PCD-400B Series
Sensor Interface
· Simple configuration and user-friendly · Carrier wave type · Sampling frequency: Max. 10 kHz · Up to 4 units are stacked for measurement in 16 channels. *Note that the PCD-400B series cannot connect with the PCD-400A/430A.
View details in the General CatalogFeatures
The PCD-400B Series are carrier-wave measuring instruments for measuring strain gages, strain-gage transducers and voltage via an USB interface.
Stress Measurement During PCB Mounting
PCD-400B Series devices contain a built-in bridge circuit, and can be used to build measurement systems with simple structures. Special software for PCB stress measurement is also available to easily handle everything from measuring strain speed to creating reports.
Contents
■Mainbody
PCD-400B (Strain)
PCD-430B (Strain, Voltage)
■Input adapters
Input adapter for strain-gage transducers UI-10A (TEDS compatible)
Input adapter for strain gages UI-11A (TEDS compatible)
Input adapter for strain gage with operating lever UI-15A
One-touch type input adapter for strain gages UI-16B
One-touch type input adapter UI-55A
Voltage input adapter (For PCD-430B) UI-30A
PCD-400B Series
PCD-400B Sensor Interface
Specifications
Measuring Targets | Strain gage and strain-gage transducers |
---|---|
Channels | 4 |
Input Form | Balanced differential input |
Synchronous Measurement | Up to 4 units by 16 channels |
Compatible Gage Resistance | Quarter bridge 2-wire system, 3-wire system: 120 Ω |
Input Connector | D-sub 37-pin connector |
Bridge Excitation | AC 2 Vrms |
Gage Factor | 2.00 fixed |
Balance Adjustment | Resistance: Within ±2% (±10 k ×10-6 strain) |
Balance Adjustment Methods | Resistance: Auto balance |
Nonlinearity | Within ±0.1% FS |
Range | 200, 500, 1 k, 2 k, 5 k, 10 k, and 20 k ×10-6 strain – 7 steps |
Frequency Response | DC to 200 Hz, deviation: Within ±10% |
Sampling Frequencies | Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz) |
LPF | Transfer characteristic: 2nd order Butterworth |
AD Converter | 24 bits |
Holding Setting Condition | The range and balance adjustment value etc. are written to nonvolatile memory. |
TEDS | Reads information from TEDS-installed sensors. |
Interfaces | USB2.0 (Conforms to High-speed USB standards.USB3.0 supported) |
Stability | Temperature |
Withstand Voltage | 250 VAC for 1 minute between input and case |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Vibration Resistance | ±29.42 m/s2 (3 G) |
Power Supply | 11 to 16 VDC |
Current Consumption | 0.7 A or less (12 VDC) |
Dimensions | 210 W x 35 H x 157.5 D mm (Excluding protrusions) |
Weight | Approx. 650 g |
Compliance | Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) Directive 2014/30/EU (EMC) |
Degree of Protection | IP30 (JIS C 0920 / IEC 60529) |
| *The specifications apply to the state in which the temperature has stabilized after a preheating time of 1 hour. |
Standard Accessories | USB cable N-38 (1 m) |
---|---|
Optional Accessories | AC adapter UNI360-1540-AC |
PCD-400B Series
PCD-430B Sensor Interface
Strain/voltage input possible
Specifications
Measuring Mode | Strain measurement mode | Voltage measurement mode |
---|---|---|
Measuring Targets | Strain gages and strain-gage transducers | Voltage |
Channels | 4 | |
Input Form | Balanced differential input | Unbalanced |
Synchronous Measurement | Up to 4 units by 16 channels | |
Compatible Gage Resistance | Quarter bridge 2-wire system, 3-wire system: 120 Ω | - |
Input Connector | D-sub 37-pin connector | |
Bridge Excitation | AC 2 Vrms | - |
Gage Factor | 2.00 fixed | - |
Balance Adjustment | Resistance: Within ±2% (±10 k ×10-6 strain) | - |
Balance Adjustment Methods | Resistance: Auto balance | - |
Nonlinearity | Within ±0.1% FS | - |
Range | 200, 500, 1 k, 2 k, 5 k, 10 k, and 20 k ×10-6 strain, -7 steps | 1, 2, 5, 10, 20, and 50 V -6 steps |
Frequency Response | DC to 200 Hz, Deviation: Within ±10% | DC to 1 kHz, Deviation: Within -3 to 1 dB |
Sampling Frequencies | Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz) | |
LPF | Transfer characteristic: 2nd order Butterworth | 2nd order Butterworth |
HPF | - | 0.2 Hz or OFF (2 steps) |
AD Converter | 24 bits | |
Setting Value Storage | - | The range and balance adjustment value etc. are written to nonvolatile memory. |
TEDS | Strain measurement mode: Reads information from TEDS-installed sensors. | - |
Interfaces | USB2.0 | |
Stability | Temperature | Temperature |
Withstand Voltage | Strain measurement mode: | - |
Operating Temperature | 0 to 40ºC | |
Operating Humidity | 20 to 85% (Non-condensing) | |
Vibration Resistance | ±29.42 m/s2 (3 G) | |
Power Supply | 11 to 16 VDC | |
Current Consumption | 0.9 A or less (12 VDC) | |
Dimensions | 210 W x 35 H x 157.5 D mm (Excluding protrusions) | |
Weight | Approx. 670 g | |
Compliance | Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) Directive 2014/30/EU (EMC) | |
Degree of Protection | IP30 (JIS C 0920 / IEC 60529) | |
| * The specifications apply to the state in which the temperature has stabilized after a preheating time of 1 hour. |
Standard Accessories | USB cable N-38 (1 m) |
---|---|
Optional Accessories | AC adapter UNI360-1540-AC |
Manual
Software
Videos related to this products
Main Compatible Accessories
Input Adapter for Strain-gage Transducers
Input Adapter for Strain Gages
Input Adapter with Operating Levers for Strain Gages
One-touch Type Input Adapter for Strain Gages
Voltage Input Adapter
One-touch Type Input Adapter
Stacking kit for PCD-400/430B
S32-7 to 4 conductors with shield wire, bared at the tip
Fork-shape crimp-style terminal (M6) to bared at the tip
S32-7 to 4 conductors with shield wire, bared at the tip
AC adapter