The PD-A is a differential pressure transducer for accurate measuring slight differential pressures. The max line pressure is 2.94 MPa. It can be used not only for differential pressure measurement, but also flow rate measurement. Rated Capacity: 10 to 200 kPa Compensated Temperature: 0 to 60ºC
Features
• Suitable for long-term measurement • For indication and control of automation systems • TEDS-installed versions can be manufactured.
Products table
All display
PD-100GA 10 kPa
Specifications
Performance
Rated Capacity
10 kPa
Nonlinearity
Within ±0.3% RO
Hysteresis
Within ±0.2% RO
Rated Output
1.5 mV/V ±1%
Environmental Characteristics
Safe Temperature
-10 to 70ºC
Compensated Temperature
0 to 60ºC
Temperature Effect on Zero
Within ±0.05% RO/ºC
Temperature Effect on Output
Within ±0.5%/ºC
Electrical Characteristics
Safe Excitation
15 V AC or DC
Recommended Excitation
1 to 10 V AC or DC
Input Resistance
350 Ω ±0.5%
Output Resistance
350 Ω ±0.5%
Cable
4-conductor (0.3 mm2) chloroprene shielded cable, 7.6 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
Differential pressure 200%
Natural Frequencies
Approx. 60 Hz
Maximum Line Pressure
2.94 MPa
Weight
Approx. 5 kg (Excluding cable)
Dimensions
PD-200GA 20 kPa
Specifications
Performance
Rated Capacity
20 kPa
Nonlinearity
Within ±0.3% RO
Hysteresis
Within ±0.2% RO
Rated Output
1.5 mV/V ±1%
Environmental Characteristics
Safe Temperature
-10 to 70ºC
Compensated Temperature
0 to 60ºC
Temperature Effect on Zero
Within ±0.01% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
15 V AC or DC
Recommended Excitation
1 to 10 V AC or DC
Input Resistance
350 Ω ±0.5%
Output Resistance
350 Ω ±0.5%
Cable
4-conductor (0.3 mm2) chloroprene shielded cable, 7.6 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
Differential pressure 200%
Natural Frequencies
Approx. 110 Hz
Maximum Line Pressure
2.94 MPa
Weight
Approx. 5 kg (Excluding cable)
Dimensions
PD-500GA 50 kPa
Specifications
Performance
Rated Capacity
50 kPa
Nonlinearity
Within ±0.3% RO
Hysteresis
Within ±0.2% RO
Rated Output
1.5 mV/V ±1%
Environmental Characteristics
Safe Temperature
-10 to 70ºC
Compensated Temperature
0 to 60ºC
Temperature Effect on Zero
Within ±0.01% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
15 V AC or DC
Recommended Excitation
1 to 10 V AC or DC
Input Resistance
350 Ω ±0.5%
Output Resistance
350 Ω ±0.5%
Cable
4-conductor (0.3 mm2) chloroprene shielded cable, 7.6 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
Differential pressure 150%
Natural Frequencies
Approx. 230 Hz
Maximum Line Pressure
2.94 MPa
Weight
Approx. 5 kg (Excluding cable)
Dimensions
PD-1KA 100 kPa
Specifications
Performance
Rated Capacity
100 kPa
Nonlinearity
Within ±0.3% RO
Hysteresis
Within ±0.2% RO
Rated Output
1.5 mV/V ±1%
Environmental Characteristics
Safe Temperature
-10 to 70ºC
Compensated Temperature
0 to 60ºC
Temperature Effect on Zero
Within ±0.01% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
15 V AC or DC
Recommended Excitation
1 to 10 V AC or DC
Input Resistance
350 Ω ±0.5%
Output Resistance
350 Ω ±0.5%
Cable
4-conductor (0.3 mm2) chloroprene shielded cable, 7.6 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads
Differential pressure 150%
Natural Frequencies
Approx. 400 Hz
Maximum Line Pressure
2.94 MPa
Weight
Approx. 5 kg (Excluding cable)
Dimensions
PD-2KA 200 kPa
Specifications
Performance
Rated Capacity
200 kPa
Nonlinearity
Within ±0.3% RO
Hysteresis
Within ±0.2% RO
Rated Output
1.5 mV/V ±1%
Environmental Characteristics
Safe Temperature
-10 to 70ºC
Compensated Temperature
0 to 60ºC
Temperature Effect on Zero
Within ±0.01% RO/ºC
Temperature Effect on Output
Within ±0.03%/ºC
Electrical Characteristics
Safe Excitation
15 V AC or DC
Recommended Excitation
1 to 10 V AC or DC
Input Resistance
350 Ω ±0.5%
Output Resistance
350 Ω ±0.5%
Cable
4-conductor (0.3 mm2) chloroprene shielded cable, 7.6 mm diameter by 5 m long, terminated with a connector plug PRC03-12A10-7M (Shield wire is connected to the case.)
Copper alloy is used for sensing element. Avoid measuring corrosive liquid or gas.
Remarks
*To use for gas, contact us. When A side is a high pressure, plus output. When B side is a high pressure, minus output. (For A-side and B-side, see the dimensions.)