UCAM-60C M14, UCAM-65C M14
Data Logger

Summary
・Saving the long-term saving measured results to the built-in memory (Approx. 1.8 GB)
・USB memory ensures easy data collection
・Up to 1000 channels measurement
(With external scanners USB-70B series)
・Measurement up to 20000 ×10-6 strain with a resolution of 0.1 ×10-6 strain (With full bridge system)
・All-in-one measuring instrument (UCAM-60C M14)
・Online data logger fully controlled from the PC (UCAM-65C M14)
Features
The data logger UCAM-60C M14 is an all-in-one measuring instrument developed in full pursuit of easier field measurement.
It has easy-to-operate keys, a bright readable display providing understandable presentation and a printer for immediate confirmation of measurement results.
Contents
【Data Loggers】
UCAM-60C-AC M14 (AC, UCS-60B: Optional, operation keys, built-in display, printer)
UCAM-60C-DC M14 (DC, UCS-60B: Optional, operation keys, built-in display, printer)
UCAM-65C-AC M14 (AC, UCS-60B: Standard, PC-controlled)
UCAM-65C-AC-0 M14 (AC, UCS-60B: Optional, PC-controlled)
UCAM-65C-DC M14 (DC, UCS-60B: Standard, PC-controlled)
UCAM-65C-DC-0 M14 (DC, UCS-60B: Optional, PC-controlled)
【Dedicated Scanners (Optional) 】
USS-61B (TEDS compatible) *
The dedicated scanner measures 10 channels/unit.
The main unit accommodates up to 3 dedicated scanners.
USS-62B (With NDIS4102 (7 pins) connectors, TEDS compatible) *
The dedicated scanner measures 10 channels/unit.
The main unit accommodates up to 3 dedicated scanners.
USS-63B (For civil engineering, with lightning arresters, TEDS compatible) *
The dedicated scanner measures 10 channels/unit.
The main unit accommodates up to 3 dedicated scanners.
* TEDS compatible function is made effective by connecting TEDS installed sensor through NDIS4102 (7 pins) connector.
【External Scanners】
The main unit is connected to the following scanners via the optional scanner interface.
USB-70B (Via scanner interface USI-67A)
【Scanner Interfaces 】
USI-67A for USB-70B
【External I/O Unit】
UIO-60A
【Control Software】
UCS-60B
Products table
All display
UCAM-60C M14
Data Logger

Specifications
Measuring Targets | Strain gages Strain-gage transducers DC voltage-output instruments DC current-output instruments Civil engineering transducers with a thermal sensor Potentiometer sensors Thermal sensors (Thermocouples and platinum resistance thermometer bulbs) |
---|---|
Connectable Scanners | USS-61B, USS-62B, USS-63B (Dedicated scanners, mounted on top of the UCAM-60C M14) The main unit is connected to the following scanners via the optional scanner interface (USI-67A). USB-70B series (via USI-67A) |
Channels | Max. 30 with dedicated scanners Max. 1000 with external scanners connected Max. 1000 with dedicated scanners and external scanners connected |
Input Terminals | Can connect to lead wires through either soldering or screwing. (Screw: M3 with metal plate) NDIS4102 (7 pins) connectors (USS-62B) |
Switching Terminal | Semiconductor relays |
Scanning Speed | 50 ms/channel (Standard mode) 0.28 s/channel (High-resolution mode) 20 ms/channel (High-speed mode) -------Scanners--------------------Line Frequencies------------ Dedicated scanner (Standard mode) 50 Hz Zone: 50 ms/channel 60 Hz Zone: 50 ms/channel Dedicated scanner (High-resolution mode) 50 Hz Zone: 0.28 s/channel 60 Hz Zone: 0.28 s/channel Dedicated scanner (High-speed mode) 50 Hz Zone: 20 ms/channel 60 Hz Zone: 20 ms/channel USB-70B (Standard mode only) 50 Hz Zone: 60 ms/channel 60 Hz Zone: 58.4 ms/channel [Note 1] Scanning speeds stated above are standard maximum speeds in respective modes. Besides these, the following speeds are set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s, and 10 s. [Note 2] High-resolution mode and high-speed mode are selectable for dedicated scanners only. [Note 3] High-resolution mode is individually switchable for desired channels. High-speed mode is only collective switching for all channels of dedicated scanners. Repeat measurement interval time = (Number of Measuring channels × scanning speed) + data processing time (2 to 20 s) *Data processing time is indeterminate, changed by measurement setting and environment. -------Measuring Targets--------------------Scanning Speed------------ Strain (Gage & transducer) Standard Mode (50 ms/channel) High-resolution Mode (0.28 s/channel) High-speed Mode (20 ms/channel) Voltage/current-output sensor Standard Mode (50 ms/channel) High-speed Mode (20 ms/channel) Civil engineering transducer Standard Mode (50 ms/channel) Temperature sensor (TC, Pt) Standard Mode (50 ms/channel) Potentiometer sensor Standard Mode (50 ms/channel) High-speed Mode (20 ms/channel) [Note 1] High-resolution mode is available only with full bridge system. [Note 2] High-speed mode is available full bridge system, voltage, current, and potentiometer sensor. |
