PG-U
Pressure Transducer

PG-U

Summary

The PG-U is a pressure transducer with high accuracy and reliability because of hermetically-sealed structure with inert gas filled in.
Rated Capacity: 200 kPa to 50 MPa
Compensated Temperature: -10 to 60ºC

Features

• High accuracy
• Various capacity
• Plenty of application results

Products table

All display

PG-2KU
200 kPa

PG-U

Specifications

Performance
Rated Capacity 200 kPa
Nonlinearity Within ±0.3% RO
Hysteresis Within ±0.3% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±1%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M

(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 2 kHz
Material Case: Anode oxidized coated aluminum; Liquid-contacting part: SUS 630
Weight Approx. 500 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-5KU
500 kPa

PG-U

Specifications

Performance
Rated Capacity 500 kPa
Nonlinearity Within ±0.3% RO
Hysteresis Within ±0.3% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±1%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 4 kHz
Material Case: Anode oxidized coated aluminum; Liquid-contacting part: SUS 630
Weight Approx. 500 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-10KU
1 MPa

PG-U

Specifications

Performance
Rated Capacity 1 MPa
Nonlinearity Within ±0.3% RO
Hysteresis Within ±0.3% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±1%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 7 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-20KU
2 MPa

PG-U

Specifications

Performance
Rated Capacity 2 MPa
Nonlinearity Within ±0.2% RO
Hysteresis Within ±0.2% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±0.5%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 13 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-50KU
5 MPa

PG-U

Specifications

Performance
Rated Capacity 5 MPa
Nonlinearity Within ±0.2% RO
Hysteresis Within ±0.2% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±0.5%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 21 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-100KU
10 MPa

PG-U

Specifications

Performance
Rated Capacity 10 MPa
Nonlinearity Within ±0.2% RO
Hysteresis Within ±0.2% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±0.5%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 29 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-200KU
20 MPa

PG-U

Specifications

Performance
Rated Capacity 20 MPa
Nonlinearity Within ±0.2% RO
Hysteresis Within ±0.2% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±0.5%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 40 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-300KU
30 MPa

PG-U

Specifications

Performance
Rated Capacity 30 MPa
Nonlinearity Within ±0.2% RO
Hysteresis Within ±0.2% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±0.5%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 45 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

PG-500KU
50 MPa

PG-U

Specifications

Performance
Rated Capacity 50 MPa
Nonlinearity Within ±0.2% RO
Hysteresis Within ±0.2% RO
Repeatability 0.1% RO or less
Rated Output 2 mV/V ±0.5%
Environmental Characteristics
Safe Temperature -20 to 70ºC
Compensated Temperature -10 to 60ºC
Temperature Effect on Zero Within ±0.02% RO/ºC
Temperature Effect on Output Within ±0.02%/ºC
Electrical Characteristics
Safe Excitation 15 V AC or DC
Recommended Excitation 1 to 10 V AC or DC
Input Resistance 350 Ω ±1%
Output Resistance 350 Ω ±1%
Accessory Cable(TT-01) 4-conductor (0.3 mm2) chloroprene shielded cable,
7.6 mm diameter by 3 m long
Sensor side: Terminated with a waterproof connector plug 1108-12A10-7M
Measuring instrument side: Terminated with a connector plug PRC03-12A10-7M
(Shield wire is connected to the case.)
Mechanical Properties
Safe Overloads 150%
Natural Frequencies Approx. 50 kHz
Material Case: Die cast zinc alloy.
Liquid-contacting part: SUS 630
Weight Approx. 300 g (Excluding cable)
Degree of Protection IP54 (IEC 60529)
Mounting Screw G3/8, male
Standard Accessories Gasket (Mild copper)

Dimensions

PG-U

Page Top

To Ensure Safe Usage

Do not use PG-200KU to PG-500KU for endurance/fatigue tests.
Avoid using for a long-term measurement of gas pressure if much importance is attached to the stability of output in a minute range.

Remarks

*CE compliant models are available. Inquiries are welcome.

eBook

Download

Recommended products for combination