WDS-10
Strain Generator

Summary

· Continuously generates equivalent strain with positive or negative polarity
· Digital dial adopted
· No power supply is required

Features

The WDS-10 is an equivalent Strain Generator for checking strain measuring instruments.

Products table

All display

WDS-10-1
Equivalent strain 0 to ±1000 ×10-6 strain

Specifications

I/O Resistance 350 ± 7 Ω
Dial Range 000 to 999
Setting Accuracy Within ±(0.5% of reading +5 digits) (Dial value)
Notes:
1. It is not recommended to use for carrier-type strain amplifiers such as
DPM series.
2. The WDS-10 is designed for checking and is not for calibration.
Dimensions 30 W ×70 H ×80 D mm
Weight Approx. 350 g

Dimensions

WDS-10-2
Equivalent strain 0 to ±2000 ×10-6 strain

Specifications

I/O Resistance 350 ± 7 Ω
Dial Range 000 to 999
Setting Accuracy Within ±(0.5% of reading +5 digits) (Dial value)
Notes:
1. It is not recommended to use for carrier-type strain amplifiers such as
DPM series.
2. The WDS-10 is designed for checking and is not for calibration.
Dimensions 30 W ×70 H ×80 D mm
Weight Approx. 350 g

Dimensions

WDS-10-3
Equivalent strain 0 to ±3000 ×10-6 strain

Specifications

I/O Resistance 350 ± 7 Ω
Dial Range 000 to 999
Setting Accuracy Within ±(0.5% of reading +5 digits) (Dial value)
Notes:
1. It is not recommended to use for carrier-type strain amplifiers such as
DPM series.
2. The WDS-10 is designed for checking and is not for calibration.
Dimensions 30 W ×70 H ×80 D mm
Weight Approx. 350 g

Dimensions

WDS-10-4
Equivalent strain 0 to ±4000 ×10-6 strain

Specifications

I/O Resistance 350 ± 7 Ω
Dial Range 000 to 999
Setting Accuracy Within ±(0.5% of reading +5 digits) (Dial value)
Notes:
1. It is not recommended to use for carrier-type strain amplifiers such as
DPM series.
2. The WDS-10 is designed for checking and is not for calibration.
Dimensions 30 W ×70 H ×80 D mm
Weight Approx. 350 g

Dimensions

Page Top

Remarks

1. Since the WDS-10 has a special circuit structure, the stated accuracy may not be satisfied depending on measuring instruments under test.
2. The WDS-10 is designed for checking and cannot be used for calibration.

Download

Catalogs