ASQ-D
Servo Type Acceleration Transducer

Summary
The ASQ-D is a servo type acceleration transducer for accurate measuring minute vibration. A dedicated signal conditioner VAQ-700A is available as the mating instrument.
Rated Capacity: ±9.807 to ±49.03 m/s2
Dimensions: 32.5 x 75 x 32.5 mm
Rated Output: ±5 V (10 V) ±5%
Frequency Response: DC to 100 Hz
Features
• High output
• Acceleration measurement in a range of DC to 100 Hz is possible.
(If desired, a model covering a range of DC to 300 Hz is available.)
• IP67 (IEC 60529)
Block diagrams

Products table
All display
ASQ-D-1
±1 G

Specifications
Performance
Rated Capacity | ±9.807 m/s2(±1 G) |
---|---|
Nonlinearity | Within ±0.03% RO *Inspect by applying the current to a single coil. |
Hysteresis | Within ±0.05% RO *Inspect by applying the current to a single coil. |
Rated Output | ±5 V (10 V) ±5% *Inspect with ±1G (inclination). |
Environmental Characteristics
Safe Temperature | -20 to 80ºC |
---|---|
Compensated Temperature | -10 to 60ºC |
Temperature Effect on Zero | Within ±0.05% RO/ºC |
Temperature Effect on Output | Within ±0.05%/ºC *Inspect with ±1G (inclination). |
Electrical Characteristics
Power Supply | ±11 to ±16 VDC, 40 mA or less |
---|
Mechanical Properties
Frequency Response | DC to 100 Hz (Sensitivity deviation ±10%) *Inspect by applying the current to a single coil. |
---|---|
Transverse Sensitivity | 0.3% RO or less *Inspect with ±1G (inclination). |
Weight | Approx. 220 g (Excluding cable) |
Degrees of Protection | IP67 (IEC 60529) |
Standard Accessories | Hexagon socket head bolts (M4 × 25) × 2 |
---|---|
Optional Accessories | Connection cable N-41 (50 m), N-42 (5 m) 1-G cancel cable N-45 (40 cm) |
Dimensions

ASQ-D-2
±2 G

Specifications
Performance
Rated Capacity | ±19.61 m/s2(±2 G) |
---|---|
Nonlinearity | Within ±0.03% RO *Inspect by applying the current to a single coil. |
Hysteresis | Within ±0.05% RO *Inspect by applying the current to a single coil. |
Rated Output | ±5 V (10 V) ±5% *Inspect with ±1G (inclination). |
Environmental Characteristics
Safe Temperature | -20 to 80ºC |
---|---|
Compensated Temperature | -10 to 60ºC |
Temperature Effect on Zero | Within ±0.05% RO/ºC |
Temperature Effect on Output | Within ±0.05%/ºC *Inspect with ±1G (inclination). |
Electrical Characteristics
Power Supply | ±11 to ±16 VDC, 40 mA or less |
---|
Mechanical Properties
Frequency Response | DC to 100 Hz (Sensitivity deviation ±10%) *Inspect by applying the current to a single coil. |
---|---|
Transverse Sensitivity | 0.3% RO or less *Inspect with ±1G (inclination). |
Weight | Approx. 220 g (Excluding cable) |
Degrees of Protection | IP67 (IEC 60529) |
Standard Accessories | Hexagon socket head bolts (M4 × 25) × 2 |
---|---|
Optional Accessories | Connection cable N-41 (50 m), N-42 (5 m) 1-G cancel cable N-45 (40 cm) |
Dimensions

ASQ-D-5
±5 G

Specifications
Performance
Rated Capacity | ±49.03 m/s2(±5 G) |
---|---|
Nonlinearity | Within ±0.03% RO *Inspect by applying the current to a single coil. |
Hysteresis | Within ±0.05% RO *Inspect by applying the current to a single coil. |
Rated Output | ±5 V (10 V) ±5% *Inspect with ±1G (inclination). |
Environmental Characteristics
Safe Temperature | -20 to 80ºC |
---|---|
Compensated Temperature | -10 to 60ºC |
Temperature Effect on Zero | Within ±0.05% RO/ºC |
Temperature Effect on Output | Within ±0.05%/ºC *Inspect with ±1G (inclination). |
Electrical Characteristics
Power Supply | ±11 to ±16 VDC, 40 mA or less |
---|
Mechanical Properties
Frequency Response | DC to 100 Hz (Sensitivity deviation ±10%) *Inspect by applying the current to a single coil. |
---|---|
Transverse Sensitivity | 0.3% RO or less *Inspect with ±1G (inclination). |
Weight | Approx. 220 g (Excluding cable) |
Degrees of Protection | IP67 (IEC 60529) |
Standard Accessories | Hexagon socket head bolts (M4 × 25) × 2 |
---|---|
Optional Accessories | Connection cable N-41 (50 m), N-42 (5 m) 1-G cancel cable N-45 (40 cm) |
Dimensions