Operating Modes | Real-time Monitor Automatic (Interval measurement) |
Measurement Functions | Initial (Initial values are measured and stored in internal memory.) Measure (Initial values are subtracted from original values.) Original (Raw values are measured without subtraction of initial values.) Easy Measure (Auto zero balancing function is activated.) *The selected function is applied to all channels. |
Coefficient Calculation Function | Multiplication by calibration coefficient Calibration by TEDS Conversion of measured values to physical quantities Scaling Correction |
Unit | 59 units |
Automatic Measurement Function
Interval Measurement | Measurement is automatically performed at preset time intervals. |
---|---|
Trigger Measurement | A relative value (certain changing quantity) or an absolute value triggers measurement. |
Trigger Interval Measurement | Combination of trigger measurement and interval measurement. |
Storage | Internal memory Capacity: Approx. 1.8 GB |
---|
Strain Measurement (Standard Mode)
Constant Voltage Excitation | Approx. 2 or 5 VDC |
---|---|
Constant Current Excitation | Approx. 5.7 mA (Bridge resistance 350 Ω) Approx. 16.7 mA (Bridge resistance 120 Ω) |
Scanning Speed | 50 ms/channel |
Gage Factor | 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks.) |
Initial Value Memory Range | Same as measuring range. |
Measuring Range, Resolution and Accuracy | Measuring Range: 0 to ±50 k ×10-6 strain Resolution: 1 ×10-6 strain Accuracy: ±(0.05% of reading +1) ×10-6 strain Measuring Range: ±50 k to ±500 k ×10-6 strain Resolution: 10 ×10-6 strain Accuracy: ±(0.05% of reading +10) ×10-6 strain *Resolution and accuracy be automatically changed by Autorange function. |
Strain Measurement (High-resolution Mode)
Constant Voltage Excitation | Approx. 5 VDC |
---|---|
Constant Current Excitation | Approx. 16.7 mA (Bridge resistance 350 Ω) |
Scanning Speed | 0.28 s/channel |
Initial Value Memory Range | Same as measuring range. |
Gage Factor | 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks.) |
Measuring Range, Resolution and Accuracy | Measuring Range: 0 to ±20 k ×10-6 strain Resolution: 0.1 ×10-6 strain Accuracy: ±(0.05% of reading +0.3) ×10-6 strain Measuring Range: ±20 k to ±200 k ×10-6 strain Resolution: 1 ×10-6 strain Accuracy: ±(0.05% of reading +3) ×10-6 strain Note 1: Available only with full bridges system. Note 2: Bridge resistance should be 120 to 1000 Ω for bridge excitation with constant voltage. Note 3: Bridge resistance should be 350 Ω for bridge excitation with constant current. [Measuring range] 0 to ±15000 ×10-6 strain 0 to ±150000 ×10-6 strain Note 4: Available only with dedicated scanners. Note 5: Resolution and accuracy be automatically changed by Autorange function. |
Strain Measurement (High-speed Mode)
Constant Voltage Excitation | Approx. 2 VDC |
---|---|
Constant Current Excitation | Approx. 5.7mA (Bridge resistance 350 Ω) Approx. 16.7mA (Bridge resistance 120 Ω) |
Scanning Speed | 20 ms/channel |
Gage Factor | 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks.) |
Initial Value Memory Range | Same as measuring range. |
Measuring Range, Resolution and Accuracy | Measuring Range: 0 to ±50 k ×10-6 strain Resolution: 1 ×10-6 strain Accuracy: ±(0.08% of reading +3) ×10-6 strain Measuring Range: ±50 k to ±500 k ×10-6 strain Resolution: 10 ×10-6 strain Accuracy: ±(0.08% of reading +30) ×10-6 strain [Note 1] Available only with full bridges system (120 to 1000 Ω). [Note 2] Available only with dedicated scanners. [Note 3] Resolution and accuracy be automatically changed by Autorange function. |
