UCAM-60C M14, UCAM-65C M14
Data Logger

  • TEDS

UCAM-60C/65C

Summary

Saving the long-term saving measured results to the built-in memory (Approx. 1.8 GB)
USB memory ensures easy data collection
Up to 1000 channels measurement with external scanners
Measurement up to 20 k ×10-6 strain with a resolution of 0.1 ×10-6 strain (With full bridge system)
The UCAM-60C M14 is an all-in-one measuring instrument
The UCAM-65C M14 is an online data logger fully controlled from the PC

Features

The data logger UCAM-60C M14 is an all-in-one measuring instrument developed in full pursuit of easier field measurement. Easy to operate keys, a bright readable display providing understandable presentation and a printer for immediate confirmation of measurement results. All these and more are incorporated in this compact unit to satisfy every need in field measurement.
The UCAM-65C M14 is an online data logger fully controlled from the PC.

Products table

All display

UCAM-60C M14
Data Logger

UCAM-60C
  • TEDS

Specifications

Model AC power supply models: UCAM-60C-AC M14
DC power supply models: UCAM-60C-DC M14
Channels Max 30 with dedicated scanners
Max 1000 with external scanners connected
Max 1000 with dedicated scanners and external scanners connected
Measuring Targets and Connectable Scanners
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor, DC voltage-output or DC current-output instruments,
potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Connectable Scanners USS-61B, USS-62B, USS-63B (Dedicated Scanners, mounted on top of the UCAM-60C)
The main unit can connect scanners USB-70B series via an optional scanner interface USI-67A
Input Terminals Capable of connecting lead wires through either soldering or screwing. (Screw: M3 with metal plate)
NDIS connectors (USS-62B)
Switching Terminals Semiconductor relays
Scanning Speed 50 ms/channel (Standard mode)
0.28 s/channel (High-resolution mode)
Note 1: Individually switchable for desired channels.
20 ms/channel (High-speed mode)
Note 2: Only collective switching for all channels of dedicated scanners.
Note 3: High-resolution mode and high-speed mode are selectable for dedicated scanners only.
Note 4: High-resolution mode is available only with full bridge system.
Note 5: High-speed mode is available full bridge system, voltage, current, and potentiometer sensor.
Operating Modes Real-time, monitor, and automatic
Measurement Functions
Initial Initial values are measured and stored in internal memory.
Original Raw values are measured without subtraction of initial values.
Measure Initial values are subtracted from original values.
Easy Measure Auto zero balancing function is activated.
Notes The selected function is applied to all channels.
Coefficient Calculation Function Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.
Units 59 units
Automatic Measurement Function
Trigger Measurement A relative value (Certain changing quantity) or an absolute value trigger measurement.
Interval Measurement Measurement is automatically performed at preset time intervals.
Trigger Interval Measurement Combination of trigger measurement and interval measurement.
Storage Internal memory, approx. 1.8 GB
Strain Measurement (Standard Mode)
Constant Voltage Excitation Approx. 2 or 5 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω)
Approx. 16.7 mA (Bridge resistance 120 Ω)
Scanning Speed 50 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ±50k ×10-6 strain
±50k to ±500k ×10-6 strain
Strain Measurement (High-resolution Mode)
Constant Voltage Excitation Approx. 5 VDC
Constant Current Excitation Approx. 16.7 mA (Bridge resistance 350 Ω)
Scanning Speed 0.28 s/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ±20k ×10-6 strain
±20k to ±200k ×10-6 strain
Notes 1. Available only with full bridge system
