Mode Transducer Interface Class 2
Applicable sensor: Should have information written in accordance with IEEE template No. 33; cable length should be 30 m or less.
*With the scanning unit USS-61B/62B/63B.
Operating Temperature 0 to 50 ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Power Supply 100 to 240 VAC (AC-operated version)
10 to 16 VDC (DC-operated version)
* DC-operated version has power control function.
Current Consumption 0.5 A or less: 100 VAC (With 3 dedicated scanners mounted)
4 A or less: 12 VDC (With 3 dedicated scanners mounted)
Dimensions 360 W x 88 H x 400 D mm (Excluding protrusions)
Weight Approx. 6.3 kg (Excluding scanner)
Standard Accessories AC power cable P-18 (With 2-pin conversion plug CM-39) (AC-operated version), DC power cable P-76 (DC-operated version), recording paper UCAM-60A-RP (1 roll for UCAM-60B only), screwdriver, spare fuse, CD-R (Instruction Manual)
Optional Accessories Recording Paper UCAM-60A-RP (10 rolls/pack)

Dimensions

UCAM-60B

UCAM-65B
Online static Data Logger

Specifications

Models UCAM-65B-AC AC only
UCAM-65B-DC DC only
Measuring Targets Strain gages, strain-gage transducers, DC voltage-output or DC currentoutput instruments, civil engineering transducers with a thermal sensor, potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Channels Max. 30 with dedicated scanners
Max. 1000 with external scanners connected
Max. 1000 with dedicated scanners + external scanners connected.
Scanning Speed 50 ms/channel (Standard mode)
280 ms/channel (High-resolution mode)
* Individually switchable for desired channels.
20 ms/channel (High-speed mode)
* Collectively switchable for all channels of dedicated scanners.
Note : Scanning speeds stated above are standard maximum speeds in respective modes. Besides these, the following speeds are set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s
Switching Terminal Semiconductor relay
Operating Modes Real-time, monitor, and automatic
Measurement Functions
Initial Initial values are measured and stored in internal memory
(Except for temperatures measured by civil engineering transducers with a thermal sensor).
Original Raw values are measured without subtraction of initial values.
Measure Initial values are subtracted from original values
(Except for temperatures measured by civil engineering transducers with a thermal sensor).
Easy Measure Auto zero balancing function is activated.
* The selected function is applied to all channels.
Calculation Functions Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.
Engineering Units 59 units
Automatic Measurement Functions
Trigger Measurement A relative value (Certain changing quantity) or an absolute value triggers measurement. In addition to the usual trigger function, a variable trigger function is provided with which the trigger value changes at each step during measurement.
With this special function, a trigger value and the measurement times (Repeat times) under the trigger condition is registered for each step to perform a series of automatic measurements in the order of steps.
The maximum steps available for setting is 15 and the repeat times may be a value selected from a range of 1 to 9999 or infinite.
Trigger channels: 1 desired channel
Trigger value: A desired real number of 6 effective figures or less
Reference value: Amount of level shift to determine the first trigger value
(Selected from the same range as for the trigger values)
Repeat times: 1 to 9999 (0 for infinite times)
Measuring steps: Max. 15
Interval Measurement Measurement is automatically performed at preset time intervals.
Starting time setting: Year, month, day, hour, minute, and second
Interval setting: Days and hours: minutes: seconds (In a range of 00 and 00:00:01 to 99 and 12:59:59)
Repeat times: 1 to 9999 (0 for infinite times)
Max. steps: 15
Trigger Interval Measurement Combination of trigger measurement and interval measurement.
Trigger value: Absolute value.
Measuring times: Max. 9999
Interval time: Available in a range of
1 second to 99 days 23 hours:59 minutes:59 seconds
Storage Internal memory, approx. 7 MB
Flash ATA card (Optional); the capacity depends on the card.
Strain Measurement (Standard Mode)
Bridge Excitation
Constant Voltage Excitation Approx. 2 or 5 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω)
(Up to 5 km with a 4-conductor (0.5 mm2) shielded cable)
Approx. 16.7 mA (Bridge resistance 120 Ω)
(Up to 2 km with a 4-conductor (0.5 mm2) shielded cable) 
Scanning Speed 50 ms/channel
Gage Factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±50 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.05% of reading + 1) ×10-6 strain