Voltage Measurement (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: V/500mV Measuring Range: 0 to ±50.000 mV Resolution: 1 μV Accuracy: ±(0.05% of reading +0.003) mV Input Resistance: 10 MΩ or more Channel Mode: V/500mV Measuring Range: ±50.00 to ±500.00 mV Resolution: 10 μV Accuracy: ±(0.05% of reading +0.03) mV Input Resistance: 10 MΩ or more Channel Mode: V/50V Measuring Range: 0 to ±5.0000 V Resolution: 100 μV Accuracy: ±(0.05% of reading +0.0002) V Input Resistance: 1 MΩ or more Channel Mode: V/50V Measuring Range: ±5.000 to ±50.000 V Resolution: 1 mV Accuracy: ±(0.05% of reading +0.002) V Input Resistance: 1 MΩ or more *Resolution and accuracy be automatically changed by Autorange function. |
Voltage Measurement (High-speed Mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: V/500mV Measuring Range: 0 to ±50.000 mV Resolution: 1 μV Accuracy: ±(0.08% of reading +0.006) mV Input Resistance: 10 MΩ or more Channel Mode: V/500mV Measuring Range: ±50.00 to ±500.00 mV Resolution: 10 μV Accuracy: ±(0.08% of reading +0.06) mV Input Resistance: 10 MΩ or more Channel Mode: V/50V Measuring Range: 0 to ±5.0000 V Resolution: 100 μV Accuracy: ±(0.08% of reading +0.0006) V Input Resistance: 1 MΩ or more Channel Mode: V/50V Measuring Range: ±5.000 to ±50.000 V Resolution: 1 mV Accuracy: ±(0.08% of reading +0.006) V Input Resistance: 1 MΩ or more [Note 1] Resolution and accuracy be automatically changed by Autorange function. [Note 2] Available only with dedicated scanners. |
Current Measurement (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: I/50 mA Measuring Range: 0 to ±50.00 mA Resolution: 10 μA Accuracy: ±(0.05% of reading +0.01) mA [Note 1] External shunt resistor (high-accuracy 250 Ω) is required. [Note 2] Stated accuracy does not include the resistance error of the external shunt resistor. |
Current Measurement (High-speed Mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: I/50 mA Measuring Range: 0 to ±50.00 mA Resolution: 10 μA Accuracy: ±(0.08% of reading +0.01) mA [Note 1] Available only with dedicated scanners. [Note 2] External shunt resistor (high-accuracy 250 Ω) is required. [Note 3] Stated accuracy does not include the resistance error of the external shunt resistor. |
Temperature Measurement with Thermocouple (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Measuring Range, Resolution and Accuracy | 【Type: K】 Measuring Range: −200.0 to 1230.0℃ Resolution: 0.1℃ Accuracy: ±0.7℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: T】 Measuring Range: −200.0 to 400.0℃ Resolution: 0.1℃ Accuracy: ±0.7℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: E】 Measuring Range: −200.0 to 660.0℃ Resolution: 0.1℃ Accuracy: ±0.5℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: J】 Measuring Range: −200.0 to 870.0℃ Resolution: 0.1℃ Accuracy: ±0.6℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: R】 Measuring Range: 0 to 1760.0℃ Resolution: 0.1℃ Accuracy: ±2.2℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) [Note 1] Accuracies do not include the internal reference junction compensator accuracy. [Note 2] The reference junction compensator is switchable between internal and external. [Note 3] Thermocouple resistance should be 1 kΩ or less. |
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Measuring Range, Resolution and Accuracy | Measuring Range: −50.0 to 200.0℃ Resolution: 0.1℃ Accuracy: ±0.5℃ [Note 1] Target physical quantity and temperature are measured in a single channel. [Note 2] Strain measuring range are the same as in strain measurement in standard mode. |
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Measuring Range, Resolution and Accuracy | 【Type: Pt100】 Measuring Range: −200.0 to 660.0℃ Resolution: 0.1℃ Accuracy: ±0.3℃ 【Type: JPt100】 Measuring Range: −200.0 to 510.0℃ Resolution: 0.1℃ Accuracy: ±0.3℃ Note: Connection is 3-wire system. |
Measurement with Potentiometer Sensor (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Sensor Power Supply | Approx. 2 VDC |
Potentiometer Resistance | 1 kΩ to 10 kΩ |
Measuring Range, Resolution and Accuracy | Channel Mode: POT. Measuring Range: 0 to ±50.00% Resolution: 0.01% Accuracy: ±0.1% FS |
Measurement with Potentiometer Sensor (High-speed Mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Sensor Power Supply | Approx. 2 VDC |
Potentiometer Resistance | 1 kΩ to 10 kΩ |
Measuring Range, Resolution and Accuracy | Channel Mode: POT. Measuring Range: 0 to ±50.00% Resolution: 0.01% Accuracy: ±0.1% FS Note: Available only with dedicated scanners. |