2. Bridge resistance should be 120 to 1000 Ω for bridge excitation with constant voltage.
3. Bridge resistance should be 350 Ω for bridge excitation with constant current.
Measuring range is 0 to 15k ×10-6 strain, 0 to 150k ×10-6 strain
4. Available only with dedicated scanners.
Strain Measurement (High-speed Mode)
Constant Voltage Excitation Approx. 2 VDC
Constant Current Excitation Approx. 5.7mA (Bridge resistance 350 Ω)
Approx. 16.7mA (Bridge resistance 120 Ω)
Scanning Speed 20 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ±50k ×10-6 strain, ±50k to ±500k ×10-6 strain
Notes 1. Available only with full bridges system (120 to 1000 Ω)
2. Available only with dedicated scanners.
Voltage Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
V/500mV Range Mode: 0 to ±50.000 mV, ±50.00 to ±500.00 mV
V/50V Range Mode: 0 to ±5.0000 V, ±5.000 to ±50.000 V
Voltage Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
V/500mV Range Mode: 0 to ±50.000 mV, ±50.00 to ±500.00 mV
V/50V Range Mode: 0 to ±5.0000 V, ±5.000 to ±50.000 V
Notes 1. Resolution and accuracy be automatically change by autorange function.
2. Available only with dedicated scanners.
Current Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
0 to ±50.00 mA
Notes External shunt resistors (High-accuracy 250 Ω are required.
Current Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
0 to ±50.00 mA
Notes 1. Available only with dedicated scanners.
2. External shunt resistors (High-accuracy 250 Ω are required.
Temperature Measurement with Thermocouple (Standard Mode)
Scanning Speed 50 ms/channel
Notes Thermocouple resistance should be 1000 Ω or less.
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Notes 1. Target physical quantity and temperature are measured in a single channel.
2. Strain measuring range is the same as in strain measurement in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed 50 ms/channel
Measurement with Potentiometer Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Same as measuring range
0% to ±50.00%
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 k Ω
Measurement with Potentiometer Sensor (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
0% to ±50.00%
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 kΩ
Notes Available only with dedicated scanners.
Internal Timer Real time clock is Built-in (Battery backup)
Display Fluorescent display tube,
128×64 dots
Printer Printing: Thermal
Paper Width: 58 mm (24 characters/line)
Printing Speed: 60 mm/s
Interfaces RS-232C
LAN (10BASE-T, 100BASE-TX)
USB2.0 (Collect measurement data by USB Flash Drive)
Notes 1: Measurement data cannot be saved directly to USB memory.
Notes 2: USB Flash Drives Capacity is 32 GB or less
File Format: FAT32
Notes 3: Recommend USB Flash Drives GH-UFI-XSC2G (The manufacturer is Greenhouse)
File Conversion Binary measurement data can be converted to CSV.
Self-diagnosis Functions Checks display, printer, bridge excitation, lead-wire-off, I/O resistance, insulation resistance, mode, etc.
TEDS ・Interface
IEEE1451.4 Mixed mode Transducer Interface Class2
・Applicable sensor
Should have information written in accordance with IEEE template No.33
Cable length should be 30m or less
(With dedicated scanners USS-61B,USS-62B, USS-63B)
Operating Temperature 0 to 50 ºC
Operating Humidity 20% to 85% (Non-condensing)
Storage Temperature -20 to 60 ºC
Power Supply AC 100 to 240 V (AC-operated models)
DC 10 to 16 V (DC-operated models)
Note: DC operated models have power control function
Current Consumption 0.5 A or less: 100 VAC (With 3 dedicated scanners mounted)
4.0 A or less: 12 VDC (With 3 dedicated scanners mounted)
Dimensions 360 W ×88 H ×400 D mm (Excluding protrusions)
Weight Approx. 6.3 kg (Excluding scanners)
Approx. 9.6 kg (With 3 dedicated scanners USS-62B mounted)

