UCS-60B
Control software

UCS-60B

Summary

User friendly of control software for data logger.
The UCS-60B enables the PC to control a data logger and to present measured/calculated data on graph and numeric windows, thereby enhancing the performance of the data logger.

Features

●Can control UCAM-60B/65B.
●Numeric window presenting data in list format
●Up to 50 graph windows on display, maximum 20 channels of data per graph
●Various data saving formats: Kyowa standard KU1, UCAM-70A, CSV and XLS (Excel format)
●Data processing (arithmetic operation, statistic operation and rosette analysis)
●Read/write of measuring/calculating condition files
●Printer output

Products table

All display

UCS-60B
Control Software

UCS-60B

Specifications

PC Requirements
OS Windows® Vista, 7, 8/8.1, English/Japanese 32, 64 bits support
CPU Core2Duo, 2 GHz or advanced
Memory If 32-bit OS, 2 GB or more
If 64-bit OS, 4 GB or more
Display 1024 x 768 pixels or more
HDD At installation, 10 MB + measurement data storage space
Interfaces UCAM-60B RS-232C/100BASE-TX or advanced
UCAM-65B RS-232C/100BASE-TX or advanced
Measurement-related Functions
Controllable Data Loggers UCAM-60B, UCAM-65B
Measuring Channels 2000 (Within 2000 channels in total of measuring, temperature and calculating channels)
Measuring Condition Setting Functions System setting (Setting of internal/ external scanners, etc.)
Measuring channel range, measuring function and scanning speed depend on the applied data logger. 
Measuring Channels Range 000 to 999
Measuring Functions EASY MEAS., MEASURE VALUE, ORIGINAL VALUE, INITIAL VALUE
Repeat Times 0 to 999 (0: Infinite)
Calibration Coefficient Calculation ON/OFF setting possible
Channel Conditions Type of scanner, measuring channel mode, calibration coefficient, number of digits below decimal point, unit, offset, temperature reference value, initial value, scanning speed, channel name (Within 18 alphanumerics) 
Interval Measurement Conditions Starting date/time, interval, measuring times (0 to 999; 0 = Infinite), steps (Up to 99)
Trigger Measurement Conditions Trigger channels (Desired 4 channels), reference values of trigger channels, AND/OR between trigger channels, trigger values, measuring times (0 to 999; 0 = infinite), steps (Up to 99)
Trigger Interval Measurement Conditions  Trigger channels (Desired 4 channels), reference values of trigger channels, AND/OR between trigger channels, interval, measuring times (0 to 999; 0 = infinite), steps (Up to 99)
TEDS Reads sensor's information and sets to channel condition automatically.
Automatic Reading of Channel Mode Possible for strain gages and strain-gage transducers connected to internal scanners of UCAM-60B and UCAM-65B
Calculation Condition File Reading and saving possible
Measuring Functions Measurement check, initial value measurement, monitor measurement (Max. 40 channels), real-time measurement, automatic measurement (Interval, trigger, trigger interval), stroke change (Single channel/measuring channel range)
 Note: Monitor, trigger or trigger interval measurement is used for setting calculation target channels.
Numeric Display Real-time measurement and automatic measurement results in a list, results of measurement check, initial value measurement and monitor measurement in a list only
 Numeric windows: Max. 1
 Monitoring windows: Max. 1
Graph Display Types of graph: Y-time, Y-Cycle, X-Y, bar graph, 1 channel/ graph, 20 channels/graph
Display channels: Max. 20 (Max. 10 sets of channels with X-Y graph) Cursor indication, scale enlargement,
auto scale, scale setting for each individual channel,
comparison graphs
Graph windows: Max. 50 Number of measured data
available on display depends on the number of channels as following.
100 channels for 10000
200 channels for 5000
500 channels for 2000
1000 channels for 1000
1001 channels for 500
(Max. monitoring measurement data available on display is 1000.)
Data Formats Kyowa standard KU1, CSV, XLS (Excel)
Printing Numeric and graphic data are printed out.
The built-in printer of UCAM-60B is set to ON or OFF.
Reproduction-related Functions
Channels 2000 (Within 2000 channels in total of measuring, temperature and calculating channels)
Reproducible File Formats KU1
In addition to reproduction of data saved in the formats, the software enables coupling of files in the same format, extracting of a desired portion, and converting to CSV, or XLS format.
Reading/Saving Calculation Condition File Possible
Numeric Indication of Measured Data Numeric window where measured data is listed and is edited as desired.
Numeric available on single screen: Max. 1
Graphic Indication of Measured Data Graph window: Y-time, Y-Cycle, X-Y 1 channel/graph, 20 channels/graph
Display Channels: Max. 20 (Max. 10 sets of channels with X-Y graph)
Cursor indication, scale enlargement, auto scale, scale setting for each individual channel
Graph Windows available on single screen: Max. 50
Reading/Saving Display Condition File Possible
Print Function Numeric and graph windows are printed out.
Setting Calculation-related Functions
Number of Characters in Expression 100
Operators +, -, *, /, (, ), ^
Applicable Function MAX To obtain the maximum value among channels
MIN To obtain the minimum value among channels
SUM To obtain the sum of data in all channels
AVG To obtain the average of data in all channels
STD To obtain the standard deviation in all channels
DEV To obtain the standard deviation in %
MAT To obtain the maximum value in a channel
MIT To obtain the minimum value in a channel
SUT To obtain the sum of data in a channel
AVT To obtain the average of data in a channel
STT To obtain the standard deviation in a channel
PRE To obtain the previous data in a channel counting
CNT To obtain the measuring times
HMX To obtain the maximum principal strain
HMN To obtain the minimum principal strain
HSM To obtain the maximum shearing strain
SMX To obtain the maximum principal stress
SMN To obtain the minimum principal stress
SSM To obtain the maximum shearing stress
DEG To obtain principal strain direction
SIN Sine
COS Cosine
TAN Tangent
ASI Arc sine
ACO Arc cosine
ATA Arc tangent
Restrictions To save measured data in the XLS format or to convert the measured data file into an XLS format file, the channels and the measured values are restricted as follows: 
 Channels: Max. 250
 Measured Values: Max. 10000

Page Top

Remarks

When connecting the data logger to the PC via LAN port,
use a straight cable and LAN hub.