Internal Timer | Real time clock is built-in. (Battery backup) |
---|---|
Display | Fluorescent display tube 128 × 64 dots |
Printer | Printing: Thermal Paper width: 58 mm (24 characters/line) Printing speed: 60 mm/s (Max.) |
Interface | RS-232C LAN (10BASE-T/100BASE-TX) USB2.0 (Collects measurement data by USB Flash Drive.) *1: Measurement data cannot be saved directly to USB memory. *2: USB Flash Drives Capacity: 32 GB or less File Format: FAT32 *3: Recommend USB Flash Drives: GH-UFI-XSD2G (Manufacturer: GREENHOUSE) |
File Conversion | Binary measurement data can be converted to CSV. |
Self Diagnosis Functions | Checks display, printer, bridge excitation, leadwire-off, input/output resistance, insulation resistance, mode, etc. |
TEDS | [Interface] IEEE1451.4 Mixed Mode Transducer Interface Class2 [Applicable sensor] Should have information written in accordance with IEEE template No.33. Cable length should be 30m or less. *With dedicated scanner USS-61B/62B/63B. |
Operating Temperature | 0 to 50ºC |
Operating Humidity | 20% to 85% (Non-condensing) |
Storage Temperature | -20 to 60ºC |
Setting Maintenance Function | ACOM at measurement circuit is switchable between floating and GND connect. |
Power Supply | 100 to 240 VAC (AC-operated version) 10 to 16 VDC (DC-operated version) *DC operated version has power control function. |
Current Consumption | 100 VAC: 0.5 A or less (With 3 dedicated scanners mounted) 12 VDC: 4.0 A or less (With 3 dedicated scanners mounted) |
Dimensions | 360 W ×88 H ×400 D mm (Excluding protrusions) |
Weight | Approx. 6.3 kg (Excluding scanner) Approx. 9.6 kg (With 3 dedicated scanners USS-62B mounted) |
Standard Accessories | AC power cable P-18 (With 2-pin conversion plug CM-52) (AC-operated version) DC power cable P-76 (DC-operated version) Recording paper UCAM-60A-RP (1 roll) Screwdriver Spare fuse CD-R (Instruction Manual) Ground wire P-72 |
---|---|
Optional Accessories | Recording paper UCAM-60A-RP (10 rolls/pack) |
Dimensions

UCAM-65C M14
Data Logger

Specifications
Measuring Targets | Strain gages Strain-gage transducers DC voltage-output instruments DC current-output instruments Civil engineering transducers with a thermal sensor Potentiometer sensors Thermal sensors (Thermocouples and platinum resistance thermometer bulbs) |
---|---|
Connectable Scanners | USS-61B, USS-62B, USS-63B (Dedicated scanners, mounted on top of the UCAM-60C M14) The main unit is connected to the following scanners via the optional scanner interface (USI-67A). USB-70B series (via USI-67A) |
Channels | Max. 30 with dedicated scanners Max. 1000 with external scanners connected Max. 1000 with dedicated scanners and external scanners connected |
Input Terminals | Can connect to lead wires through either soldering or screwing. (Screw: M3 with metal plate) NDIS4102 (7 pins) connectors (USS-62B) |
Switching Terminal | Semiconductor relays |
Scanning Speed | 50 ms/channel (Standard mode) 0.28 s/channel (High-resolution mode) 20 ms/channel (High-speed mode) -------Scanners--------------------Line Frequencies------------ Dedicated scanner (Standard mode) 50 Hz Zone: 50 ms/channel 60 Hz Zone: 50 ms/channel Dedicated scanner (High-resolution mode) 50 Hz Zone: 0.28 s/channel 60 Hz Zone: 0.28 s/channel Dedicated scanner (High-speed mode) 50 Hz Zone: 20 ms/channel 60 Hz Zone: 20 ms/channel USB-70B (Standard mode only) 50 Hz Zone: 60 ms/channel 60 Hz Zone: 58.4 ms/channel [Note 1] Scanning speeds stated above are standard maximum speeds in respective modes. Besides these, the following speeds are set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s, and 10 s. [Note 2] High-resolution mode and high-speed mode are selectable for dedicated scanners only. [Note 3] High-resolution mode is individually switchable for desired channels. High-speed mode is only collective switching for all channels of dedicated scanners. Repeat measurement interval time = (Number of Measuring channels × scanning speed) + data processing time (2 to 20 s) *Data processing time is indeterminate, changed by measurement setting and environment. -------Measuring Targets--------------------Scanning Speed------------ Strain (Gage & transducer) Standard Mode (50 ms/channel) High-resolution Mode (0.28 s/channel) High-speed Mode (20 ms/channel) Voltage/current-output sensor Standard Mode (50 ms/channel) High-speed Mode (20 ms/channel) Civil engineering transducer Standard Mode (50 ms/channel) Temperature sensor (TC, Pt) Standard Mode (50 ms/channel) Potentiometer sensor Standard Mode (50 ms/channel) High-speed Mode (20 ms/channel) [Note 1] High-resolution mode is available only with full bridge system. [Note 2] High-speed mode is available full bridge system, voltage, current, and potentiometer sensor. |