VAQ-700A
ASQ Dedicated Conditioner

Specifications
When using ASQ-D-1
Channels | 1 |
---|---|
Measuring Modes | Simultaneous measurement and output of acceleration, velocity and displacement |
Acceleration
Measuring Range | ±9.807 m/s2 (±1 G) |
---|---|
Range Select | 0.01, 0.1, 1 G, OFF |
Calibration | Linked with selected range × 0.51 |
Accuracy | ±0.5% FS to reading with both measuring range and calibration |
Frequency Response | DC to 300 Hz (±5%) in DC acceleration mode 0.1 to 300 Hz (±5%) in AC acceleration mode |
Velocity
Measuring Range | ±100 cm/s |
---|---|
Range Select | 1, 10, 100 cm/s, OFF |
Calibration | Linked with selected range × 0.5 |
Accuracy | ±1% FS to reading with both measuring range and calibration |
Frequency Response | 0.1 to 100 Hz (±5%) |
Displacement
Measuring Range | ±100 mm |
---|---|
Range Select | 1, 10, 100 mm, OFF |
Calibration | Linked with selected range × 0.5 |
Accuracy | ±1% FS to reading with both measuring range and calibration |
Frequency Response | 0.1 to 60 Hz (±5%) (1 to 60 Hz, ±0.5% with a maximum sensitivity selected for the measuring range) |
Output | ±10 V (Load 5 kΩ or more) |
---|---|
Operating Temperature & Humidity | -10 to 50ºC, 20 to 85% (Non-condensing) |
Power Supply | 100 VAC or 12 VDC |
Dimensions-weight | 49 W x 128.5 H x 262.5 D mm, approx. 1.4 kg |
When using ASQ-D-2
Channels | 1 |
---|---|
Measuring Modes | Simultaneous measurement and output of acceleration, velocity and displacement |
Acceleration
Measuring Range | ±19.61 m/s2(±2 G) |
---|---|
Range Select | 0.02, 0.2, 2 G, OFF |
Calibration | Linked with selected range x 0.51 |
Accuracy | ±0.5% FS to reading with both measuring range and calibration |
Frequency Response | DC to 300 Hz (±5%) in DC acceleration mode 0.1 to 300 Hz (±5%) in AC acceleration mode |
Velocity
Measuring Range | ±200 cm/s |
---|---|
Range Select | 2, 20, 200 cm/s, OFF |
Calibration | Linked with selected range x 0.5 |
Accuracy | ±1% FS to reading with both measuring range and calibration |
Frequency Response | 0.1 to 100 Hz (±5%) |
Displacement
Measuring Range | ±200 mm |
---|---|
Range Select | 2, 20, 200 mm, OFF |
Calibration | Linked with selected range x 0.5 |
Accuracy | ±1% FS to reading with both measuring range and calibration |
Frequency Response | 0.1 to 60 Hz (±5%) (1 to 60 Hz, ±0.5% with a maximum sensitivity selected for the measuring range) |
Output | ±10 V (Load 5 kΩ or more) |
---|---|
Operating Temperature & Humidity | -10 to 50ºC, 20 to 85% (Non-condensing) |
Power Supply | 100 VAC or 12 VDC |
Dimensions-weight | 49 W x 128.5 H x 262.5 D mm, approx. 1.4 kg |
When using ASQ-D-5
Channels | 1 |
---|---|
Measuring Modes | Simultaneous measurement and output of acceleration, velocity and displacement |
Acceleration
Measuring Range | ±49.03 m/s2(±5 G) |
---|---|
Range Select | 0.05, 0.5, 5 G, OFF |
Calibration | Linked with selected range x 0.51 |
Accuracy | ±0.5% FS to reading with both measuring range and calibration |
Frequency Response | DC to 300 Hz (±5%) in DC acceleration mode 0.1 to 300 Hz (±5%) in AC acceleration mode |
Velocity
Measuring Range | ±500 cm/s |
---|---|
Range Select | 5, 50, 500 cm/s, OFF |
Calibration | Linked with selected range x 0.5 |
Accuracy | ±1% FS to reading with both measuring range and calibration |
Frequency Response | 0.1 to 100 Hz (±5%) |
Displacement
Measuring Range | ±500 mm |
---|---|
Range Select | 5, 50, 500 mm, OFF |
Calibration | Linked with selected range x 0.5 |
Accuracy | ±1% FS to reading with both measuring range and calibration |
Frequency Response | 0.1 to 60 Hz (±5%) (1 to 60 Hz, ±0.5% with a maximum sensitivity selected for the measuring range) |
Output | ±10 V (Load 5 kΩ or more) |
---|---|
Operating Temperature & Humidity | -10 to 50ºC, 20 to 85% (Non-condensing) |
Power Supply | 100 VAC or 12 VDC |
Dimensions-weight | 49 W x 128.5 H x 262.5 D mm, approx. 1.4 kg |
Standard Accessories | AC power cable P-25 (With 2-pin conversion plug CM-52) Output cable U-59 × 3 Shorting bar Midget type fuse (0.5 A) Midget type fuse (3 A) Miniature screwdriver Instruction manual |
---|---|
Optional Accessories | Portable housing case YC-A Amplifier stand FA-1B |
Dimensions