PCD-400A
Sensor Interface

  • CE
  • TEDS

PCD-400A

Summary

· Synchronous sampling in 4 units connected by an option Connection Connector Set
· TEDS compliant when using Input Adapter UI-10A or UI-11A
· Max sampling frequency 10 kHz

Features

The PCD-400A is a carrier-wave measuring instrument for connecting strain gages, strain-gage transducers.
Up to 16 channels is possible, simply dock 4 units.

Products table

All display

PCD-400A
Sensor Interface

PCD-400A
  • CE
  • TEDS

Specifications

Measuring Targets Strain gages and strain-gage transducers
Channels 4
Input Format Balanced differential input
Synchronous Operation A maximum of 4 units for 16 channels
Applicable Gage Resistance Quarter bridge 2-wire system, 3-wire system: 120 Ω
Half bridge system, full bridge system: 120 to 1000 Ω
Input Connector D-sub 37-pin connector
Bridge Excitation 2 Vrms
Gage Factor 2.00 fixed
Balance Adjustment Resistance: Within ±2% (±10k ×10-6 strain)
Capacity: Within 5000 pF
Balance Adjustment Methods Resistance: Auto balance
Capacitance: CST method (Capacitance self-tracking)
Nonlinearity Within ±0.1%FS
Range 200, 500, 1k, 2k, 5k, 10k, 20k ×10-6 strain – 7 steps
Accuracy: within ±0.5%FS
Frequency Response DC to 200 Hz
Deviation: within ±10%
Sampling Frequencies Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz)
LPF Transfer characteristic : 2nd order Butterworth
Cutoff frequencies : 10, 30, 100 Hz, and FLAT (4 steps)
Amplitude ratio at cutoff point : -3±1 dB
Attenuation : (-12 ±1) dB/oct.
A/D Converter 24 bits
Setting Value Storage The range and balance adjustment value etc. are written to nonvolatile memory.
TEDS Reads information from TEDS-installed sensors.
(Input adapters: UI-10A and UI-11A only)
Channel name writing
(If the manufacturer’s ID is Kyowa)
Interfaces USB2.0 (Conforms to High-speed USB standards.
USB3.0 supports.)
Stability Temperature
Zero point: Within ±0.2 ×10-6 strain per ºC
Sensitivity: Within ±0.05%/ºC
Time
Zero point: Within ±1 ×10-6 strain per 8h.
Sensitivity: Within ±0.3%/8h.
Withstand Voltage 250 VAC for 1 minute between input and case
Operating Temperature 0 to 40ºC
Operating Humidity 20 to 85%RH (Non-condensing)
Vibration Resistance ±29.42 m/s2 (3 G), 5 to 200 Hz, (12 cycles for each axis, 10 minutes/cycle)
Power Supply 11 to 16 VDC
Connector type: RM12BRD-4PH (Hirose)
Current Consumption 0.7 A or less (12 VDC)
Dimensions 210 W x 35 H x 157.5 D mm (Excluding protrusions)
Weight Approx. 700 g
EMC Directive EN61326-1 (Class A)
RoHS Directive EN50581
Standard Accessories USB cable N-38 (1 m)
Ground wire P-72 (5 m)
Dynamic data acquisition software DCS-100A (DVD)
Optional Accessories DC power cable P-76 (11 to 16 VDC, 1.8 m)
USB cable N-39 (2 m)
Connection cable N-97 (10 cm)
Stacking connector set ST-1A
Stack fixture CN-20
AC adapter UIA345-12 (For 1 to 4 units of PCD) (For U.S.A.: UNI345-1238)
Conversion adapter FV-1A
Input adapters (At least one is required)
Input adapter for strain-gage transducers UI-10A (TEDS compatible)
Input adapter for strain gages UI-11A (TEDS compatible)
Input adapter for strain gage with operating lever UI-15A
One-touch type input adapter for strain gages UI-16B
One-touch type input adapter UI-55A
Voltage input adapter UI-30A
Data analysis software DAS-200A

Dimensions

PCD-400A

UI-10A
Input Adapter for Strain Gage Transducers

  • TEDS

Specifications

Channels 4
Sensor Connection P12-7
Main Measuring Instruments PCD-400A/430A
Connecting Targets strain gage transducer

UI-11A
Input Adapter for Strain Gages

UI-11A
  • TEDS

Specifications

Channels 4
Sensor Connection Clamp-type terminals
Main Measuring Instruments PCD-400A/430A
Measuring Targets quater bridge (2-wire and 3-wire), half bridge, full bridge

UI-15A
Input Adapter with Operating Levers for Strain Gages

UI-15A

Specifications

Channels 4
Sensor Connection Clamp-type terminals with operating levers
Main Measuring Instruments PCD-400A/430A
Measuring Targets quarter bridge(2-wire and 3-wire), half bridge, full bridge strain gage, strain gage transducer

UI-16B
One-touch Type Input Adapter for Strain Gages

UI-16B

Specifications

Channels 4
Sensor Connection Clamp-type one-touch terminals
Main Measuring Instruments PCD-400A/430A
Measuring Targets quarter bridge (2-wire and 3-wire), half bridge, full bridge strain gage, strain gage transducer

UI-55A
One-touch Type Input Adapter

UI-55A

Specifications

Channels 4
Sensor Connection One-touch type terminals
Main Measuring Instruments PCD-400A/430A
EDX-11A/14A/15A(For full bridge only)
Measuring Targets quarter bridge (2-wire and 3-wire), half bridge, full bridge strain gage, strain gage transducer
Dimensions 105 W × 18.5 H × 35 D mm
Weight 90 g

Page Top