UCAM-65C M14
Data Logger

UCAM-65C
  • TEDS

Specifications

Model AC power supply models: UCAM-65C-AC M14, UCAM-65C-AC-0 M14*
DC power supply models: UCAM-65C-DC M14, UCAM-65C-DC-0 M14*
The suffix -0 means without Control Software UCS-60B
Channels Max 30 with dedicated scanners
Max 1000 with external scanners connected
Max 1000 with dedicated scanners and external scanners connected
Measuring Targets and Connectable Scanners
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor, DC voltage-output or DC current-output instruments,
potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Connectable Scanners USS-61B, USS-62B, USS-63B (Dedicated Scanners, mounted on top of the UCAM-65C)
The main unit can connect scanners USB-70B series via an optional scanner interface USI-67A
Input Terminals Capable of connecting lead wires through either soldering or screwing. (Screw: M3 with metal plate)
NDIS connectors (USS-62B)
Switching Terminals Semiconductor relays
Scanning Speed 50 ms/channel (Standard mode)
0.28 s/channel (High-resolution mode)
Note 1: Individually switchable for desired channels.
20 ms/channel (High-speed mode)
Note 2: Only collective switching for all channels of dedicated scanners.
Note 3: High-resolution mode and high-speed mode are selectable for dedicated scanners only.
Note 4: High-resolution mode is available only with full bridge system.
Note 5: High-speed mode is available full bridge system, voltage, current, and potentiometer sensor.
Operating Modes Real-time, monitor, and automatic
Measurement Functions
Initial Initial values are measured and stored in internal memory.
Original Raw values are measured without subtraction of initial values.
Measure Initial values are subtracted from original values.
Easy Measure Auto zero balancing function is activated.
Notes The selected function is applied to all channels.
Coefficient Calculation Function Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.
Units 59 units
Automatic Measurement Function
Trigger Measurement A relative value (Certain changing quantity) or an absolute value trigger measurement.
Interval Measurement Measurement is automatically performed at preset time intervals.
Trigger Interval Measurement Combination of trigger measurement and interval measurement.
Storage Internal memory, approx. 1.8 GB
Strain Measurement (Standard Mode)
Constant Voltage Excitation Approx. 2 or 5 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω)
Approx. 16.7 mA (Bridge resistance 120 Ω)
Scanning Speed 50 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ±50k ×10-6 strain
±50k to ±500k ×10-6 strain
Strain Measurement (High-resolution Mode)
Constant Voltage Excitation Approx. 5 VDC
Constant Current Excitation Approx. 16.7 mA (Bridge resistance 350 Ω)
Scanning Speed 0.28 s/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ±20k ×10-6 strain
±20k to ±200k ×10-6 strain
Notes 1. Available only with full bridge system
2. Bridge resistance should be 120 to 1000 Ω for bridge excitation with constant voltage.
3. Bridge resistance should be 350 Ω for bridge excitation with constant current.
Measuring range is 0 to 15k ×10-6 strain, 0 to 150k ×10-6 strain
4. Available only with dedicated scanners.
Strain Measurement (High-speed Mode)
Constant Voltage Excitation Approx. 2 VDC
Constant Current Excitation Approx. 5.7mA (Bridge resistance 350 Ω)
Approx. 16.7mA (Bridge resistance 120 Ω)
Scanning Speed 20 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ±50k ×10-6 strain, ±50k to ±500k ×10-6 strain
Notes 1. Available only with full bridges system (120 to 1000 Ω)
2. Available only with dedicated scanners.
Voltage Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
V/500mV Range Mode: 0 to ±50.000 mV, ±50.00 to ±500.00 mV
V/50V Range Mode: 0 to ±5.0000 V, ±5.000 to ±50.000 V
Voltage Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
V/500mV Range Mode: 0 to ±50.000 mV, ±50.00 to ±500.00 mV
V/50V Range Mode: 0 to ±5.0000 V, ±5.000 to ±50.000 V
Notes 1. Resolution and accuracy be automatically change by autorange function.
2. Available only with dedicated scanners.
Current Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
0 to ±50.00 mA
Notes External shunt resistors (High-accuracy 250 Ω are required.
Current Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
0 to ±50.00 mA
Notes 1. Available only with dedicated scanners.
2. External shunt resistors (High-accuracy 250 Ω are required.
Temperature Measurement with Thermocouple (Standard Mode)
Scanning Speed 50 ms/channel
Notes Thermocouple resistance should be 1000 Ω or less.
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Notes 1. Target physical quantity and temperature are measured in a single channel.
2. Strain measuring range is the same as in strain measurement in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed 50 ms/channel
Measurement with Potentiometer Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Same as measuring range
0% to ±50.00%
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 kΩ
Measurement with Potentiometer Sensor (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
0% to ±50.00%
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 kΩ
Notes Available only with dedicated scanners.
Internal Timer Real time clock is Built-in (Battery backup)
Interfaces RS-232C
LAN (10BASE-T, 100BASE-TX)
USB2.0 (Collect measurement data by USB Flash Drive)
Notes 1: Measurement data cannot be saved directly to USB memory.
Notes 2: USB Flash Drives Capacity is 32 GB or less
File Format: FAT32
Notes 3: Recommend USB Flash Drives GH-UFI-XSC2G (The manufacturer is Greenhouse)
File Conversion Binary measurement data can be converted to CSV.
Self-diagnosis Functions Checks display, printer, bridge excitation, lead-wire-off, I/O resistance, insulation resistance, mode, etc.
TEDS ・Interface
IEEE1451.4 Mixed mode Transducer Interface Class2
・Applicable sensor
Should have information written in accordance with IEEE template No.33
Cable length should be 30m or less
(With dedicated scanners USS-61B,USS-62B, USS-63B)
Operating Temperature 0 to 50 ºC
Operating Humidity 20% to 85% (Non-condensing)
Storage Temperature -20 to 60 ºC
Power Supply AC 100 to 240 V (AC-operated models)
DC 10 to 16 V (DC-operated models)
Note: DC operated models have power control function
Current Consumption 0.5 A or less: 100 VAC (With 3 dedicated scanners mounted)
4.0 A or less: 12 VDC (With 3 dedicated scanners mounted)
Dimensions 360 W ×88 H ×400 D mm (Excluding protrusions)
Weight Approx. 6.3 kg (Excluding scanners)
Approx. 9.6 kg (With 3 dedicated scanners USS-62B mounted)