Operating Modes | Real-time Monitor Automatic (Interval measurement) |
Measurement Functions | Initial (Initial values are measured and stored in internal memory.) Measure (Initial values are subtracted from original values.) Original (Raw values are measured without subtraction of initial values.) Easy Measure (Auto zero balancing function is activated.) *The selected function is applied to all channels. |
Coefficient Calculation Function | Multiplication by calibration coefficient Calibration by TEDS Conversion of measured values to physical quantities Scaling Correction |
Unit | 59 units |
Automatic Measurement Function
Interval Measurement | Measurement is automatically performed at preset time intervals. |
---|---|
Trigger Measurement | A relative value (certain changing quantity) or an absolute value triggers measurement. |
Trigger Interval Measurement | Combination of trigger measurement and interval measurement. |
Storage | Internal memory Capacity: Approx. 1.8 GB |
---|
Strain Measurement (Standard Mode)
Constant Voltage Excitation | Approx. 2 or 5 VDC |
---|---|
Constant Current Excitation | Approx. 5.7 mA (Bridge resistance 350 Ω) Approx. 16.7 mA (Bridge resistance 120 Ω) |
Scanning Speed | 50 ms/channel |
Gage Factor | 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks.) |
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Measuring Range: 0 to ±50 k ×10-6 strain Resolution: 1 ×10-6 strain Accuracy: ±(0.05% of reading +1) ×10-6 strain Measuring Range: ±50 k to ±500 k ×10-6 strain Resolution: 10 ×10-6 strain Accuracy: ±(0.05% of reading +10) ×10-6 strain *Resolution and accuracy be automatically changed by Autorange function. |
Strain Measurement (High-resolution Mode)
Constant Voltage Excitation | Approx. 5 VDC |
---|---|
Constant Current Excitation | Approx. 16.7 mA (Bridge resistance 350 Ω) |
Scanning Speed | 0.28 s/channel |
Initial Value Memory Range | Same as measuring range. |
Gage Factor | 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks.) |
Measuring Range, Resolution and Accuracy | Measuring Range: 0 to ±20 k ×10-6 strain Resolution: 0.1 ×10-6 strain Accuracy: ±(0.05% of reading +0.3) ×10-6 strain Measuring Range: ±20 k to ±200 k ×10-6 strain Resolution: 1 ×10-6 strain Accuracy: ±(0.05% of reading +3) ×10-6 strain Note 1: Available only with full bridges system. Note 2: Bridge resistance should be 120 to 1000 Ω for bridge excitation with constant voltage. Note 3: Bridge resistance should be 350 Ω for bridge excitation with constant current. [Measuring range] 0 to ±15000 ×10-6 strain 0 to ±150000 ×10-6 strain Note 4: Available only with dedicated scanners. Note 5: Resolution and accuracy be automatically changed by Autorange function. |
Strain Measurement (High-speed Mode)
Constant Voltage Excitation | Approx. 2 VDC |
---|---|
Constant Current Excitation | Approx. 5.7mA (Bridge resistance 350 Ω) Approx. 16.7mA (Bridge resistance 120 Ω) |
Scanning Speed | 20 ms/channel |
Gage Factor | 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks.) |
Initial Value Memory Range | Same as measuring range. |
Measuring Range, Resolution and Accuracy | Measuring Range: 0 to ±50 k ×10-6 strain Resolution: 1 ×10-6 strain Accuracy: ±(0.08% of reading +3) ×10-6 strain Measuring Range: ±50 k to ±500 k ×10-6 strain Resolution: 10 ×10-6 strain Accuracy: ±(0.08% of reading +30) ×10-6 strain [Note 1] Available only with full bridges system (120 to 1000 Ω). [Note 2] Available only with dedicated scanners. [Note 3] Resolution and accuracy be automatically changed by Autorange function. |
