DTJ-A-200
Displacement Transducer

Summary
The DTJ-A-200 is a displacement transducer for both tension and compression measurement. It can measure up to 200 mm displacement.
Rated Capacity: 200 mm
Nonlinearity: Within ±0.3% RO
Measuring Force: Approx. 5.9 N
Rated Output: 5 mV/V ±0.3%
Features
• High output 5 mV/V
• Temperature effect on zero within ±0.02% RO per ºC
• With scale
• TEDS-installed versions can be manufactured.
Products table
All display
DTJ-A-200
200 mm

Specifications
Performance
Rated Capacity | 200 mm |
---|---|
Nonlinearity | Within ±0.3% RO |
Hysteresis | Within ±0.3% RO |
Repeatability | 0.3% RO or less |
Rated Output | 5 mV/V ±0.3% |
Environmental Characteristics
Safe Temperature | -10 to 70ºC (Non-condensing) |
---|---|
Compensated Temperature | 0 to 60ºC (Non-condensing) |
Temperature Effect on Zero | Within ±0.02% RO/ºC |
Temperature Effect on Output | Within ±0.02%/ºC |
Electrical Characteristics
Safe Excitation | 6 V AC or DC |
---|---|
Recommended Excitation | 1 to 4 V AC or DC |
Input Resistance | 350 Ω ±1% |
Output Resistance | 350 Ω ±1% |
Cable | 4-conductor (0.065 mm2) vinyl shielded cable, 4 mm diameter by 2 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is not connected to the case.) |
Mechanical Properties
Frequency Response | DC to approx. 2 Hz |
---|---|
Measuring Force | Approx. 5.9 N |
Weight | Approx. 560 g (Excluding cable) |
Optional Accessories | Extension rods EB-50, EB-100, EB-200, EB-300 Replacement probes X, XS, SH Magnet base MB-B |
---|
Dimensions

To Ensure Safe Usage
Fix the transducer to the steady point using two M6 bolts.
Remarks
(Note 1) Initial unbalance with the rod fully extended is approximately-5000 to -6000 ×10-6 strain.
(Note 2) Avoid usage in vibration.
(Note 3) If large displacement is applied momentarily, it takes some time that output is settled.
(Note 4) Do not apply any displacement in other than expansion/ contraction direction of the rod.