ASQ-D
Servo Type Acceleration Transducer

Summary

The ASQ-D is a servo type acceleration transducer for accurate measuring minute vibration. A dedicated signal conditioner VAQ-700A is available as the mating instrument.

Rated Capacity: ±9.807 to ±49.03 m/s2
Dimensions: 32.5 x 75 x 32.5 mm
Rated Output: ±5 V (10 V) ±5%
Frequency Response: DC to 100 Hz

Features

• High output
• Acceleration measurement in a range of DC to 100 Hz is possible.
 (If desired, a model covering a range of DC to 300 Hz is available.)
• IP67 (IEC 60529)

Block diagrams

ASQ-D

Products table

All display

ASQ-D-1
±1 G

Specifications

Performance
Rated Capacity ±9.807 m/s2(±1 G)
Nonlinearity Within ±0.03% RO
*Inspect by applying the current to a single coil.
Hysteresis Within ±0.05% RO
*Inspect by applying the current to a single coil.
Rated Output ±5 V (10 V) ±5%
*Inspect with ±1G (inclination).
Environmental Characteristics
Safe Temperature -20 to 80ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.05% RO/ºC
Temperature Effect on Output Within ±0.05%/ºC
*Inspect with ±1G (inclination).
Electrical Characteristics
Power Supply ±11 to ±16 VDC, 40 mA or less
Mechanical Properties
Frequency Response DC to 100 Hz (Sensitivity deviation ±10%)
*Inspect by applying the current to a single coil.
Transverse Sensitivity 0.3% RO or less
*Inspect with ±1G (inclination).
Weight Approx. 220 g (Excluding cable)
Degrees of Protection IP67 (IEC 60529)
Standard Accessories Hexagon socket head bolts (M4 × 25) × 2
Optional Accessories Connection cable N-41 (50 m), N-42 (5 m)
1-G cancel cable N-45 (40 cm)

Dimensions

ASQ-D

ASQ-D-2
±2 G

Specifications

Performance
Rated Capacity ±19.61 m/s2(±2 G)
Nonlinearity Within ±0.03% RO
*Inspect by applying the current to a single coil.
Hysteresis Within ±0.05% RO
*Inspect by applying the current to a single coil.
Rated Output ±5 V (10 V) ±5%
*Inspect with ±1G (inclination).
Environmental Characteristics
Safe Temperature -20 to 80ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.05% RO/ºC
Temperature Effect on Output Within ±0.05%/ºC
*Inspect with ±1G (inclination).
Electrical Characteristics
Power Supply ±11 to ±16 VDC, 40 mA or less
Mechanical Properties
Frequency Response DC to 100 Hz (Sensitivity deviation ±10%)
*Inspect by applying the current to a single coil.
Transverse Sensitivity 0.3% RO or less
*Inspect with ±1G (inclination).
Weight Approx. 220 g (Excluding cable)
Degrees of Protection IP67 (IEC 60529)
Standard Accessories Hexagon socket head bolts (M4 × 25) × 2
Optional Accessories Connection cable N-41 (50 m), N-42 (5 m)
1-G cancel cable N-45 (40 cm)

Dimensions

ASQ-D

ASQ-D-5
±5 G

Specifications

Performance
Rated Capacity ±49.03 m/s2(±5 G)
Nonlinearity Within ±0.03% RO
*Inspect by applying the current to a single coil.
Hysteresis Within ±0.05% RO
*Inspect by applying the current to a single coil.
Rated Output ±5 V (10 V) ±5%
*Inspect with ±1G (inclination).
Environmental Characteristics
Safe Temperature -20 to 80ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.05% RO/ºC
Temperature Effect on Output Within ±0.05%/ºC
*Inspect with ±1G (inclination).
Electrical Characteristics
Power Supply ±11 to ±16 VDC, 40 mA or less
Mechanical Properties
Frequency Response DC to 100 Hz (Sensitivity deviation ±10%)
*Inspect by applying the current to a single coil.
Transverse Sensitivity 0.3% RO or less
*Inspect with ±1G (inclination).
Weight Approx. 220 g (Excluding cable)
Degrees of Protection IP67 (IEC 60529)
Standard Accessories Hexagon socket head bolts (M4 × 25) × 2
Optional Accessories Connection cable N-41 (50 m), N-42 (5 m)
1-G cancel cable N-45 (40 cm)