Voltage Measurement (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: V/500mV Measuring Range: 0 to ±50.000 mV Resolution: 1 μV Accuracy: ±(0.05% of reading +0.003) mV Input Resistance: 10 MΩ or more Channel Mode: V/500mV Measuring Range: ±50.00 to ±500.00 mV Resolution: 10 μV Accuracy: ±(0.05% of reading +0.03) mV Input Resistance: 10 MΩ or more Channel Mode: V/50V Measuring Range: 0 to ±5.0000 V Resolution: 100 μV Accuracy: ±(0.05% of reading +0.0002) V Input Resistance: 1 MΩ or more Channel Mode: V/50V Measuring Range: ±5.000 to ±50.000 V Resolution: 1 mV Accuracy: ±(0.05% of reading +0.002) V Input Resistance: 1 MΩ or more *Resolution and accuracy be automatically changed by Autorange function. |
Voltage Measurement (High-speed Mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: V/500mV Measuring Range: 0 to ±50.000 mV Resolution: 1 μV Accuracy: ±(0.08% of reading +0.006) mV Input Resistance: 10 MΩ or more Channel Mode: V/500mV Measuring Range: ±50.00 to ±500.00 mV Resolution: 10 μV Accuracy: ±(0.08% of reading +0.06) mV Input Resistance: 10 MΩ or more Channel Mode: V/50V Measuring Range: 0 to ±5.0000 V Resolution: 100 μV Accuracy: ±(0.08% of reading +0.0006) V Input Resistance: 1 MΩ or more Channel Mode: V/50V Measuring Range: ±5.000 to ±50.000 V Resolution: 1 mV Accuracy: ±(0.08% of reading +0.006) V Input Resistance: 1 MΩ or more [Note 1] Resolution and accuracy be automatically changed by Autorange function. [Note 2] Available only with dedicated scanners. |
Current Measurement (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: I/50 mA Measuring Range: 0 to ±50.00 mA Resolution: 10 μA Accuracy: ±(0.05% of reading +0.01) mA [Note 1] External shunt resistor (high-accuracy 250 Ω) is required. [Note 2] Stated accuracy does not include the resistance error of the external shunt resistor. |
Current Measurement (High-speed Mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Measuring Range, Resolution and Accuracy | Channel Mode: I/50 mA Measuring Range: 0 to ±50.00 mA Resolution: 10 μA Accuracy: ±(0.08% of reading +0.01) mA [Note 1] Available only with dedicated scanners. [Note 2] External shunt resistor (high-accuracy 250 Ω) is required. [Note 3] Stated accuracy does not include the resistance error of the external shunt resistor. |
Temperature Measurement with Thermocouple (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Measuring Range, Resolution and Accuracy | 【Type: K】 Measuring Range: −200.0 to 1230.0℃ Resolution: 0.1℃ Accuracy: ±0.7℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: T】 Measuring Range: −200.0 to 400.0℃ Resolution: 0.1℃ Accuracy: ±0.7℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: E】 Measuring Range: −200.0 to 660.0℃ Resolution: 0.1℃ Accuracy: ±0.5℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: J】 Measuring Range: −200.0 to 870.0℃ Resolution: 0.1℃ Accuracy: ±0.6℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) 【Type: R】 Measuring Range: 0 to 1760.0℃ Resolution: 0.1℃ Accuracy: ±2.2℃ Internal Reference Junction Compensator Accuracy: ±0.5℃ (With input terminal temperature balanced in an ambient) (Temp. range of 0 to 50°C) [Note 1] Accuracies do not include the internal reference junction compensator accuracy. [Note 2] The reference junction compensator is switchable between internal and external. [Note 3] Thermocouple resistance should be 1 kΩ or less. |
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Measuring Range, Resolution and Accuracy | Measuring Range: −50.0 to 200.0℃ Resolution: 0.1℃ Accuracy: ±0.5℃ [Note 1] Target physical quantity and temperature are measured in a single channel. [Note 2] Strain measuring range are the same as in strain measurement in standard mode. |
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Measuring Range, Resolution and Accuracy | 【Type: Pt100】 Measuring Range: −200.0 to 660.0℃ Resolution: 0.1℃ Accuracy: ±0.3℃ 【Type: JPt100】 Measuring Range: −200.0 to 510.0℃ Resolution: 0.1℃ Accuracy: ±0.3℃ Note: Connection is 3-wire system. |
Measurement with Potentiometer Sensor (Standard Mode)
Scanning Speed | 50 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Sensor Power Supply | Approx. 2 VDC |
Potentiometer Resistance | 1 kΩ to 10 kΩ |
Measuring Range, Resolution and Accuracy | Channel Mode: POT. Measuring Range: 0 to ±50.00% Resolution: 0.01% Accuracy: ±0.1% FS |
Measurement with Potentiometer Sensor (High-speed Mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial Value Memory Range | Same as measuring range |
Sensor Power Supply | Approx. 2 VDC |