USS-61B
Dedicated Scanner (TEDS compatible)

USS-61B

Specifications

Model TEDS compatible
Channels 10/Unit
Switching Terminals Semiconductor relays
Input Terminals Connect to leadwires through either soldering or screwing.
One-touch terminal block JT-1A (Optional)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 80% (Non-condensing)
Dimensions 320 W x 28 H x 80 D mm (Excluding protrusions)
Weight Approx800g(Including terminal cover)
Standard Accessories Terminal cover, Channel label

USS-62B
Dedicated Scanner (With NDIS connectors, TEDS compatible)

USS-62B

Specifications

Model With NDIS connectors,TEDS compatible
Channels 10/Unit
Switching Terminals Semiconductor relays
Input Terminals Connect to leadwires through either soldering or screwing NDIS connectors One-touch terminal block JT-1A (Optional)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 80% (Non-condensing)
Dimensions 320 W x 28 H x 80 D mm (Excluding protrusions)
Weight Approx. 1 kg(Including terminal cover)
Standard Accessories NDIS connector caps (Pre-attached to connectors),
terminal cover, channel label

USS-63B
Dedicated Scanner (For civil engineering measurement,TEDS compatible, with lightning arresters)

USS-63B

Specifications

Model For civil engineering measurement, TEDS compatible, with lightning Arresters
Channels 10/Unit
Switching Terminals Semiconductor relays
Input Terminals Connect to lead wire by either soldering or screwing.
One-touch terminal block JT-1A (Optional)
Lightning Arresters Built
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Dimensions 320 W x 28 H x 80 D mm (Excluding protrusions)
Weight Approx. 900 g (Including terminal cover)
Standard Accessories Terminal cover, Channel label

