The DT-D is a displacement transducer equiped with dial enables to visual measured displacement. Rated Capacity: 10, 20, 30, 50 mm Nonlinearity: Within ±0.5% RO Measuring Force: Approx. 1.7 N (10 mm), approx. 2.1 N (20 mm), approx. 2.2 N (30 mm), approx. 2.7 N (50mm) Rated Output: 1.5 mV/V or more
Features
• Easy installation, handling & maintenance • Temperature effect on zero within 0.03% RO per ºC • TEDS-installed versions can be manufactured.
Products table
All display
DT-10D 10 mm
Specifications
Performance
Rated Capacity
10 mm
Nonlinearity
Within ±0.5% RO
Hysteresis
Within ±0.5% RO
Repeatability
0.3% RO or less
Rated Output
1.5 mV/V or more
Environmental Characteristics
Safe Temperature
0 to 55ºC (Non-condensing)
Compensated Temperature
0 to 50ºC (Non-condensing)
Temperature Effect on Zero
Within ±0.03% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
12 V AC or DC
Recommended Excitation
1 to 4 V AC or DC
Input Resistance
350 Ω ±2%
Output Resistance
350 Ω ±2%
Cable
4-conductor (0.08 mm2) chloroprene shielded cable, 4 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
100%
Frequency Response
DC to approx. 0.8Hz
Measuring Force
Approx. 1.7 N
Weight
Approx. 160 g (Excluding cable)
Optional Accessories
Replacement probes X, XS, SH Magnet base MB-B
Dimensions
DT-20D 20 mm
Specifications
Performance
Rated Capacity
20 mm
Nonlinearity
Within ±0.5% RO
Hysteresis
Within ±0.5% RO
Repeatability
0.3% RO or less
Rated Output
1.5 mV/V or more
Environmental Characteristics
Safe Temperature
0 to 55ºC (Non-condensing)
Compensated Temperature
0 to 50ºC (Non-condensing)
Temperature Effect on Zero
Within ±0.03% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
12 V AC or DC
Recommended Excitation
1 to 4 V AC or DC
Input Resistance
350 Ω ±2%
Output Resistance
350 Ω ±2%
Cable
4-conductor (0.08 mm2) chloroprene shielded cable, 4 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
100%
Frequency Response
DC to approx. 0.8Hz
Measuring Force
Approx. 2.1 N
Weight
Approx. 310 g (Excluding cable)
Optional Accessories
Replacement probes X, XS, SH Magnet base MB-B
Dimensions
DT-30D M150 30 mm
Specifications
Performance
Rated Capacity
30 mm
Nonlinearity
Within ±0.5% RO
Hysteresis
Within ±0.5% RO
Repeatability
0.3% RO or less
Rated Output
1.5 mV/V or more
Environmental Characteristics
Safe Temperature
0 to 55ºC (Non-condensing)
Compensated Temperature
0 to 50ºC (Non-condensing)
Temperature Effect on Zero
Within ±0.03% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
12 V AC or DC
Recommended Excitation
1 to 4 V AC or DC
Input Resistance
350 Ω ±2%
Output Resistance
350 Ω ±2%
Cable
4-conductor (0.08 mm2) chloroprene shielded cable, 4 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
100%
Frequency Response
DC to approx. 0.8Hz
Measuring Force
Approx. 2.2 N
Weight
Approx. 260 g (Excluding cable)
Optional Accessories
Replacement probes X, XS, SH Magnet base MB-B
Dimensions
DT-50D M150 50 mm
Specifications
Performance
Rated Capacity
50 mm
Nonlinearity
Within ±0.5% RO
Hysteresis
Within ±0.5% RO
Repeatability
0.3% RO or less
Rated Output
1.5 mV/V or more
Environmental Characteristics
Safe Temperature
0 to 55ºC (Non-condensing)
Compensated Temperature
0 to 50ºC (Non-condensing)
Temperature Effect on Zero
Within ±0.03% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
12 V AC or DC
Recommended Excitation
1 to 4 V AC or DC
Input Resistance
350 Ω ±2%
Output Resistance
350 Ω ±2%
Cable
4-conductor (0.08 mm2) chloroprene shielded cable, 4 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
(Note 1) Avoid usage in vibration. (Note 2) If large displacement is applied momentarily, it takes some time that output is settled. (Note 3) Do not apply any displacement in other than expansion/contraction direction of the rod. (Note 4) If the DT-50D M150 is used in horizontal position, the rod inclines by approximately 10 mm due to its own weight and may not follow displacement.
Remarks
To Ensure Safe Usage • Fix the transducer to a steady point by the M6 bolt. • DT-D series transducers are designed to output signal proportional to the magnitude of displacement, while indicating it on the dial gage.