Potentiometer Resistance | 1 kΩ to 10 kΩ |
Measuring Range, Resolution and Accuracy | Channel Mode: POT. Measuring Range: 0 to ±50.00% Resolution: 0.01% Accuracy: ±0.1% FS Note: Available only with dedicated scanners. |
Internal Timer | Real time clock is built-in. (Battery backup) |
---|---|
Interface | RS-232C LAN (10BASE-T/100BASE-TX) USB2.0 (Collects measurement data by USB Flash Drive.) *1: Measurement data cannot be saved directly to USB memory. *2: USB Flash Drives Capacity: 32 GB or less File Format: FAT32 *3: Recommend USB Flash Drives: GH-UFI-XSD2G (Manufacturer: GREENHOUSE) |
Self Diagnosis Functions | Checks bridge excitation, leadwire-off, input/output resistance, insulation resistance, mode, etc. |
TEDS | [Interface] IEEE1451.4 Mixed Mode Transducer Interface Class2 [Applicable sensor] Should have information written in accordance with IEEE template No.33. Cable length should be 30m or less. *With dedicated scanner USS-61B/62B/63B. |
Operating Temperature | 0 to 50ºC |
Operating Humidity | 20% to 85% (Non-condensing) |
Storage Temperature | -20 to 60ºC |
Setting Maintenance Function | ACOM at measurement circuit is switchable between floating and GND connect. |
Power Supply | 100 to 240 VAC (AC-operated version) 10 to 16 VDC (DC-operated version) *DC operated version has power control function. |
Current Consumption | 100 VAC: 0.5 A or less (With 3 dedicated scanners mounted) 12 VDC: 4.0 A or less (With 3 dedicated scanners mounted) |
Dimensions | 327 W ×88 H ×365 D mm (Excluding protrusions) |
Weight | Approx. 5.0 kg (Excluding scanners) Approx. 8.3 kg (With 3 dedicated scanners USS-62B mounted) |
Standard Accessories | AC power cable P-18 (With 2-pin conversion plug CM-52) (AC-operated version) DC power cable P-76 (DC-operated version) Screwdriver Spare fuse CD-R (Instruction Manual) CD-R (Control software UCS-60B) Ground wire P-72 |
---|
Dimensions

USS-61B
Dedicated Scanner (TEDS compatible)

Specifications
Model | TEDS compatible |
---|---|
Channels | 10/unit |
Switching Terminals | Semiconductor relays |
Input Terminals | Connect to lead wire by either soldering or screwing. One-touch terminal block (JT-1A) (Optional) |
Operating Temperature | 0 to 50ºC |
Operating Humidity | 20 to 80% (Non-condensing) |
Dimensions | 320 W x 28 H x 80 D mm (Excluding protrusions) |
Weight | Approx. 0.8 kg |
Standard Accessories | Terminal cover Channel label Short bar ML-1000-3H1 |
---|
USS-62B
Dedicated Scanner (With NDIS4102 (7 pins) connectors, TEDS compatible)

Specifications
Model | With NDIS4102 (7 pins) connectors, TEDS compatible |
---|---|
Channels | 10/unit |
Switching Terminals | Semiconductor relays |
Input Terminals | Connect to lead wire by either soldering or screwing. NDIS4102 (7 pins) connectors One-touch terminal block (JT-1A) (Optional) |
Operating Temperature | 0 to 50ºC |
Operating Humidity | 20 to 80% (Non-condensing) |
Dimensions | 320 W x 28 H x 80 D mm (Excluding protrusions) |
Weight | Approx. 1.1 kg |
Standard Accessories | NDIS4102 (7 pins) connector caps (Pre-attached to connectors) Terminal cover Channel label Short bar ML-1000-3H1 |
---|
USS-63B
Dedicated Scanner (For civil engineering measurement, TEDS compatible, with lightning arresters)

Specifications
Model | For civil engineering measurement, TEDS compatible, with lightning arresters |
---|---|
Channels | 10/unit |
Switching Terminals | Semiconductor relays |
Input Terminals | Connect to lead wire by either soldering or screwing. One-touch terminal block (JT-1A) (Optional) |
Lightning Arresters | Built in USS-63B |
Operating Temperature | 0 to 50ºC |
Operating Humidity | 20 to 80% (Non-condensing) |
Dimensions | 320 W x 28 H x 80 D mm (Excluding protrusions) |
Weight | Approx. 1.0 kg |
Standard Accessories | Terminal cover Channel label Short bar ML-1000-3H1 |
---|
USI-67A
Scanner Interfaces

Specifications
Connectable Scanners | USB-70B |
---|---|
Number of Scanners | Max. 20 |
Cable Length | Max. 1 km (When connecting the UPS-70B to the USB-70B.) |
Operating Temperature | 0 to 50ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Dimensions | 99 W x 50 H x 163 D mm (Excluding protrusions) |
Weight | Approx. 170 g |
UIO-60A
External I/O Unit

Specifications
Output
ALARM Signal | 4 channels (High/low limit checking) |