Measuring Range: ±50 k to 500 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±(0.05% of reading + 10) ×10-6 strain
Strain Measurement (High-resolution Mode)
Bridge Excitation
Constant Voltage Excitation Approx. 5 VDC
Constant Current Excitation Approx. 16.7 mA (Bridge resistance 350 Ω)
(Up to 2 km with a 4-conductor (0.5 mm2) shielded cable)
Scanning Speed 280 ms/channel
Gage Factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±20 k ×10-6 strain
Resolution: 0.1 ×10-6 strain
Accuracy: ±(0.05% of reading + 0.3)×10-6 strain

Measuring Range: ±20 k to 200 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.05% of reading + 3)×10-6 strain
Note : 1. Available only with full bridge system (120 to 1000 Ω)
2. Bridge resistance should be 350 Ω for bridge excitation with constant current.
3. Measuring range is 0 to 15 k ×10-6 strain for bridge excitation with constant current.
4. Available only with dedicated scanners.
Strain Measurement (High-speed Mode)
Bridge Excitation
Constant Voltage Excitation Approx. 2 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω )
(Up to 5 km with a 4-conductor (0.5 mm2) shielded cable)
Approx. 16.7 mA (Bridge resistance 120 Ω )
(Up to 2 km with a 4-conductor (0.5 mm2) shielded cable) 
Scanning Speed 20 ms/channel
Gage Factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±50 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.08% of reading + 3) ×10-6 strain
Measuring Range: ±50 k to 500 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±(0.08% of reading + 30) ×10-6 strain
Notes : 1. Available only with full bridge system
2. Available only with dedicated scanners.
Voltage Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Range Mode:V/500 mV
Measuring Range:0 to ±50.000 mV
Resolution:1 μV
Accuracy:±(0.05% of reading + 3) mV
Input Resistance:10 M Ω or more
Range Mode:V/500 mV
Measuring Range:±50.00 to 500.00 mV
Resolution:10 μV
Accuracy:±(0.05% of reading + 3) mV
Input Resistance:10 M Ω or more
Range Mode:V/50 V
Measuring Range:0 to ±5.0000 V
Resolution:100 μV
Accuracy±(0.05% of reading + 2) V
Input Resistance:1 M Ω or more
Range Mode:V/50 V
Measuring Range:±5.000 to 50.000 V
Resolution:1 mV
Accuracy±(0.05% of reading + 2) V
Input Resistance:1 M Ω or more
Voltage Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Range Mode:V/500 mV
Measuring Range:0 to ±50.000 mV
Resolution:1 μV
Accuracy:±(0.08% of reading + 6) mV
Input Resistance:10 M Ω or more
Range Mode:V/500 mV
Measuring Range:±50.00 to 500.00 mV
Resolution:10 μV
Accuracy:±(0.08% of reading + 6) mV
Input Resistance:10 M Ω or more
Range Mode:V/50 V
Measuring Range:0 to ±5.0000 V
Resolution:100 μV
Accuracy±(0.08% of reading + 6) V
Input Resistance:1 M Ω or more
Range Mode:V/50 V
Measuring Range:±5.000 to 50.000 V
Resolution:1 mV
Accuracy±(0.08% of reading + 6) V
Input Resistance:1 M Ω or more
Current Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode:1/50 mA
Measuring Range: 0 to ±50.00 mA
Resolution: 10 μA
Accuracy: ±(0.08% of reading + 0.01) mA
Notes 1. External shunt resistor (High-accuracy 250 Ω ) is required.
2. Stated accuracy does not include the external shunt resistor.
Current Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode:1/50 mA
