Upgrade history UCAM-60C M14
Ver.01.08
Fixed the following bugs.
- When the language settings was English and when you executed the “Convert File” on the [File Manage], each unit row shifted to the next row.
(Note that the UCAM-60C worked properly when the language settings was Japanese.) - When you executed the thermocouple measurement, the UCAM-60C did not measure data by the Scan Speed of 0.5 sec, 1 sec, 2 sec, 5 sec, or 10 sec.
(Note that the UCAM-60C measured data correctly when the Scan Speed was Normal or 0.28 sec.) - When you connected the 5-wire LVDT or 5-wire LVDT/T (Sakata Denki Co., Ltd.), the UCAM-60C did not measure data by the Scan Speed of Normal, 0.28 sec, 0.5 sec, or 1 sec.
(Note that the UCAM-60C measured data correctly when the Scan Speed was 2 sec, 5 sec, or 10 sec.) - When you pressed the SET key, etc. for checking conditions, etc. during the interval measurement state, the UCAM-60C entered standby mode and reset the measurement count to 0.
(Note that the UCAM-60C measured data correctly even through the measurement count was 0.)
Ver.01.07
- Fixed the following bug.
When the “Monitor Font” was “Large” during monitor measurements, the characters overlapped with each other on the windows.
*The display glitch occurred on the window only. Note that the unit monitored data correctly. - Fixed the following bug.
When the unit was AC power supply device, the “Asynchronous” couldn’t be selected on the “MEAS Timing”. - Fixed the following bug.
When you frequently executed the “CAL” command (start monitor measurement) and “STP” command (interrupt or stop measurement), some unit froze.
Ver.01.06
Fixed the following bug.
- We corrected the following bug: When the display was set to English, the displayed text sometimes overlapped and resulted in a distorted screen during file management.
* This bug only causes screen display distortion and does not prevent normal operation. - Incorrect data very rarely appeared as the real-time measurement data, when the number of “Monitor CH” is 1, the “MEAS CH” is the same as the “Monitor CH”, and when you performed the real-time measurement at freely-selectable timing during the monitor measurement.
Ver.01.04
1. Fixed the following bug.
When controlling the UCAM-60C by using commands with the [Buzzer: ON], the communication error occurred after you sent commands and buzzer sounded approx. 400 times.
*To recover the communication, turn OFF and ON the UCAM-60C.
●The commands which makes buzzer sounds are described as follows.
- STA (Start measurement.) command with the [LOCK] LED lights out.
- CAL (Start monitor measurement.) command with the [LOCK] LED lights out.
- CHK (Checking conditions.) command with the [LOCK] LED lights out.
- ION (Turn interval measurement ON.) command with the [LOCK] LED lights out.
- SCS (Match measursing conditions between mainframe and scanners.) command with the [LOCK] LED lights out.
- Above commands + LOC (Exit the on-line status.) command.
[Example]
This trouble occurred after sending the following commands approx. 400 times with the [LOCK] LED lights out.
1. Send the ION command. (The buzzer sounds and the [LOCK] LED lights up.) → 1st buzzer
2. Send the SCS command. (The buzzer does not sounds.)
3.Send the LOC command. (The buzzer sounds and the [LOCK] LED lights out.) → 2nd buzzer
*This trouble does not occur when the [7: System Control] → [0: Environment] → [1: Buzzer] is OFF.
*The initial status is [Buzzer: OFF].
*This trouble does not occur when using the UCS-60B (control software for UCAM).
This trouble occurred when using the custom-made software.
2. Fixed the following bug.
Could not turn ON the Auto-MEAS (Interval, Trigger, Trigger interval) from the UCAM-60C.
Note that you are able to turn ON/OFF the Auto-MEAS by the following operations.。
- Press the [0] key for approx. more than 3 seconds in ready state. → [Interval: ON]
- Press the [0] key for approx. more than 3 seconds when waiting for the measurement start. → [Interval: OFF]
- Press the [.] key for approx. more than 3 seconds in ready state. → [Trigger: ON]
- Press the [.] key for approx. more than 3 seconds during trigger measurements.→ [Trigger: OFF]
- Press the [-] key for approx. more than 3 seconds in ready state. → [Trigger interval: ON]
- Press the [-] key for approx. more than 3 seconds during trigger interval measurements. → [Trigger interval: OFF]
Ver.01.03
Fixed the following bug.
- The UCAM-60C did not stop measuring data by sending the STP (Stop measurement or setting.) command.