USB-70B-10
External Scanner (for general strain measurement)

USB-70B-10

Specifications

Channels 50/unit
Measuring Channel Mode Selected for each channel from the mainframe
Measuring Targets Strain gages, strain-gage transducers,potentiometer, DC voltage-output instruments,thermocouples
Power Supply Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into scanners.
UPS-70B operates on 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx7.3 kg
Standard Accessories Connection cable N-24 (1 m)

USB-70B-20
External Scanner (for general strain measurement, with NDIS connectors)

Specifications

Channels 50/unit
Measuring Channel Modes Selected for each channel from the mainframe
Measuring Targets Strain gages, strain-gagee transducers, potentiometer,DC voltage-output instruments, thermocouples
(Transducer with NDIS connector is required)
Power Supply Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into scanners.
UPS-70B operates on 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx. 8.5 kg
Standard Accessories Connection cable N-24 (1 m)

USB-70B-30
External Scanner(for civil engineering, with lightning arresters)

USB-70B

Specifications

Channels 50/unit
Measuring Channel Mode Selected for each channel from the mainframe
Measuring Targets Strain gages, strain-gage transducers, potentiometer,DC voltage-output instruments, thermal sensors
(Thermocouples, platinum resistance thermometer bulbs, civil engineering transducers with a thermal sensor), lightning arresters built in
Power Supply Supplied from data logger. If the cable is extended or if 4 or more scanners are connected, an optional UPS-70B should be mounted into scanners.
UPS-70B operates on 100 to 240 VAC (100 to 127 VAC or 220 to 240 VAC automatic switchover)
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% RH (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx7.7 kg
Standard Accessories Connection cable N-24 (1 m)

USB-65A
External Scanners

Specifications

Measuring Targets
Connectable Data Loggers UCAM-60B, 65B*Dedicated Scanners USI-67A is required, Connect with USB-70B is possible.
Channels 50/unit
Input Terminals Can connect to leadwires through either soldering
or screwing.
Switching Terminals Semiconductor relays
Differential Transformer Types Transducers
Sacnning Speed 1s/channel
Power Supply to Sensors 500 Hz 1 Vrms sine wave
Measuring Range, Resolution, Accuracy Measuring range: 0 to ±1000.00
Resolution: 0.01
Accuracy: ±1%FS as follows
SlidingResistance Types Transducers
Sacnning Speed 50 ms/channel
Power Supply to Sensors 2 VDC (From the unit base)
Measuring Range, Resolution, Accuracy Measuring range: 0 to ±100%
Resolution: 0.01%
Accuracy: ±1%Fs as follows
Power Supply Supplied from data logger If the cable is extended
or if 4 or more scanners are connected, an optional
UPS-70B should be mounted into scanners.
UPS-70B operates on 90 to 132/180 to 264 VAC
(No switchover required), 50/60 Hz
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% (Non-condensing)
Dimensions 302 W x 107 H x 500 D mm (Excluding protrusions)
Weight Approx. 10 kg

UIO-60A
External I/O Unit

UIO-60A

Specifications

Output
ALARM Signal 4 channels (High/low limit checking)
BUSY Signal 1 channel
Input
START Signal 1 channel
STOP Signal 1 channel
RESET Signal 1 channel
RAINFALL Signals 1 channel
Operating Temperature 0 to 50 ºC
Operating Humidity 20% to 85% (Non-condensing)
Dimensions 90 W x 50 H x 180 D mm (Excluding protrusions)
Weight Approx. 140 g

USI-67A
Scanner Interfaces

Specifications

Connectable Scanners USB-70B, USB-65A
Number of Scanners Max. 20
Operating Temperature 0 to 50ºC
Operating Humidity 20 to 85% (Non-condensing)
Dimensions 99 W x 50 H x 163 D mm (Excluding protrusions)
Weight Approx. 160 g