Measuring Range: 0 to ±50.00 mA
Resolution: 10 μA
Accuracy: ±(0.08% of reading + 0.01) mA
Notes : 1. Available only with dedicated scanners.
2. External shunt resistor (High-accuracy 250 Ω) is required.
3. Stated accuracy does not include the external shunt resistor accuracy.
Temperature Measurement with Thermocouples (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy K
Measuring Range:-200.0 to 1230.0 °C
Resolution: 0.1 ºC
Accuracy: ±0.7°C
T
Measuring Range: -200.0 to 400.0 °C
Resolution: 0.1 ºC
Accuracy: ±0.7°C
E
Measuring Range: -200.0 to 660.0 ºC
Resolution: 0.1 ºC
Accuracy: ±0.7°C
J
Measuring Range:-200.0 to 870.0 °C
Resolution: 0.6 ºC
Accuracy: ±0.7°C
R
Measuring Range: -0 to 1760.0 °C
Resolution: 0.1 ºC
Accuracy: ±2.2°C
Notes : 1. Accuracies do not include the internal reference junction
compensator accuracy.
2. The reference junction compensator is switchable between internal and external.
3. Thermocouple resistance should be 1 kΩ or less.
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy Measuring Range:-50.0 to 200.0°C
Resolution:0.1°C
Accuracy:±0.5°C
Notes : 1. Target physical quantity and temperature are measured in a single channel.
2. Strain measuring ranges are the same as in strain measurement in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy Pt100
Measuring Range: -200.0 to 660.0 ºC
Resolution: 0.1 ºC
Accuracy: ±0.3 ºC
JPt100
Measuring Range: -200.0 to 510.0 ºC
Resolution: 0.1 ºC
Accuracy: ±0.3 ºC
Note : Connection is 3-wire system
Measurement with Potentiometer Sensor
Scanning Speed 50 ms/channel (Standard mode)
20 ms/channel (High-speed mode)
Initial Value Memory Range Same as measuring range
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 kΩ
Measuring Range, Resolution and Accuracy Channel Mode: POT.
Measuring Range: 0 to ±50.00%
Resolution: 0.01%
Accuracy: ±0.1% FS
PC Card Slot Conforms with PCMCIA Ver. 4.2.
Accepts a commercially available flash ATA card or CF card (ATA card adapter required).
Interfaces RS-232C and LAN (10BASE-T/100BASE-TX)
File Conversion CSV conversion
Self-diagnosis Functions Checks display, printer, bridge excitation, disconnection, I/O resistance, insulation resistance, mode, etc. Checking of I/O resistance and mode is available only for dedicated scanners.
TEDS Interfaces: IEEE 1451.4 Mixed Mode Transducer Interface Class 2
Applicable sensor: Should have information written in accordance with IEEE template No. 33; cable length should be 30 m or less.
*With the scanning unit USS-61B/62B/63B.
Operating Temperature 0 to 50 ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Power Supply 100 to 240 VAC (AC-operated version)
10 to 16 VDC (DC-operated version)
* DC-operated version has power control function.
Current Consumption 0.5 A or less: 100 VAC (With 3 dedicated scanners mounted)
4 A or less: 12 VDC (With 3 dedicated scanners mounted)
Dimensions 327 W x 88 H x 365 D mm (Excluding protrusions)
Weight Approx. 5.0 kg
Standard Accessories AC power cable P-18 (With 2-pin conversion plug CM-39) (ACoperated version), DC power cable P-76 (DC-operated version), recording paper UCAM-60A-RP (1 roll for UCAM-60B only), screwdriver, spare fuse, CD-R (Instruction Manual)CD-R (Control
software UCS-60B)