Dimensions

ASQ-D

VAQ-700A
ASQ Dedicated Conditioner

VAQ-700A

Specifications

When using ASQ-D-1
Channels 1
Measuring Modes Simultaneous measurement and output of acceleration, velocity and displacement
Acceleration
Measuring Range ±9.807 m/s2 (±1 G)
Range Select 0.01, 0.1, 1 G, OFF
Calibration Linked with selected range × 0.51
Accuracy ±0.5% FS to reading with both measuring range and calibration
Frequency Response DC to 300 Hz (±5%) in DC acceleration mode
0.1 to 300 Hz (±5%) in AC acceleration mode
Velocity
Measuring Range ±100 cm/s
Range Select 1, 10, 100 cm/s, OFF
Calibration Linked with selected range × 0.5
Accuracy ±1% FS to reading with both measuring range and calibration
Frequency Response 0.1 to 100 Hz (±5%)
Displacement
Measuring Range ±100 mm
Range Select 1, 10, 100 mm, OFF
Calibration Linked with selected range × 0.5
Accuracy ±1% FS to reading with both measuring range and calibration
Frequency Response 0.1 to 60 Hz (±5%)
(1 to 60 Hz, ±0.5% with a maximum sensitivity selected for the measuring range)
Output ±10 V (Load 5 kΩ or more)
Operating Temperature & Humidity -10 to 50ºC, 20 to 85% (Non-condensing)
Power Supply 100 VAC or 12 VDC
Dimensions-weight 49 W x 128.5 H x 262.5 D mm, approx. 1.4 kg
When using ASQ-D-2
Channels 1
Measuring Modes Simultaneous measurement and output of acceleration, velocity and displacement
Acceleration
Measuring Range ±19.61 m/s2(±2 G)
Range Select 0.02, 0.2, 2 G, OFF
Calibration Linked with selected range x 0.51
Accuracy ±0.5% FS to reading with both measuring range and calibration
Frequency Response DC to 300 Hz (±5%) in DC acceleration mode
0.1 to 300 Hz (±5%) in AC acceleration mode
Velocity
Measuring Range ±200 cm/s
Range Select 2, 20, 200 cm/s, OFF
Calibration Linked with selected range x 0.5
Accuracy ±1% FS to reading with both measuring range and calibration
Frequency Response 0.1 to 100 Hz (±5%)
Displacement
Measuring Range ±200 mm
Range Select 2, 20, 200 mm, OFF
Calibration Linked with selected range x 0.5
Accuracy ±1% FS to reading with both measuring range and calibration
Frequency Response 0.1 to 60 Hz (±5%)
(1 to 60 Hz, ±0.5% with a maximum sensitivity selected for the measuring range)
Output ±10 V (Load 5 kΩ or more)
Operating Temperature & Humidity -10 to 50ºC, 20 to 85% (Non-condensing)
Power Supply 100 VAC or 12 VDC
Dimensions-weight 49 W x 128.5 H x 262.5 D mm, approx. 1.4 kg
When using ASQ-D-5
Channels 1
Measuring Modes Simultaneous measurement and output of acceleration, velocity and displacement
Acceleration
Measuring Range ±49.03 m/s2(±5 G)
Range Select 0.05, 0.5, 5 G, OFF
Calibration Linked with selected range x 0.51
Accuracy ±0.5% FS to reading with both measuring range and calibration
Frequency Response DC to 300 Hz (±5%) in DC acceleration mode
0.1 to 300 Hz (±5%) in AC acceleration mode
Velocity
Measuring Range ±500 cm/s
Range Select 5, 50, 500 cm/s, OFF
Calibration Linked with selected range x 0.5
Accuracy ±1% FS to reading with both measuring range and calibration
Frequency Response 0.1 to 100 Hz (±5%)
Displacement
Measuring Range ±500 mm
Range Select 5, 50, 500 mm, OFF
Calibration Linked with selected range x 0.5
Accuracy ±1% FS to reading with both measuring range and calibration
Frequency Response 0.1 to 60 Hz (±5%)
(1 to 60 Hz, ±0.5% with a maximum sensitivity selected for the measuring range)
Output ±10 V (Load 5 kΩ or more)
Operating Temperature & Humidity -10 to 50ºC, 20 to 85% (Non-condensing)
Power Supply 100 VAC or 12 VDC
Dimensions-weight 49 W x 128.5 H x 262.5 D mm, approx. 1.4 kg
Standard Accessories AC power cable P-25 (With 2-pin conversion plug CM-52)
Output cable U-59 × 3
Shorting bar
Midget type fuse (0.5 A)
Midget type fuse (3 A)
Miniature screwdriver
Instruction manual
Optional Accessories Portable housing case YC-A
Amplifier stand FA-1B

Dimensions

VAQ-700A

Page Top