---|---|
BUSY Signal | 1 channel |
Input
START Signal | 1 channel |
---|---|
STOP Signal | 1 channel |
RESET Signal | 1 channel |
RAINFALL Signals | 1 channel |
Operating Temperature | 0 to 50 ºC |
---|---|
Operating Humidity | 20 to 85% (Non-condensing) |
Dimensions | 90 W x 50 H x 180 D mm (Excluding protrusions) |
Weight | Approx. 140 g |
Standard Accessories | Screwdriver |
---|
USB-70B-10
External Scanner (For general measurement)

Specifications
Models | USB-70B-10 (For general measurement) |
---|---|
Measuring Targets | Strain gages Strain-gage transducers DC voltage 0 to ±50 V DC current 0 to ±50 mA Thermocouples K, T, E, J, N, B, S, R Potentiometer sensors |
Channels | 50/unit |
Connection Data Loggers | UCAM-60C M14 UCAM-65C M14 *An optional Scanner Interface USI-67A is required. |
Number of Connection | Up to 20 *19 if there are 30 channels in the data logger |
Input Terminals | Screw-soldering terminal blocks (M3 screw with washer) [Receptacle] PRC90-231BR10-7F [Applicable plug] ex. PRC03-12A10-7M10.5 |
Switching Elements | Semiconductor relays |
Measuring Mode | Every channel set by the data logger |
Measuring Channel Number | The upper 2-digit from every 10 channels are set with digital switches |
Power Supply | Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into the scanners. |
Dimensions | 302 W x 107 H x 500 D mm (Excluding protrusions) |
Weight | Approx. 7.3 kg (Excluding UPS-70B) |
Option | Built-in AC Power Unit UPS-70B 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover) |
Standard Accessories | Phillips precision screwdriver Connection cable N-24 (1 m) |
---|
USB-70B-20
External Scanner (For general measurement, with NDIS4102 (7 pins) connectors)

Specifications
Models | USB-70B-20 (For general measurement, with NDIS4102 (7 pins) connectors) |
---|---|
Measuring Targets | Strain gages Strain-gage transducers DC voltage 0 to ±50 V DC current 0 to ±50 mA Thermocouples K, T, E, J, N, B, S, R Potentiometer sensor |
Channels | 50/unit |
Connection Data Loggers | UCAM-60C M14 UCAM-65C M14 *An optional Scanner Interface USI-67A is required. |
Number of Connection | Up to 20 *19 if there are 30 channels in the data logger |
Input Terminals | Screw-soldering terminal blocks (M3 screw with washer) NDIS4102 (7 pins) connectors [Receptacle] PRC90-231BR10-7F [Applicable plug] ex. PRC03-12A10-7M10.5 |
Switching Elements | Semiconductor relays |
Measuring Modes | Every channel set by the data logger |
Measuring Channel Number | The upper 2-digit from every 10 channels are set with digital switches |
Power Supply | Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into the scanners. |
Dimensions | 302 W x 107 H x 500 D mm (Excluding protrusions) |
Weight | Approx. 8.5 kg (Excluding UPS-70B) |
Option | Built-in AC Power Unit UPS-70B 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover) |
Standard Accessories | Phillips precision screwdriver Connection cable N-24 (1 m) |
---|
USB-70B-30
External Scanner (For civil engineering)

Specifications
Model | USB-70B-30 (For civil engineering) |
---|---|
Measuring Targets | Strain gages Strain-gage transducers DC voltage 0 to ±50 V DC current 0 to ±50 mA Thermocouples K, T, E, J, N, B, S, R Platinum resistance thermometer bulbs Pt100, JPt100 Civil engineering transducers with a thermal sensor Potentiometer sensor |
Channels | 50/unit |
Connection Data Loggers | UCAM-60C M14 UCAM-65C M14 *An optional Scanner Interface USI-67A is required. |
Number of Connection | Up to 20 *19 if there are 30 channels in the data logger |
Input Terminals | Screw-soldering terminal blocks (M3 screw with washer) [Receptacle] PRC90-231BR10-7F [Applicable plug] ex. PRC03-12A10-7M10.5 Built-in lightning arresters |
Switching Elements | Semiconductor relays |
Measuring Mode | Every channel set by the data logger |
Measuring Channel Number | The upper 2-digit from every 10 channels are set with digital switches |
Power Supply | Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into the scanners. |
Dimensions | 302 W x 107 H x 500 D mm (Excluding protrusions) |
Weight | Approx. 7.7 kg (Excluding UPS-70B) |
Option | Built-in AC Power Unit UPS-70B 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover) |
Standard Accessories | Phillips precision screwdriver Connection cable N-24 (1 m) |
---|