Dimensions

UCAM-65B

USS-61B
Dedicated Scanner (TEDS compatible)

USS-61B

Specifications

Model TEDS compatible
Channels 10/Unit
Switching Terminals Semiconductor relays
Input Terminals Connect to leadwires through either soldering or screwing.
One-touch terminal block JT-1A (Optional)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 80%RH (Non-condensing)
Dimensions 320 W x 28 H x 80 D mm (Excluding protrusions)
Weight Approx800g(Including terminal cover)
Standard Accessories Terminal cover, Channel label

USS-62B
Dedicated Scanner (With NDIS connectors, TEDS compatible)

USS-62B

Specifications

Model With NDIS connectors,TEDS compatible
Channels 10/Unit
Switching Terminals Semiconductor relays
Input Terminals Connect to leadwires through either soldering or screwing NDIS connectors One-touch terminal block JT-1A (Optional)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 80%RH (Non-condensing)
Dimensions 320 W x 28 H x 80 D mm (Excluding protrusions)
Weight Approx. 1 kg(Including terminal cover)
Standard Accessories NDIS connector caps (Pre-attached to connectors),
terminal cover, channel label

USS-63B
Dedicated Scanner (For civil engineering measurement,TEDS compatible, with lightning arresters)

USS-63B

Specifications

Model For civil engineering measurement, TEDS compatible, with lightning Arresters
Channels 10/Unit
Switching Terminals Semiconductor relays
Input Terminals Connect to lead wire by either soldering or screwing.
One-touch terminal block JT-1A (Optional)
Lightning Arresters Built
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Dimensions 320 W x 28 H x 80 D mm (Excluding protrusions)
Weight Approx. 900 g (Including terminal cover)
Standard Accessories Terminal cover, Channel label

USI-67A
Scanner Interfaces

Specifications

Connectable Scanners USB-70B, USB-65A
Number of Scanners Max. 20
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85%RH (Non-condensing)
Dimensions 99 W x 50 H x 163 D mm (Excluding protrusions)
Weight Approx. 160 g

UIO-60A
External I/O Unit

UIO-60A

Specifications

Output
ALARM Signal 4 channels (High/low limit checking)
BUSY Signal 1 channel
Input
START Signal 1 channel
STOP Signal 1 channel
RESET Signal 1 channel
RAINFALL Signals 1 channel
Operating Temperature 0 to 50 ºC
Operating Humidity 20% to 85%RH (Non-condensing)
Dimensions 90 W x 50 H x 180 D mm (Excluding protrusions)
Weight Approx. 140 g

USB-70B-10
External Scanner (for general strain measurement)

USB-70B-10

Specifications

Channels 50/unit
Measuring Channel Mode Selected for each channel from the mainframe
Measuring Targets Strain gages, strain-gage transducers,potentiometer, DC voltage-output instruments,thermocouples
Power Supply Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into scanners.
UPS-70B operates on 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85%RH (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx7.3 kg
Standard Accessories Connection cable N-24 (1 m)

USB-70B-20
External Scanner (for general strain measurement, with NDIS connectors)

Specifications

Channels 50/unit
Measuring Channel Modes Selected for each channel from the mainframe
Measuring Targets Strain gages, strain-gagee transducers, potentiometer,DC voltage-output instruments, thermocouples
(Transducer with NDIS connector is required)
Power Supply Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into scanners.
UPS-70B operates on 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx. 8.5 kg
Standard Accessories Connection cable N-24 (1 m)

USB-70B-30
External Scanner(for civil engineering, with lightning arresters)

USB-70B

Specifications

Channels 50/unit
Measuring Channel Mode Selected for each channel from the mainframe
Measuring Targets Strain gages, strain-gage transducers, potentiometer,DC voltage-output instruments, thermal sensors
(Thermocouples, platinum resistance thermometer bulbs, civil engineering transducers with a thermal sensor), lightning arresters built in
Power Supply Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into scanners.
UPS-70B operates on 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx7.7 kg
Standard Accessories Connection cable N-24 (1 m)

USB-65A
External Scanners

Specifications

Measuring Targets
Connectable Data Loggers UCAM-60B, 65B*Dedicated Scanners USI-67A is required, Connect with USB-70B is possible.
Channels 50/unit
Input Terminals Can connect to leadwires through either soldering
or screwing.
Switching Terminals Semiconductor relays
Differential Transformer Types Transducers
Sacnning Speed 1s/channel
Power Supply to Sensors 500 Hz 1 Vrms sine wave
Measuring Range, Resolution, Accuracy Measuring range: 0 to ±1000.00
Resolution: 0.01
Accuracy: ±1%FS as follows
SlidingResistance Types Transducers
Sacnning Speed 50 ms/channel
Power Supply to Sensors 2 VDC (From the unit base)
Measuring Range, Resolution, Accuracy Measuring range: 0 to ±100%
Resolution: 0.01%
Accuracy: ±1%Fs as follows
Power Supply Supplied from data logger If the cable is extended
or if 4 or more scanners are connected, an optional
UPS-70B should be mounted into scanners.
UPS-70B operates on 90 to 132/180 to 264 VAC
(No switchover required), 50/60 Hz
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85%RH (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx. 10 kg