The AS-TG is an oil damping type small-sized triaxial acceleration transducer for synchronously measuring acceleration in 3 directions of back & forth, right & left, and up & down. Rated Capacity: ±9.807 to ±196.1 m/s2 Dimensions: 44 x 20 x 44 mm Rated Output: 0.5 mV/V Frequency Response: DC to (40 to 250) Hz
Features
• Synchronous measurement in directions X, Y and Z • Overload Limit 1000% due to built-in stopper • TEDS-installed versions can be manufactured.
Products table
All display
AS-1TG ±1 G
Specifications
Performance
Rated Capacity
±9.807 m/s2(±1 G)
Nonlinearity
Within ±1% RO
Hysteresis
Within ±1% RO
Rated Output
0.5 mV/V
Environmental Characteristics
Safe Temperature
-10 to 60ºC
Electrical Characteristics
Safe Excitation
6 V AC or DC
Recommended Excitation
1 to 3 V AC or DC
Input Resistance
121 Ω ±1.7%
Output Resistance
121 Ω ±1.7%
Cable
4-conductor (0.08 mm2) vinyl shielded cable, 3.2 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M, 3 cables (Shield wire is not connected to the case.)
Mechanical Properties
Safe Overloads
300%
Overload Limit
1000% (With a stopper)
Frequency Response
DC to 40 Hz at 23ºC (Sensitivity deviation ±5%)
Transverse Sensitivity
4% RO or less
Weight
Approx. 110 g (Excluding cable)
Dimensions
AS-2TG ±2 G
Specifications
Performance
Rated Capacity
±19.61 m/s2(±2 G)
Nonlinearity
Within ±1% RO
Hysteresis
Within ±1% RO
Rated Output
0.5 mV/V
Environmental Characteristics
Safe Temperature
-10 to 60ºC
Electrical Characteristics
Safe Excitation
6 V AC or DC
Recommended Excitation
1 to 3 V AC or DC
Input Resistance
121 Ω ±1.7%
Output Resistance
121 Ω ±1.7%
Cable
4-conductor (0.08 mm2) vinyl shielded cable, 3.2 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is not connected to the case.)
Mechanical Properties
Safe Overloads
300%
Overload Limit
1000% (With a stopper)
Frequency Response
DC to 60 Hz at 23ºC (Sensitivity deviation ±5%)
Transverse Sensitivity
4% RO or less
Weight
Approx. 110 g (Excluding cable)
Dimensions
AS-5TG ±5 G
Specifications
Performance
Rated Capacity
±49.03 m/s2(±5 G)
Nonlinearity
Within ±1% RO
Hysteresis
Within ±1% RO
Rated Output
0.5 mV/V
Environmental Characteristics
Safe Temperature
-10 to 60ºC
Electrical Characteristics
Safe Excitation
6 V AC or DC
Recommended Excitation
1 to 3 V AC or DC
Input Resistance
121 Ω ±1.7%
Output Resistance
121 Ω ±1.7%
Cable
4-conductor (0.08 mm2) vinyl shielded cable, 3.2 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is not connected to the case.)
Mechanical Properties
Safe Overloads
300%
Overload Limit
1000% (With a stopper)
Frequency Response
DC to 100 Hz at 23ºC (Sensitivity deviation ±5%)
Transverse Sensitivity
4% RO or less
Weight
Approx. 110 g (Excluding cable)
Dimensions
AS-10TG ±10 G
Specifications
Performance
Rated Capacity
±98.07 m/s2(±10 G)
Nonlinearity
Within ±1% RO
Hysteresis
Within ±1% RO
Rated Output
0.5 mV/V
Environmental Characteristics
Safe Temperature
-10 to 60ºC
Electrical Characteristics
Safe Excitation
6 V AC or DC
Recommended Excitation
1 to 3 V AC or DC
Input Resistance
121 Ω ±1.7%
Output Resistance
121 Ω ±1.7%
Cable
4-conductor (0.08 mm2) vinyl shielded cable, 3.2 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is not connected to the case.)
Mechanical Properties
Safe Overloads
300%
Overload Limit
1000% (With a stopper)
Frequency Response
DC to 150 Hz at 23ºC (Sensitivity deviation ±5%)
Transverse Sensitivity
4% RO or less
Weight
Approx. 110 g (Excluding cable)
Dimensions
AS-20TG ±20 G
Specifications
Performance
Rated Capacity
±196.1 m/s2(±20 G)
Nonlinearity
Within ±1% RO
Hysteresis
Within ±1% RO
Rated Output
0.5 mV/V
Environmental Characteristics
Safe Temperature
-10 to 60ºC
Electrical Characteristics
Safe Excitation
6 V AC or DC
Recommended Excitation
1 to 3 V AC or DC
Input Resistance
121 Ω ±1.7%
Output Resistance
121 Ω ±1.7%
Cable
4-conductor (0.08 mm2) vinyl shielded cable, 3.2 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is not connected to the case.)
Measurement range of up-down direction is narrowed by 1 G (9.807 m/s 2) because gravity acceleration is always applied to acceleration transducers. In terms of AS-1TG, detected value exceeds rated capacity in positive side toward gravidy direction. Although AS-TG acceleration transducers are not broked under safe overload rating, characteristics will be outside the guaranteed range.